Neonatal and Adult ICU Ventilators to Provide Ventilation in Neonates, Infants, and Children: A Bench Model Study

被引:25
|
作者
Vignaux, Laurence [1 ,2 ,3 ]
Piquilloud, Lise [4 ]
Tourneux, Pierre [2 ,5 ]
Jolliet, Philippe [4 ]
Rimensberger, Peter C. [6 ]
机构
[1] Hop La Tour, Cardioresp Physiotherapy Unit, CH-1217 Meyrin, Switzerland
[2] Univ Picardie Jules Verne, Inst Natl Environm Ind & Risques, EA 4284, Peritox,Unite Mixte 101, Amiens, France
[3] Univ Hosp Geneva, Adult Intens Care Unit, Geneva, Switzerland
[4] Univ Lausanne Hosp, Adult Intens Care & Burn Unit, Lausanne, Switzerland
[5] Univ Hosp North, Pediat Intens Care Unit, Amiens, France
[6] Univ Hosp Geneva, Neonatal & Pediat Intens Care Unit, Geneva, Switzerland
关键词
mechanical ventilators; ventilatory support; neonatal; pediatrics; intensive care units; equipment safety; respiration; artificial; models; PRESSURE-SUPPORT VENTILATION; ACUTE RESPIRATORY-FAILURE; NONINVASIVE VENTILATION; MANDATORY VENTILATION; TIDAL VOLUME; ASSIST; ASYNCHRONY; PERFORMANCE; NAVA; FLOW;
D O I
10.4187/respcare.02540
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Using a bench test model, we investigated the hypothesis that neonatal and/or adult ventilators equipped with neonatal/pediatric modes currently do not reliably administer pressure support (PS) in neonatal or pediatric patient groups in either the absence or presence of air leaks. METHODS: PS was evaluated in 4 neonatal and 6 adult ventilators using a bench model to evaluate triggering, pressurization, and cycling in both the absence and presence of leaks. Delivered tidal volumes were also assessed. Three patients were simulated: a preterm infant (resistance 100 cm H2O/L/s, compliance 2 mL/cm H2O, inspiratory time of the patient [T-1] 400 ms, inspiratory effort 1 and 2 cm H2O), a full-term infant (resistance 50 cm H2O/L/s, compliance 5 mL/cm H2O, T-1 500 ms, inspiratory effort 2 and 4 cm H2O), and a child (resistance 30 cm H2O/L/s, compliance 10 mL/cm H2O, T-1 600 ms, inspiratory effort 5 and 10 cm H2O). Two PS levels were tested (10 and 15 cm H2O) with and without leaks and with and without the leak compensation algorithm activated. RESULTS: Without leaks, only 2 neonatal ventilators and one adult ventilator had trigger delays under a given predefined acceptable limit (1/8 T-1). Pressurization showed high variability between ventilators. Most ventilators showed T-1 in excess high enough to seriously impair patient-ventilator synchronization (> 50% of the T-1 of the subject). In some ventilators, leaks led to autotriggering and impairment of ventilation performance, but the influence of leaks was generally lower in neonatal ventilators. When a noninvasive ventilation algorithm was available, this was partially corrected. In general, tidal volume was calculated too low by the ventilators in the presence of leaks; the noninvasive ventilation algorithm was able to correct this difference in only 2 adult ventilators. CONCLUSIONS: No ventilator performed equally well under all tested conditions for all explored parameters. However, neonatal ventilators tended to perform better in the presence of leaks. These findings emphasize the need to improve algorithms for assisted ventilation modes to better deal with situations of high airway resistance, low pulmonary compliance, and the presence of leaks.
引用
收藏
页码:1463 / 1475
页数:13
相关论文
共 50 条
  • [31] Bilevel vs ICU Ventilators Providing Noninvasive Ventilation: Effect of System Leaks A COPD Lung Model Comparison
    Ferreira, Juliana C.
    Chipman, Daniel W.
    Hill, Nicholas S.
    Kacmarek, Robert M.
    CHEST, 2009, 136 (02) : 448 - 456
  • [32] Unexpected persistence of platelet hyporeactivity beyond the neonatal period:: a flow cytometric study in neonates, infants and older children
    Hézard, N
    Potron, G
    Schlegel, N
    Amory, C
    Leroux, B
    Nguyen, P
    THROMBOSIS AND HAEMOSTASIS, 2003, 90 (01) : 116 - 123
  • [33] Accuracy of tidal volume delivery by paediatric intensive care ventilators: A bench-model study
    Vedrenne-Cloquet, Meryl
    Tuffet, Samuel
    Louis, Bruno
    Khirani, Sonia
    Collignon, Charlotte
    Renolleau, Sylvain
    Fauroux, Brigitte
    Carteaux, Guillaume
    AUSTRALIAN CRITICAL CARE, 2024, 37 (04) : 600 - 605
  • [34] A study of breathing pattern and ventilation in newborn infants and adult subjects
    Al-Hathlol, K
    Idiong, N
    Hussain, A
    Kwiatkowski, K
    Alvaro, RE
    Weintraub, Z
    Cates, DB
    Rigatto, H
    ACTA PAEDIATRICA, 2000, 89 (12) : 1420 - 1425
  • [35] INFUSION OF ATRACURIUM IN NEONATES, INFANTS AND CHILDREN - A STUDY OF DOSE REQUIREMENTS
    KALLI, I
    MERETOJA, OA
    BRITISH JOURNAL OF ANAESTHESIA, 1988, 60 (06) : 651 - 654
  • [36] Effect of Jet Nebulization on Noninvasive Positive-Pressure Ventilation Administered with Noninvasive or Intensive Care Unit Ventilators: A Bench Study
    Xu, Meng-Jiao
    Dai, Bing
    Peng, Yun
    Su, Jia
    Tan, Wei
    Zhao, Hong-Wen
    RESPIRATION, 2019, 97 (04) : 355 - 362
  • [37] Correction to the paper 'Delivery of tidal volume from four anaesthesia ventilators during volume-controlled ventilation: a bench study'
    Hallback, M.
    Wallin, M. K. E. B.
    BRITISH JOURNAL OF ANAESTHESIA, 2016, 116 (05) : 720 - 721
  • [38] Effects on mechanical power of different devices used for inhaled sedation in a bench model of protective ventilation in ICU
    Pellet, Pierre-Louis
    Stevic, Neven
    Degivry, Florian
    Louis, Bruno
    Argaud, Laurent
    Guerin, Claude
    Cour, Martin
    ANNALS OF INTENSIVE CARE, 2024, 14 (01)
  • [39] Effects on mechanical power of different devices used for inhaled sedation in a bench model of protective ventilation in ICU
    Pierre-Louis Pellet
    Neven Stevic
    Florian Degivry
    Bruno Louis
    Laurent Argaud
    Claude Guérin
    Martin Cour
    Annals of Intensive Care, 14
  • [40] A model for size and age changes in the pharmacokinetics of paracetamol in neonates, infants and children
    Anderson, BJ
    Woollard, GA
    Holford, NHG
    BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 2000, 50 (02) : 125 - 134