Novel 6 MV X-ray photoneutron detection and dosimetry of medical accelerators

被引:20
|
作者
Sohrabi, Mehdi [1 ]
Hakimi, Amir [1 ]
机构
[1] Amirkabir Univ Technol, Dept Energy Engn & Phys, Hlth Phys & Dosimetry Res Lab, Tehran, Iran
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2017年 / 36卷
关键词
Novel; 6 MV X-rays; Photoneutrons; Detection; Dosimetry; Polycarbonate/B-10 (with/without cadmium); DEPTH-DOSE-EQUIVALENT; DYNAMIC MLC IMRT; LINEAR-ACCELERATOR; NEUTRON DOSIMETRY; CONTAMINATION; RADIOTHERAPY; BEAMS; FIELD; SIZE;
D O I
10.1016/j.ejmp.2017.03.020
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Dosimetry of fast, epithermal and thermal photoneutrons in 6 MV X-ray beams of two medical accelerators were studied by novel dosimetry methods. Methods: A Siemens ONCOR and an Elekta COMPACT medical accelerators were used. Fast, epithermal and thermal photoneutron dose equivalents in 10 cm x 10 cm 6 MV X-rays fields were determined in air and on surface of a polyethylene phantom in X and Y directions. Polycarbonate dosimeters as bare or with enriched B-10 convertors (with or without cadmium covers) were used applying a 50 Hz-HV electrochemical etching method. Results: Fast, epithermal and thermal photoneutron dose equivalents were efficiently determined respectively as similar to 1145.8, similar to 45.3 and similar to 170.6 mu Sv in air and similar to 1888.5, similar to 96.1 and similar to 640.6 mu Sv on phantom per 100 Gy X-rays at the isocenter of Siemens ONCOR accelerator in air. The dose equivalent is maximum at the isocenter which decreases as distance from it increases reaching a constant level. Tissue-to-air ratios are constants up to 15 cm from the isocenter. No photoneutrons was detected in the Elekta COMPACT accelerator. Conclusions: Fast, epithermal and thermal photoneutron dosimetry of 6 MV X-rays were made by novel dosimetry methods in a Siemens ONCOR accelerator with sum dose equivalent per Gy of similar to 0.0014% mSv with similar to 0.21 MeV mean energy at the isocenter; i. e. similar to 150 times smaller than that of 18 MV X-rays. This observation assures clinical safety of 6 MV X-rays in particular in single-mode machines like Elekta COMPACT producing no photoneutrons due to no "beryllium exit window" in the head structure. (C) 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:103 / 109
页数:7
相关论文
共 50 条
  • [1] NEUTRON DOSIMETRY IN HIGH-ENERGY X-RAY BEAMS PRODUCED IN MEDICAL ACCELERATORS
    SOHRABI, M
    MORGAN, KZ
    MCGINLEY, PH
    HEALTH PHYSICS, 1977, 33 (06): : 687 - 687
  • [2] Photoneutron depth dose equivalent distributions in high-energy X-ray medical accelerators by a novel position-sensitive dosimeter
    Hakimi, Amir
    Sohrabi, Mehdi
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 36 : 73 - 80
  • [3] EFFECTS OF FIELD SIZE AND DEPTH ON PHOTONEUTRON DOSE EQUIVALENT DISTRIBUTIONS IN AN 18 MV X-RAY MEDICAL ACCELERATOR
    Hakimi, Amir
    Sohrabi, Mehdi
    Mahdavi, Seied Rabie
    RADIATION PROTECTION DOSIMETRY, 2017, 176 (04) : 354 - 364
  • [4] Scatter fractions from linear accelerators with x-ray energies from 6 to 24 MV
    Taylor, PL
    Rodgers, JE
    Shobe, J
    MEDICAL PHYSICS, 1999, 26 (08) : 1442 - 1446
  • [5] Thermal limits on MV x-ray production by bremsstrahlung targets in the context of novel linear accelerators
    Wang, Jinghui
    Trovati, Stefania
    Borchard, Philipp M.
    Loo, Billy W., Jr.
    Maxim, Peter G.
    Fahrig, Rebecca
    MEDICAL PHYSICS, 2017, 44 (12) : 6610 - 6620
  • [6] Photoneutron radiation field of ducts in barrier of 15 MV medical electron accelerators
    Deng, Lei
    Zhou, Ning
    Chen, Yi-shui
    Tu, Yu
    RADIATION PHYSICS AND CHEMISTRY, 2017, 140 : 340 - 344
  • [7] Effect of electron contamination of a 6 MV x-ray beam on near surface diode dosimetry
    Edwards, C. R.
    Mountford, P. J.
    Moloney, A. J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (24): : 6471 - 6482
  • [9] DOSIMETRY OF 6-MV X-RAY-BEAM PENUMBRA
    METCALFE, P
    KRON, T
    ELLIOTT, A
    WONG, T
    HOBAN, P
    MEDICAL PHYSICS, 1993, 20 (05) : 1439 - 1445
  • [10] The influence of a novel transmission detector on 6 MV x-ray beam characteristics
    Venkataraman, Sankar
    Malkoske, Kyle E.
    Jensen, Martin
    Nakonechny, Keith D.
    Asuni, Ganiyu
    McCurdy, Boyd M. C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (10): : 3173 - 3183