A free-air ionization chamber with a large aperture diaphragm

被引:1
|
作者
Takata, N. [1 ]
机构
[1] AIST, NMIJ, Ionizing Radiat Sect, Tsukuba, Ibaraki 3058568, Japan
来源
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT | 2010年 / 619卷 / 1-3期
关键词
Parallel-plate free-air ionization chamber; Ionization volume; Charge collection volume; Electric field distortion; Guard electrode; Scattered X-rays; Air kerma; Exposure; X-RAYS;
D O I
10.1016/j.nima.2010.02.083
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Calculations of the electric field distributions in free-air ionization chambers reveal that the distortion of the charge collection volume is small even for wide X-ray beam passage if the diaphragm and the X-ray shielding box are kept at a potential equal to half that applied to the high-voltage electrode. Applying this potential to the diaphragm and the shielding box permits a larger aperture diaphragm to be used. This will allow a wider X-ray beam to enter the chamber, thus generating a larger signal. In addition, the distance between the diaphragm and the charge collection volume can be shortened to reduce the amount of X-ray attenuation. It is also possible to calibrate a dosimeter against a free-air ionization chamber that has a diaphragm whose aperture size is equal to the size of the dosimeter in an X-ray field that is collimated to the same size. This is important since free-air ionization chambers are not sensitive to X-rays that are incident at large angles, such as those scattered by the collimator, filters and air. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:216 / 220
页数:5
相关论文
共 50 条
  • [41] COMPARISON OF UNITED-STATES AND CANADIAN FREE-AIR IONIZATION CHAMBERS
    AITKEN, JH
    DELAVERGNE, L
    HENRY, WH
    LOFTUS, TP
    BRITISH JOURNAL OF RADIOLOGY, 1962, 35 (409): : 65 - 70
  • [42] PLATE-SEPARATION REQUIREMENTS FOR STANDARD FREE-AIR IONIZATION CHAMBERS
    ATTIX, FH
    DELAVERGNE, L
    JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1954, 53 (06): : 393 - 402
  • [43] FREE-AIR TURBULENCE
    ANDERSON, AD
    JOURNAL OF METEOROLOGY, 1957, 14 (06): : 477 - 494
  • [44] THE NEW NPL STANDARD FREE-AIR IONIZATION-CHAMBER FOR X-RAYS GENERATED AT UP TO 50KV
    MARSH, ARS
    PHYSICS IN MEDICINE AND BIOLOGY, 1980, 25 (05): : 949 - 949
  • [45] Conception and realization of a parallel-plate free-air ionization chamber for the absolute dosimetry of an ultrasoft X-ray beam
    Groetz, J. -E.
    Ounoughi, N.
    Mavon, C.
    Belafrites, A.
    Fromm, M.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2014, 85 (08):
  • [46] AN INVESTIGATION OF THE INFLUENCE OF APPLIED VOLTAGE ON IONIZATION DUE TO ELECTRONS IN FREE-AIR IONIZATION-CHAMBERS
    TAKATA, N
    MATIULLAH
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 324 (1-2): : 226 - 231
  • [47] FIELD DISTORTION IN A PARALLEL-PLATE FREE-AIR CHAMBER - A COMPENSATORY EFFECT
    KEMP, LAW
    READ, LR
    WOODALL, JE
    PHYSICS IN MEDICINE AND BIOLOGY, 1968, 13 (01): : 91 - &
  • [48] THE CANADIAN STANDARD FREE-AIR CHAMBER FOR MEDIUM QUALITY X-RAYS
    HENRY, WH
    GARRETT, C
    CANADIAN JOURNAL OF PHYSICS, 1960, 38 (12) : 1677 - 1689
  • [49] Novel approach for the Monte Carlo calculation of free-air chamber correction factors
    Mainegra-Hing, Ernesto
    Reynaert, Nick
    Kawrakow, Iwan
    MEDICAL PHYSICS, 2008, 35 (08) : 3650 - 3660
  • [50] NBS FREE-AIR CHAMBER FOR MEASUREMENT OF 10 TO 60 KV X RAYS
    LAMPERTI, PJ
    WYCKOFF, HO
    JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS SECTION C-ENGINEERING AND INSTRUMENTATION, 1965, C 69 (01): : 39 - +