Performance characteristics of the new Eurosets magnetically suspended centrifugal pump

被引:2
|
作者
Berdajs, Denis [1 ]
von Segesser, Ludwig K. [2 ]
Maisano, Francesco [3 ]
Milano, Guiseppina [2 ]
Ferrari, Enrico [2 ]
机构
[1] Univ Hosp Basel, Dept Cardiac Surg, Spitalstr 21, CH-4031 Basel, Switzerland
[2] Dept Surg & Anesthesiol, Cardiovasc Res, Lausanne, Switzerland
[3] Univ Hosp Zurich, Dept Cardiac Surg, Zurich, Switzerland
来源
PERFUSION-UK | 2021年 / 36卷 / 02期
关键词
ECLS; ECMO; centrifugal pump; hydrodynamic performance; hemolysis; EXTRACORPOREAL MEMBRANE-OXYGENATION; LIFE-SUPPORT; BLOOD PUMP; CENTRIMAG; COMPLICATIONS;
D O I
10.1177/0267659120931990
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: The aim was to evaluate the performance of a newly developed magnetically suspended centrifugal pump head intended for use as a ventricular assistance device with a newly developed extracorporeal membrane oxygenator setup. Methods: In an experimental setup, an extracorporeal membrane oxygenator circuit was established in three calves with a mean weight of 68.2 +/- 2.0 kg. A magnetically levitated centrifugal pump was tested, along with a newly designed extracorporeal membrane oxygenator console, at three different flow ranges: (a) 0.0 to 5.2 L/min, (b) 0.0 to 7.1 L/min, and (c) 0.0 to 6.0 L/min. For each setup, the animals were supported by a circuit for 6 h. Blood samples were collected just before caridiopulmonary bypass (CPB) after 10 min on bypass and after 1, 2, 5, and 6 h of perfusion for hemolysis determination and biochemical tests. Values were recorded for blood pressure, mean flow, and pump rotational speed. Analysis of variance was used for repeated measurements. Results: Mean pump flows achieved during the three 6 h pump runs for the three pump heads studied were as follows: (a) flow range 0.0 to 5.2 L/min, 3.6 +/- 1.5 L/min, (b) flow range 0.0 to 7.1 L/min, 4.9 +/- 1.3 L/min, and (c) flow range 0.0 to 6.0 L/min, 3.8 +/- 1.5 L/min. Blood trauma, evaluated by plasma hemoglobin and lactate dehydrogenase levels, did not help in detecting any significant hemolysis. Thrombocytes and white blood cell count profiles showed no significant differences between the groups at the end of the 6 h perfusion. At the end of testing, no clot deposition was found in the oxygenator, and there was no evidence of peripheral emboli. Conclusion: The results suggest that the newly developed magnetically suspended centrifugal pump head provides satisfactory hydrodynamic performance in an acute perfusion scenario without increasing hemolysis.
引用
收藏
页码:183 / 189
页数:7
相关论文
共 50 条
  • [41] Performance characteristics of centrifugal pump using central composite design
    Seenivasan, M.
    Venugopal, T.
    Gopalsamy, N.
    Ulaganathan, K.
    Vineeth, S.
    MATERIALS TODAY-PROCEEDINGS, 2022, 62 : 2105 - 2108
  • [42] PERFORMANCE-CHARACTERISTICS OF A MULTIPLE-DISK CENTRIFUGAL PUMP
    RODDY, PJ
    DARBY, R
    MORRISON, GL
    JENKINS, PE
    JOURNAL OF FLUIDS ENGINEERING-TRANSACTIONS OF THE ASME, 1987, 109 (01): : 51 - 57
  • [43] A Passive Magnetically and Hydrodynamically Suspended Rotary Blood Pump
    Stoiber, Martin
    Grasl, Christian
    Pirker, Stefan
    Raderer, Franz
    Schistek, Roland
    Huber, Leopold
    Gittler, Philipp
    Schima, Heinrich
    ARTIFICIAL ORGANS, 2009, 33 (03) : 250 - 257
  • [44] Field Dynamic Balancing for Magnetically Suspended Turbomolecular Pump
    Chen, Qi
    Li, Jinlei
    SENSORS, 2023, 23 (13)
  • [45] Estimation of changes in dynamic hydraulic force in a magnetically suspended centrifugal blood pump with transient computational fluid dynamics analysis
    Masuzawa, Toru
    Ohta, Akiko
    Tanaka, Nobuatu
    Qian, Yi
    Tsukiya, Tomonori
    JOURNAL OF ARTIFICIAL ORGANS, 2009, 12 (03) : 150 - 159
  • [46] Estimation of changes in dynamic hydraulic force in a magnetically suspended centrifugal blood pump with transient computational fluid dynamics analysis
    Toru Masuzawa
    Akiko Ohta
    Nobuatu Tanaka
    Yi Qian
    Tomonori Tsukiya
    Journal of Artificial Organs, 2009, 12 : 150 - 159
  • [47] VentrAssist hydrodynamically suspended, open, centrifugal blood pump
    Watterson, PA
    Woodard, JC
    Ramsden, VS
    Reizes, JA
    ARTIFICIAL ORGANS, 2000, 24 (06) : 475 - 477
  • [48] EVALUATING CENTRIFUGAL PUMP CHARACTERISTICS
    BAIBAKOV, OV
    MATVEEV, IV
    RUSSIAN ENGINEERING JOURNAL, 1973, 53 (10): : 13 - 16
  • [49] INVESTIGATION OF CENTRIFUGAL PUMP CHARACTERISTICS
    Repsa, Edgars
    Kronbergs, Eriks
    20TH INTERNATIONAL SCIENTIFIC CONFERENCE ENGINEERING FOR RURAL DEVELOPMENT, 2021, : 551 - 556
  • [50] Dimensionless characteristics of centrifugal pump
    Timár, P