Biomechanical Comparison of Fixation With a Single Screw Versus Two Kirschner Wires in Distal Chevron Osteotomies of the First Metatarsal: A Cadaver Study

被引:16
|
作者
Trost, Matthias [1 ]
Bredow, Jan [2 ]
Boese, Christoph Kolja [3 ]
Loweg, Lennard
Schulte, Tobias Ludger [1 ]
Scaal, Martin [4 ]
Eysel, Peer [5 ,6 ]
Oppermann, Johannes [6 ]
机构
[1] Univ Bochum, St Josef Hosp, Dept Orthopaed & Traumatol, Gudrunstr 56, D-44791 Bochum, Germany
[2] Schon Klin, Dept Spinal Surg, Dusseldorf, Germany
[3] Cologne Univ Hosp, Dept Orthopaed & Traumatol, Cologne, Germany
[4] Univ Cologne, Inst Anat 2, Cologne, Germany
[5] Cologne Univ Hosp, Dept Orthopaed & Traumatol, Cologne, Germany
[6] Univ Cologne, Cologne Ctr Musculoskeletal Biomech, Cologne, Germany
来源
JOURNAL OF FOOT & ANKLE SURGERY | 2018年 / 57卷 / 01期
关键词
cantilever; foot; hallux valgus; load; osteosynthesis; RANDOMIZED CONTROLLED-TRIAL; SEVERE HALLUX-VALGUS; SOFT-TISSUE RELEASE; SHAFT OSTEOTOMIES; ANKLE ARTHRODESIS; STABILITY;
D O I
10.1053/j.jfas.2017.08.014
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Distal chevron osteotomy is a common procedure for surgical correction of hallux valgus. Osteosynthesis with 1 screw or 2 Kirschner wires has been commonly used. We compared the stability of the 2 techniques in distal chevron osteotomy. Sixteen first metatarsals from fresh-frozen human cadaver feet (9 different cadaveric specimens) were used. A standardized distal chevron osteotomy was performed. One first metatarsal from each pair was assigned to group 1 (3.5-mm cortical screw; n = 8) and one to group 2 (two 1.6-mm Kirschner wires; n = 8). Using a materials testing machine, the head of the first metatarsals was loaded in 2 different configurations (cantilever and physiologic) in succession. In the cantilever configuration, the relative stiffness of the osteosynthesis compared with intact bone was 59% +/- 27% in group 1 and 68% +/- 18% in group 2 (p = .50). In the physiologic configuration, it was 38% +/- 25% in group 1 and 35% +/- 7% in group 2 (p = .75). The failure strength in the cantilever configuration was 187 +/- 105 N in group 1 and 259 +/- 71 N in group 2 (p = .21). No statistically significant differences were found in stability between the 2 techniques. The use of 1 screw or 2 Kirschner wires had no significant differences in their biomechanical loading capacity for osteosynthesis in distal chevron osteotomies for treatment of hallux valgus. (C) 2017 by the American College of Foot and Ankle Surgeons. All rights reserved.
引用
收藏
页码:95 / 99
页数:5
相关论文
共 50 条
  • [1] Biomechanical Comparison of a Novel Method of Tricortical Kirschner Wire Fixation of Distal Chevron Osteotomies of the First Metatarsal Versus Traditional Kirschner Wire and Versus Screw Fixation
    Chen, Jie
    Black, Natalie R.
    Morris, Randall
    Panchbhavi, Vinod K.
    FOOT & ANKLE SPECIALIST, 2022, 15 (05) : 426 - 431
  • [2] Biomechanical comparison of conventional versus modified technique in distal chevron osteotomies of the first metatarsal: A cadaver study
    Trost, Matthias
    Knieps, Matthias
    Loweg, Lennard
    Neiss, Wolfram Friedrich
    Eysel, Peer
    Frey, Soenke
    Schulte, Tobias Ludger
    Bredow, Jan
    FOOT AND ANKLE SURGERY, 2019, 25 (05) : 665 - 669
  • [3] Comparison of Clinical and Radiological Results of Two Fixation Materials after Distal Chevron Osteotomy for Hallux Valgus? - Two Kirschner Wires versus Single Screw Fixation
    Bilgin, E.
    Kececi, T.
    Turgut, A.
    Adiyeke, L.
    Kilinc, B. E.
    ACTA CHIRURGIAE ORTHOPAEDICAE ET TRAUMATOLOGIAE CECHOSLOVACA, 2020, 87 (05) : 350 - 355
  • [4] Comparison of screw-fixation stabilities of first metatarsal shaft osteotomies: a biomechanical study
    Unal, A. Meric
    Baran, Onder
    Uzun, Bora
    Turan, Ahmet C.
    ACTA ORTHOPAEDICA ET TRAUMATOLOGICA TURCICA, 2010, 44 (01) : 70 - 75
  • [5] Axial Loading Screw Fixation for Chevron Type Osteotomies of the Distal First Metatarsal: A Retrospective Outcomes Analysis
    Murphy, Ryan M.
    Fallat, Lawrence M.
    Kish, John P.
    JOURNAL OF FOOT & ANKLE SURGERY, 2014, 53 (01): : 52 - 54
  • [6] Comparison of bioabsorbable pins and Kirschner wires in the fixation of chevron osteotomies for hallux valgus
    Winemaker, MJ
    Amendola, A
    FOOT & ANKLE INTERNATIONAL, 1996, 17 (10) : 623 - 628
  • [7] Is the magnesium screw as stable as the titanium screw in the fixation of first metatarsal distal chevron osteotomy? A comparative biomechanical study on sawbones models
    Sahin, Adem
    Gulabi, Deniz
    Buyukdogan, Halil
    Agar, Anil
    Kilic, Bulent
    Mutlu, Ibrahim
    Erturk, Cemil
    JOURNAL OF ORTHOPAEDIC SURGERY, 2021, 29 (03)
  • [8] Supplementary axial kirschner wire fixation for crescentic and ludloff proximal metatarsal osteotomies: A biomechanical study
    Jung, HG
    Guyton, GP
    Parks, BG
    Title, CI
    Dom, KJ
    Nguyen, A
    Schon, LC
    FOOT & ANKLE INTERNATIONAL, 2005, 26 (08) : 620 - 626
  • [9] Biomechanical analysis of two fixation methods for proximal chevron osteotomy of the first metatarsal
    Reinhard Schuh
    Jochen Gerhard Hofstaetter
    Emir Benca
    Madeleine Willegger
    Gobert von Skrbensky
    Shahin Zandieh
    Axel Wanivenhaus
    Johannes Holinka
    Reinhard Windhager
    International Orthopaedics, 2014, 38 : 983 - 989
  • [10] Biomechanical analysis of two fixation methods for proximal chevron osteotomy of the first metatarsal
    Schuh, Reinhard
    Hofstaetter, Jochen Gerhard
    Benca, Emir
    Willegger, Madeleine
    von Skrbensky, Gobert
    Zandieh, Shahin
    Wanivenhaus, Axel
    Holinka, Johannes
    Windhager, Reinhard
    INTERNATIONAL ORTHOPAEDICS, 2014, 38 (05) : 983 - 989