A study of long term operation and reliability of heated atomic force microscope cantilevers

被引:4
|
作者
Kim, Hoe Joon [1 ]
King, William P. [1 ]
机构
[1] Univ Illinois, Dept Mech Sci & Engn, Urbana, IL 61801 USA
关键词
AFM; microcantilever; reliability; dopant diffusion; DATA-STORAGE; SILICON; DIFFUSION; PHOSPHORUS; MECHANISMS; AFM;
D O I
10.1088/0960-1317/25/6/065003
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper presents a detailed investigation of the long term operation and reliability of heated atomic force microscope (AFM) cantilevers. The electrical and thermal operation of these heated AFM cantilevers depends upon the distribution of dopants within the silicon cantilever. Over long term operation, or for operation at very high temperatures, the cantilevers' electro-thermal properties can change due to dopant diffusion at the junction between the cantilever heater and the cantilever leg. We analyze this change in cantilever behavior at high heating temperatures, up to 1100 degrees C, and long heating times, up to 108 h. At extreme heating conditions, the cantilevers' electro-thermal properties change by as much as 24%, resulting in significant errors in cantilever calibration. The main failure mechanism of these heated AFM cantilevers is the melting of heater-leg junction at high cantilever temperatures, as high junction resistances induced the localized heating. Understanding and accounting for long term operational reliability is crucial for consistent and repeatable uses of heated cantilevers, especially for applications in material property measurements and nanomanufacturing.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Direct three dimensional nanoscale thermal lithography at high speeds using heated atomic force microscope cantilevers
    Hua, Yueming
    Saxena, Shubham
    Lee, Jung Chul
    King, William P.
    Henderson, Clifford L.
    EMERGING LITHOGRAPHIC TECHNOLOGIES XI, PTS 1 AND 2, 2007, 6517
  • [22] Tilt of atomic force microscope cantilevers: Effect on friction measurements
    Wang, Fei
    Zhao, Xuezeng
    PROGRESSES IN FRACTURE AND STRENGTH OF MATERIALS AND STRUCTURES, 1-4, 2007, 353-358 : 742 - 745
  • [23] Note: Improved calibration of atomic force microscope cantilevers using multiple reference cantilevers
    Sader, John E.
    Friend, James R.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2015, 86 (05):
  • [24] Normal and torsional spring constants of atomic force microscope cantilevers
    Green, CP
    Lioe, H
    Cleveland, JP
    Proksch, R
    Mulvaney, P
    Sader, JE
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (06): : 1988 - 1996
  • [25] Dynamic simulation of atomic force microscope cantilevers oscillating in liquid
    Martin, Michael James
    Fathy, Hosam K.
    Houston, Brian H.
    JOURNAL OF APPLIED PHYSICS, 2008, 104 (04)
  • [26] Calibration of the lateral spring constant of atomic force microscope cantilevers
    Song, Yunpeng
    Wu, Sen
    Xu, Linyan
    Fu, Xing
    AOPC 2015: MICRO/NANO OPTICAL MANUFACTURING TECHNOLOGIES; AND LASER PROCESSING AND RAPID PROTOTYPING TECHNIQUES, 2015, 9673
  • [27] Method to determine the spring constant of atomic force microscope cantilevers
    Gibson, CT
    Johnson, DJ
    Anderson, C
    Abell, C
    Rayment, T
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (02): : 565 - 567
  • [28] Actuation of atomic force microscope cantilevers by acoustic radiation pressure
    Onaran, AG
    Degertekin, FL
    Hadimioglu, B
    Sulchek, T
    Quate, CF
    2001 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2001, : 509 - 512
  • [29] FRICTION EFFECTS IN THE DEFLECTION OF ATOMIC-FORCE MICROSCOPE CANTILEVERS
    WARMACK, RJ
    ZHENG, XY
    THUNDAT, T
    ALLISON, DP
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1994, 65 (02): : 394 - 399
  • [30] Improved acoustic excitation of atomic force microscope cantilevers in liquids
    Maali, A
    Hurth, C
    Cohen-Bouhacina, T
    Couturier, G
    Aimé, JP
    APPLIED PHYSICS LETTERS, 2006, 88 (16)