A free-air ionization chamber with a large aperture diaphragm

被引:1
|
作者
Takata, N. [1 ]
机构
[1] AIST, NMIJ, Ionizing Radiat Sect, Tsukuba, Ibaraki 3058568, Japan
来源
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT | 2010年 / 619卷 / 1-3期
关键词
Parallel-plate free-air ionization chamber; Ionization volume; Charge collection volume; Electric field distortion; Guard electrode; Scattered X-rays; Air kerma; Exposure; X-RAYS;
D O I
10.1016/j.nima.2010.02.083
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Calculations of the electric field distributions in free-air ionization chambers reveal that the distortion of the charge collection volume is small even for wide X-ray beam passage if the diaphragm and the X-ray shielding box are kept at a potential equal to half that applied to the high-voltage electrode. Applying this potential to the diaphragm and the shielding box permits a larger aperture diaphragm to be used. This will allow a wider X-ray beam to enter the chamber, thus generating a larger signal. In addition, the distance between the diaphragm and the charge collection volume can be shortened to reduce the amount of X-ray attenuation. It is also possible to calibrate a dosimeter against a free-air ionization chamber that has a diaphragm whose aperture size is equal to the size of the dosimeter in an X-ray field that is collimated to the same size. This is important since free-air ionization chambers are not sensitive to X-rays that are incident at large angles, such as those scattered by the collimator, filters and air. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:216 / 220
页数:5
相关论文
共 50 条
  • [31] THE RONTGEN - DEFICIENCIES IN PRESENT-DAY FREE-AIR CHAMBER STANDARDS
    KEMP, LAW
    NATURE, 1953, 172 (4366) : 35 - 35
  • [32] Characteristics of a miniature parallel-plate free-air ionization chamber for measuring the intensity of synchrotron radiation from an undulator
    Nariyama, N
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (09): : 2860 - 2862
  • [33] FREE-AIR IONIZATION DUE TO COSMIC RADIATIONS AT SEA-LEVEL
    KATSURAYAMA, K
    TSUJIMOTO, T
    YAMASAKI, K
    URABE, I
    OKAMOTO, K
    YOSHIMOTO, T
    NAKASHIMA, Y
    FURUTA, S
    SASAGAWA, S
    ITOH, K
    YAMAMOTO, H
    MARUYAMA, S
    OSHIMA, T
    TATEISHI, N
    HEALTH PHYSICS, 1979, 36 (04): : 558 - 558
  • [34] FREE-AIR AND THIMBLE IONIZATION CHAMBERS FOR GRENZ-RAY DOSIMETRY
    OOSTERKAMP, WJ
    PROPER, J
    ACTA RADIOLOGICA, 1952, 37 (01) : 33 - 43
  • [35] MEASUREMENT OF FIELD DISTORTION IN FREE-AIR IONIZATION CHAMBERS BY ANALOG METHOD
    MILLER, W
    KENNEDY, RJ
    JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1955, 55 (05): : 291 - 297
  • [36] The performance of the INER improved free-air ionization chamber in the comparison of air kerma calibration coefficients for medium-energy X-rays
    Lee, JH
    Kotler, LH
    Büermann, L
    Hwang, WS
    Chiu, JH
    Wang, CF
    RADIATION MEASUREMENTS, 2005, 39 (01) : 1 - 10
  • [37] FIELD DISTORTION IN GUARDED-WALL FREE-AIR IONIZATION CHAMBERS
    UOTILA, IK
    METROLOGIA, 1980, 16 (01) : 15 - 19
  • [38] DETERMINATION OF SATURATION CUREENTS IN FREE-AIR IONIZATION CHAMBERS BY EXTRAPOLATION METHODS
    SCOTT, PB
    GREENING, JR
    PHYSICS IN MEDICINE AND BIOLOGY, 1963, 8 (1-4): : 51 - &
  • [39] Significant discrepancies in air kerma rates measured with free-air and cavity ionization chambers
    Bueermann, L.
    Hilgers, G.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (01): : 477 - 480
  • [40] Current saturation in free-air ionization chambers with chopped synchrotron radiation
    Nariyama, Nobuteru
    JOURNAL OF SYNCHROTRON RADIATION, 2013, 20 : 698 - 704