Neonatal and Adult ICU Ventilators to Provide Ventilation in Neonates, Infants, and Children: A Bench Model Study

被引:25
|
作者
Vignaux, Laurence [1 ,2 ,3 ]
Piquilloud, Lise [4 ]
Tourneux, Pierre [2 ,5 ]
Jolliet, Philippe [4 ]
Rimensberger, Peter C. [6 ]
机构
[1] Hop La Tour, Cardioresp Physiotherapy Unit, CH-1217 Meyrin, Switzerland
[2] Univ Picardie Jules Verne, Inst Natl Environm Ind & Risques, EA 4284, Peritox,Unite Mixte 101, Amiens, France
[3] Univ Hosp Geneva, Adult Intens Care Unit, Geneva, Switzerland
[4] Univ Lausanne Hosp, Adult Intens Care & Burn Unit, Lausanne, Switzerland
[5] Univ Hosp North, Pediat Intens Care Unit, Amiens, France
[6] Univ Hosp Geneva, Neonatal & Pediat Intens Care Unit, Geneva, Switzerland
关键词
mechanical ventilators; ventilatory support; neonatal; pediatrics; intensive care units; equipment safety; respiration; artificial; models; PRESSURE-SUPPORT VENTILATION; ACUTE RESPIRATORY-FAILURE; NONINVASIVE VENTILATION; MANDATORY VENTILATION; TIDAL VOLUME; ASSIST; ASYNCHRONY; PERFORMANCE; NAVA; FLOW;
D O I
10.4187/respcare.02540
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Using a bench test model, we investigated the hypothesis that neonatal and/or adult ventilators equipped with neonatal/pediatric modes currently do not reliably administer pressure support (PS) in neonatal or pediatric patient groups in either the absence or presence of air leaks. METHODS: PS was evaluated in 4 neonatal and 6 adult ventilators using a bench model to evaluate triggering, pressurization, and cycling in both the absence and presence of leaks. Delivered tidal volumes were also assessed. Three patients were simulated: a preterm infant (resistance 100 cm H2O/L/s, compliance 2 mL/cm H2O, inspiratory time of the patient [T-1] 400 ms, inspiratory effort 1 and 2 cm H2O), a full-term infant (resistance 50 cm H2O/L/s, compliance 5 mL/cm H2O, T-1 500 ms, inspiratory effort 2 and 4 cm H2O), and a child (resistance 30 cm H2O/L/s, compliance 10 mL/cm H2O, T-1 600 ms, inspiratory effort 5 and 10 cm H2O). Two PS levels were tested (10 and 15 cm H2O) with and without leaks and with and without the leak compensation algorithm activated. RESULTS: Without leaks, only 2 neonatal ventilators and one adult ventilator had trigger delays under a given predefined acceptable limit (1/8 T-1). Pressurization showed high variability between ventilators. Most ventilators showed T-1 in excess high enough to seriously impair patient-ventilator synchronization (> 50% of the T-1 of the subject). In some ventilators, leaks led to autotriggering and impairment of ventilation performance, but the influence of leaks was generally lower in neonatal ventilators. When a noninvasive ventilation algorithm was available, this was partially corrected. In general, tidal volume was calculated too low by the ventilators in the presence of leaks; the noninvasive ventilation algorithm was able to correct this difference in only 2 adult ventilators. CONCLUSIONS: No ventilator performed equally well under all tested conditions for all explored parameters. However, neonatal ventilators tended to perform better in the presence of leaks. These findings emphasize the need to improve algorithms for assisted ventilation modes to better deal with situations of high airway resistance, low pulmonary compliance, and the presence of leaks.
引用
收藏
页码:1463 / 1475
页数:13
相关论文
共 50 条
  • [21] Monitoring of Noninvasive Ventilation by Built-in Software of Home Bilevel Ventilators A Bench Study
    Contal, Olivier
    Vignaux, Laurence
    Combescure, Christophe
    Pepin, Jean-Louis
    Jolliet, Philippe
    Janssens, Jean-Paul
    CHEST, 2012, 141 (02) : 469 - 476
  • [22] Performance of noninvasive ventilation algorithms on ICU ventilators during pressure support: a clinical study
    Vignaux, Laurence
    Tassaux, Didier
    Carteaux, Guillaume
    Roeseler, Jean
    Piquilloud, Lise
    Brochard, Laurent
    Jolliet, Philippe
    INTENSIVE CARE MEDICINE, 2010, 36 (12) : 2053 - 2059
  • [23] Performance of noninvasive ventilation algorithms on ICU ventilators during pressure support: a clinical study
    Laurence Vignaux
    Didier Tassaux
    Guillaume Carteaux
    Jean Roeseler
    Lise Piquilloud
    Laurent Brochard
    Philippe Jolliet
    Intensive Care Medicine, 2010, 36 : 2053 - 2059
  • [24] Comparative bench study of triggering, pressurization, and cycling between the home ventilator VPAP II and three ICU ventilators
    Didier Tassaux
    Susannah Strasser
    Silvana Fonseca
    Eric Dalmas
    Philippe Jolliet
    Intensive Care Medicine, 2002, 28 : 1254 - 1261
  • [25] Assessment of open-source, intermediate and ICU ventilators to face the COVID-19 pandemic A bench study
    Guerin, Claude
    Cour, Martin
    Degivry, Florian
    Charbon, Francois
    Louis, Bruno
    Argaud, Laurent
    Terzi, Nicolas
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2022, 39 (05) : 474 - 476
  • [26] Delivery of tidal volume from four anaesthesia ventilators during volume-controlled ventilation: a bench study
    Wallon, G.
    Bonnet, A.
    Guerin, C.
    BRITISH JOURNAL OF ANAESTHESIA, 2013, 110 (06) : 1045 - 1051
  • [27] Comparative bench study of triggering pressurization, and cycling between the home ventilator VPAP II and three ICU ventilators
    Tassaux, D
    Strasser, S
    Fonseca, S
    Dalmas, E
    Jolliet, P
    INTENSIVE CARE MEDICINE, 2002, 28 (09) : 1254 - 1261
  • [28] DELIVERY OF TIDAL VOLUME FROM FOUR ANAESTHESIA VENTILATORS DURING VOLUME-CONTROLLED VENTILATION: A BENCH STUDY
    Wallon, G.
    Bonnet, A.
    Guerin, C.
    INTENSIVE CARE MEDICINE, 2012, 38 : S266 - S267
  • [29] Comparison of actual tidal volume in neonatal lung model volume control ventilation using three ventilators
    Toyama, H.
    Endo, Y.
    Ejima, Y.
    Matsubara, M.
    Kurosawa, S.
    ANAESTHESIA AND INTENSIVE CARE, 2011, 39 (04) : 599 - 606
  • [30] PERFORMANCE OF AN ICU VENTILATOR AND TWO TURBIN-BASED VENTILATORS DEDICATED TO NON INVASIVE VENTILATION (NIV) IN SIMULATED HIGH INSPIRATORY EFFORT AND RATE: A NIV BENCH-STUDY
    Piquilloud, L.
    Reichmuth, P.
    Thevoz, D.
    Jolliet, P.
    Revelly, J. -P.
    INTENSIVE CARE MEDICINE, 2012, 38 : S67 - S67