Skyshine photon doses from 6 and 10 MV medical linear accelerators

被引:0
|
作者
de Paiva, Eduardo [1 ]
da Rosa, Luiz A. R. [1 ]
机构
[1] IRD CNEN MCT, Serv Fis Med Radioterapia & Med Nucl, BR-22780160 Rio De Janeiro, RJ, Brazil
来源
关键词
skyshine; shielding; medical accelerator;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The skyshine radiation phenomenon consists of the scattering of primary photon beams in the atmosphere above the roof of a medical linear accelerator facility, generating an additional dose at ground level in the vicinity of the treatment room. Thus, with respect to radioprotection, this situation plays an important role when the roof is designed with little shielding and there are buildings next to the radiotherapy treatment room. In literature, there are few reported skyshine-measured doses and these contain poor agreement with empirical calculations. In this work, we carried out measurements of skyshine photon dose rates produced from eight different 6 and 10 MV medical accelerators. Each measurement was performed outside the room facility, with the beam positioned in the upward direction, at a horizontal distance from the target and for a 40 cm x 40 cm maximum photon field size at the accelerator isocenter. Measured dose-equivalent rates results were compared with calculations obtained by an empirical expression, and differences between them deviated in one or more order of magnitude.
引用
收藏
页码:215 / 219
页数:5
相关论文
共 50 条
  • [21] Dosimetric characteristics of 6 and 10 MV unflattened photon beams
    Kragl, Gabriele
    Wetterstedt, Sacha Af
    Knaeusl, Barbara
    Lind, Marten
    McCavana, Patrick
    Knoos, Tommy
    McClean, Brendan
    Georg, Dietmar
    RADIOTHERAPY AND ONCOLOGY, 2009, 93 (01) : 141 - 146
  • [22] Measurement of the leakage radiation from linear accelerators in the backward direction for 4, 6, 10, 15, and 18 MV X-ray energies
    Jaradat, Adnan K.
    Biggs, Peter J.
    HEALTH PHYSICS, 2007, 92 (04): : 387 - 395
  • [23] INVESTIGATION OF PHOTONEUTRON CONTAMINATION FROM THE 18-MV PHOTON BEAM IN A MEDICAL LINEAR ACCELERATOR
    Karaman, Onur
    Tanir, Ayse Gunes
    Karaman, Ceren
    MATERIALI IN TEHNOLOGIJE, 2019, 53 (05): : 699 - 704
  • [24] SURFACE DOSES FROM A 4MV LINEAR ACCELERATOR
    GRAY, L
    PHYSICS IN MEDICINE AND BIOLOGY, 1972, 17 (05): : 672 - &
  • [25] Dosimetric Comparison of SBRT VMAT Treatment Plans Generated for 6 MV, 6 MV FFF, and 10 MV FFF Photon Beams
    Wilson, D.
    Wang, B.
    Dunlap, N.
    MEDICAL PHYSICS, 2015, 42 (06) : 3404 - 3404
  • [26] Novel 6 MV X-ray photoneutron detection and dosimetry of medical accelerators
    Sohrabi, Mehdi
    Hakimi, Amir
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 36 : 103 - 109
  • [27] Redesign of a 6 MV Linear Accelerator Waveguide to Produce Energies Up to 10 MV
    Baillie, D.
    Aubin, J. St.
    Fallone, B.
    Steciw, S.
    MEDICAL PHYSICS, 2012, 39 (06) : 3859 - 3859
  • [28] CHARACTERISTICS OF AN 18 MV PHOTON-BEAM FROM A THERAC 20 MEDICAL LINEAR-ACCELERATOR
    PATTERSON, MS
    SHRAGGE, PC
    MEDICAL PHYSICS, 1981, 8 (03) : 312 - 318
  • [29] Distribution of secondary electrons in the photon beam of a 6-MV and 15-MV linear accelerator
    Wiezorek, T.
    Scheithauer, M.
    Salz, H.
    Meininger, G.
    Harder, D.
    Zeitschrift fur Medizinische Physik, 2000, 10 (04): : 265 - 273
  • [30] X-RAY DEPTH DOSES FROM LINEAR ACCELERATORS IN ENERGY-RANGE FROM 10 TO 32 MEV
    PODGORSAK, EB
    RAWLINSON, JA
    JOHNS, HE
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1975, 123 (01) : 182 - 191