Gas chromatographic determination of catecholestrogens following isolation by solid-phase extraction

被引:11
|
作者
Pinnella, KD
Cranmer, BK
Tessari, JD [1 ]
Cosma, GN
Veeramachaneni, DNR
机构
[1] Colorado State Univ, Dept Environm Hlth, Ft Collins, CO 80523 USA
[2] Pharmacia & Upjohn Inc, Kalamazoo, MI 49001 USA
[3] Colorado State Univ, Dept Physiol, Ft Collins, CO 80523 USA
来源
JOURNAL OF CHROMATOGRAPHY B | 2001年 / 758卷 / 02期
关键词
catecholestrogens;
D O I
10.1016/S0378-4347(01)00164-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A sensitive and specific assay for the determination of the catecholestrogens 2-hydroxyestradiol (2-OHE2) and 4-hydroxyestradiol (4-OHE2) using gas chromatography with electron-capture detection (GC-ECD) is described. The formation of 2- and 4-OHE2 was assessed following activation of 17 beta -estradiol in the microsomal fraction of female rat livers. The analytes were isolated by solid-phase extraction. derivatized to their heptafluorobutyryl esters with heptafluoro-butyric acid anhydride, and subjected to solvent exchange prior to analysis: this resulted in minimal chromatographic interference, long column life, and stable derivatized analytes. Derivatized catechols were separated and confirmed with dual column chromatography (DB-5 and DB-608) and quantitated using GC-ECD. The DB-608 column was preferred for quantitation as it provided better 4-OHE2 resolution from interference. Key validation parameters for the assay include sensitivity, intra- and inter-assay precision, and accuracy. instrument sensitivity and limits of detection (LOD) and quantitation (LOQ) were determined statistically from fortification data approaching expected limits. For 2-OHE2 and 4-OHE2, respective values for these parameters were. instrument sensitivities of 0.4 and 0.7 pg, LODs of 0.8 and 1.3 ng/mg, and LOQs of 2.6 and 4.3 ng/mg. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:145 / 152
页数:8
相关论文
共 50 条