Performance characteristics of the new Eurosets magnetically suspended centrifugal pump

被引:2
|
作者
Berdajs, Denis [1 ]
von Segesser, Ludwig K. [2 ]
Maisano, Francesco [3 ]
Milano, Guiseppina [2 ]
Ferrari, Enrico [2 ]
机构
[1] Univ Hosp Basel, Dept Cardiac Surg, Spitalstr 21, CH-4031 Basel, Switzerland
[2] Dept Surg & Anesthesiol, Cardiovasc Res, Lausanne, Switzerland
[3] Univ Hosp Zurich, Dept Cardiac Surg, Zurich, Switzerland
来源
PERFUSION-UK | 2021年 / 36卷 / 02期
关键词
ECLS; ECMO; centrifugal pump; hydrodynamic performance; hemolysis; EXTRACORPOREAL MEMBRANE-OXYGENATION; LIFE-SUPPORT; BLOOD PUMP; CENTRIMAG; COMPLICATIONS;
D O I
10.1177/0267659120931990
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: The aim was to evaluate the performance of a newly developed magnetically suspended centrifugal pump head intended for use as a ventricular assistance device with a newly developed extracorporeal membrane oxygenator setup. Methods: In an experimental setup, an extracorporeal membrane oxygenator circuit was established in three calves with a mean weight of 68.2 +/- 2.0 kg. A magnetically levitated centrifugal pump was tested, along with a newly designed extracorporeal membrane oxygenator console, at three different flow ranges: (a) 0.0 to 5.2 L/min, (b) 0.0 to 7.1 L/min, and (c) 0.0 to 6.0 L/min. For each setup, the animals were supported by a circuit for 6 h. Blood samples were collected just before caridiopulmonary bypass (CPB) after 10 min on bypass and after 1, 2, 5, and 6 h of perfusion for hemolysis determination and biochemical tests. Values were recorded for blood pressure, mean flow, and pump rotational speed. Analysis of variance was used for repeated measurements. Results: Mean pump flows achieved during the three 6 h pump runs for the three pump heads studied were as follows: (a) flow range 0.0 to 5.2 L/min, 3.6 +/- 1.5 L/min, (b) flow range 0.0 to 7.1 L/min, 4.9 +/- 1.3 L/min, and (c) flow range 0.0 to 6.0 L/min, 3.8 +/- 1.5 L/min. Blood trauma, evaluated by plasma hemoglobin and lactate dehydrogenase levels, did not help in detecting any significant hemolysis. Thrombocytes and white blood cell count profiles showed no significant differences between the groups at the end of the 6 h perfusion. At the end of testing, no clot deposition was found in the oxygenator, and there was no evidence of peripheral emboli. Conclusion: The results suggest that the newly developed magnetically suspended centrifugal pump head provides satisfactory hydrodynamic performance in an acute perfusion scenario without increasing hemolysis.
引用
收藏
页码:183 / 189
页数:7
相关论文
共 50 条
  • [31] Hemocompatibility of new magnetically-levitated centrifugal pump technology compared to the CentriMag adult pump
    David Schibilsky
    Setsuo Takatani
    Barbara Schibilsky
    Tobias Graf
    Diana Michels da Silva
    Hans Peter Wendel
    Meltem Avci-Adali
    Christian Schlensak
    Scientific Reports, 10
  • [32] Hemocompatibility of new magnetically-levitated centrifugal pump technology compared to the CentriMag adult pump
    Schibilsky, David
    Takatani, Setsuo
    Schibilsky, Barbara
    Graf, Tobias
    da Silva, Diana Michels
    Wendel, Hans Peter
    Avci-Adali, Meltem
    Schlensak, Christian
    SCIENTIFIC REPORTS, 2020, 10 (01)
  • [33] Evaluation of a new magnetically suspended centrifugal neonatal pump in healthy animals using a veno-venous extracorporeal membrane oxygenation configuration
    Di Nardo, Matteo
    Moreau, Anthony
    Annoni, Filippo
    Su, Fuhong
    Belliato, Mirko
    Broman, Lars Mikael
    Malfertheiner, Maximilian
    Lorusso, Roberto
    Taccone, Fabio Silvio
    PERFUSION-UK, 2024, 39 (07): : 1462 - 1470
  • [34] Effect of speed on the performance characteristics of a centrifugal slurry pump
    Gandhi, BK
    Singh, SN
    Seshadri, V
    JOURNAL OF HYDRAULIC ENGINEERING-ASCE, 2002, 128 (02): : 225 - 233
  • [36] Some performance characteristics of a centrifugal pump with recessed impeller
    LUBIENIECKI VM
    1972,
  • [37] Development of Terumo implantable left ventricular assist system (T-ILVAS) with a magnetically suspended centrifugal pump
    Nojiri C.
    Kijima T.
    Maekawa J.
    Horiuchi K.
    Kido T.
    Sugiyama T.
    Mori T.
    Sugiura N.
    Asada T.
    Ozaki T.
    Suzuki M.
    Akamatsu T.
    Akutsu T.
    Journal of Artificial Organs, 1999, 2 (1) : 3 - 7
  • [38] Optimal synchronous vibration control for magnetically suspended centrifugal compressor
    Peng, Cong
    He, Jiaxi
    Zhu, Mengting
    Deng, Zhiquan
    Zhen, Ziyang
    Liu, Qiang
    MECHANICAL SYSTEMS AND SIGNAL PROCESSING, 2019, 132 (776-789) : 776 - 789
  • [39] Performance of a new implantable cardiac assist centrifugal pump
    Tevaearai, HT
    Mueller, XM
    Jegger, D
    Augsburger, M
    Burki, M
    von Segesser, LK
    ARTIFICIAL ORGANS, 2001, 25 (01) : 67 - 69
  • [40] Performance characteristics of centrifugal pump using central composite design
    Seenivasan, M.
    Venugopal, T.
    Gopalsamy, N.
    Ulaganathan, K.
    Vineeth, S.
    MATERIALS TODAY-PROCEEDINGS, 2022, 62 : 2105 - 2108