Investigation of dose homogeneity for loose helical tomotherapy delivery in the context of breath-hold radiation therapy

被引:12
|
作者
Kim, B
Kron, T
Battista, J
Van Dyk, J
机构
[1] London Reg Canc Program, London, ON N6A 4L6, Canada
[2] Univ Western Ontario, Dept Med Biophys, London, ON N6A 5C1, Canada
[3] Univ Western Ontario, Dept Oncol, London, ON N6A 4L6, Canada
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2005年 / 50卷 / 10期
关键词
D O I
10.1088/0031-9155/50/10/014
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Loose helical delivery is a potential solution to account for respiration-driven tumour motion in helical tomotherapy (HT). In this approach, a treatment is divided into a set of interlaced 'loose' helices commencing at different gantry angles. Each loose helix covers the entire target length in one gantry rotation during a single breath-hold. The dosimetric characteristics of loose helical delivery were investigated by delivering a 6 MV photon beam in a HT-like manner. Multiple scenarios of conventional 'tight' HT and loose helical deliveries were modelled in treatment planning software, and carried out experimentally with Kodak EDR2 film. The advantage of loose helical delivery lies in its ability to produce a more homogeneous dose distribution by eliminating the 'thread' effect-an inherent characteristic of HT, which results in dose modulations away from the axis of gantry rotation. However, loose helical delivery was also subjected to undesirable dose modulations in the direction of couch motion (termed 'beating' effect), when the ratio between the number of beam projections per gantry rotation (n) and pitch factor (p) was a non-integer. The magnitude of dose modulations decreased with an increasing n/p ratio. The results suggest that for the current HT unit (n = 51), dose modulations could be kept under 5% by selecting a pitch factor smaller than 7. A pitch factor of this magnitude should be able to treat a target up to 30 cm in length. Loose helical delivery should increase the total session time only by a factor of 2, while the planning time should stay the same since the total number of beam projections remains unchanged. Considering its dosimetric advantage and clinical practicality, loose helical delivery is a promising solution for the future HT treatments of respiration-driven targets.
引用
收藏
页码:2387 / 2404
页数:18
相关论文
共 50 条
  • [21] Deep Inspiration Breath-hold (DIBH) Technique to Reduce Cardiac Radiation Dose in the Management of Breast Cancer
    Sayan, Mutlay
    Hopkins, William E.
    Heimann, Ruth
    CURRENT CANCER THERAPY REVIEWS, 2016, 12 (01) : 2 - 14
  • [22] Deep-Inspiration Breath-Hold Stereotactic Body Radiation Therapy by Combining Spirometer-Guided Breath-Hold and a Real-Time Tumor Tracking System: A Novel Approach
    Tanaka, H.
    Ueda, K.
    Karita, M.
    Ono, T.
    Kajima, M.
    Manabe, Y.
    Sera, T.
    Fujimoto, K.
    Yuasa, Y.
    Shiinoki, T.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2023, 117 (02): : E63 - E64
  • [23] Dosimetric evaluation of modern radiation therapy techniques for left breast in deep-inspiration breath-hold
    Koivumakia, Tuomas
    Fogliata, Antonella
    Zeverino, Michele
    Boman, Eeva
    Sierpowska, Joanna
    Moeckli, Raphael
    Vallet, Veronique
    Cozzi, Luca
    Heikkila, Janne
    Vaananen, Anssi
    Vuolukka, Kristiina
    Sillanmaki, Saara
    Seppala, Jan
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 45 : 82 - 87
  • [24] Static Breath-Hold MRI Based Measurement of Change in Pulmonary Function Following a Course of Radiation Therapy
    Ding, K.
    Cao, K.
    Du, K.
    Chen, Q.
    Ennis, D.
    Christensen, G.
    Reinhardt, J.
    Libby, B.
    Benedict, S.
    Sheng, K.
    MEDICAL PHYSICS, 2012, 39 (06) : 3697 - 3697
  • [25] Verification of Breath-Hold Technique in Stereotactic Body Radiation Therapy for Lung Cancer: What Are the Optimal Margins?
    Okazaki, D.
    Murai, T.
    Nakajima, K.
    Kondo, T.
    Yanagi, T.
    Fukuma, H.
    Eguchi, Y.
    Kitagawa, H.
    Tuchiya, T.
    Hirose, Y.
    Shibamoto, Y.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E706 - E707
  • [26] Fraction-specific post-treatment quality assurance for active breath-hold radiation therapy
    Thiyagarajan, Rajesh
    Nambiraj, Arunai
    Manigandan, Durai
    Singaravelu, Tamilseivan
    Selvaraj, Rajesh
    Kataria, Tejinder
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2019, 18 (03) : 262 - 270
  • [27] Predictors of cardiac dose sparing in deep inspiratory breath-hold technique for radiation of left breast cancer patients
    Nagpal, Puneet
    Pruthi, Deep Shankar
    Shanmugam, Prabakar
    Pandey, Manish Bhushan
    Singh, Harpreet
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2023, 19 (03) : 573 - 578
  • [28] Variation in Boost Dose to Breast Cancer Tumor Bed in Deep Inspiration Breath-Hold Radiation Therapy Evaluated With Feedback-Guided CBCT
    Shen, S.
    Benhabib, S. M.
    Duan, J.
    Keene, K.
    Delossantos, J.
    Meredith, R. F.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S251 - S252
  • [29] Deep Inspiration Breath-Hold Radiation Therapy for Locally Advanced Non-Small Cell Lung Cancer: Is Dose to the Mediastinal Structures Reduced?
    Persson, G. F.
    Josipovic, M.
    Maraldo, M. V.
    Nygard, L.
    Specht, L.
    Aznar, M. C.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S618 - S618
  • [30] Clinical experience using respiratory gated radiation therapy: Comparison of free-breathing and breath-hold techniques
    Berson, AM
    Emery, R
    Rodriguez, L
    Richards, GM
    Ng, T
    Sanghavi, S
    Barsa, J
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (02): : 419 - 426