Microscopic examination of the sonothrombolysis process inside the transparent to ultrasound parallel plate flow chamber

被引:0
|
作者
Secomski, Wojciech [1 ]
Klimonda, Ziemowit [1 ]
Majka, Katarzyna [2 ]
Olszewski, Robert [1 ,3 ]
Nowicki, Andrzej [1 ]
机构
[1] Polish Acad Sci, Dept Ultrasound, Inst Fundamental Technol Res, Warsaw, Poland
[2] Med Univ Warsaw, Clin Traumatol & Orthoped, Mil Inst Med Warsaw, Dept Basic Nursing, Warsaw, Poland
[3] Polish Acad Sci, Dept Gerontol, Publ Hlth & Didact, Natl Inst Geriatr Rheumatol & Rehabil, Warsaw, Poland
关键词
ultrasound; blood; thrombus; thrombolysis; parallel plate flow chamber;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Parallel plate flow chambers are widely used for cell research in conditions of constant or pulsatile liquid flow. They are also used to analyze the formation of thrombi. The authors designed a flow chamber that is transparent to ultrasound, thus enabling the microscopic observation of the thrombus dissolution process by interaction of drug, microbubbles and ultrasound in real time, in-vitro. Sonothrombolysis was performed at 1 MHz and 3.3 MHz ultrasound frequencies, at space-averaged and time-averaged intensities I-SATA of 0.2 - 1.6 W/cm(2).1 mm thick slices of the human blood clots were exposed to ultrasound in the presence of the Actilyse tissue plasminogen activator at a concentration of 10 mu g/ml flowing around the sample. Next, the effect of SonoVue microbubbles at a concentration of 5x10(5)/ml on the dissolution of the thrombus was investigated. Thrombus size changes were observed under a 4x magnification microscope and were recorded as a function of time. The best result was achieved for the Actilyse tissue plasminogen activator at the 1 MHz and 1.6/cm(2) I-SATA, the thrombus was sonolysed within 5.5 minutes. The noticeable effect of the SonoVue microbubbles on the thrombolysis process appeared at 3.3 MHz, the thrombus was sonolysed in 12 min at I-SATA = 1.6 W/cm(2).
引用
收藏
页数:4
相关论文
共 50 条
  • [1] Drug Adhesion Parallel Plate Flow Chamber
    Han, G.
    Hardiman, K.
    Hyer, R.
    2012 38TH ANNUAL NORTHEAST BIOENGINEERING CONFERENCE (NEBEC), 2012, : 145 - 146
  • [2] Numerical optimum design for parallel plate flow chamber
    Jiang, Wen-Tao
    Fan, Yu-Bo
    Zou, Yuan-Wen
    Chen, Jun-Kai
    Jisuan Lixue Xuebao/Chinese Journal of Computational Mechanics, 2007, 24 (04): : 420 - 424
  • [3] On analysis of the steady flow in an irrectangular parallel-plate flow chamber
    Kairong Q.
    Weiyuan J.
    Xixi L.
    Zhaorong L.
    Applied Mathematics and Mechanics, 1998, 19 (9) : 851 - 859
  • [4] On analysis of the steady flow in an irrectangular parallel-plate flow chamber
    Qin, KR
    Jiang, WY
    Li, XX
    Liu, ZR
    APPLIED MATHEMATICS AND MECHANICS-ENGLISH EDITION, 1998, 19 (09) : 851 - 859
  • [5] On analysis of the steady flow in an irrectangular parallel-plate flow chamber
    Fudan Univ, Shanghai, China
    Appl Math Mech Engl Ed, 9 (851-859):
  • [6] Monitoring cell detachment by surfactants in a parallel plate flow chamber
    Azeredo, J
    Pacheco, AP
    Lopes, I
    Oliveira, R
    Vieira, MJ
    WATER SCIENCE AND TECHNOLOGY, 2003, 47 (05) : 77 - 82
  • [7] Measurement of microbial adhesive forces with a parallel plate flow chamber
    Yoshihara, Akinori
    Narahara, Hisaya
    Kuriyama, Yuta
    Toyoda, Syunsuke
    Tokumoto, Hayato
    Konishi, Yasuhiro
    Nomura, Toshiyuki
    JOURNAL OF COLLOID AND INTERFACE SCIENCE, 2014, 432 : 77 - 85
  • [8] Quantitative Assessment of Steady and Pulsatile Flow Fields in a Parallel Plate Flow Chamber
    Eric A. Nauman
    Kurtis J. Risic
    Tony M. Keaveny
    Robert L. Satcher
    Annals of Biomedical Engineering, 1999, 27 : 194 - 199
  • [9] Quantitative assessment of steady and pulsatile flow fields in a parallel plate flow chamber
    Nauman, EA
    Risic, KJ
    Keaveny, TM
    Satcher, RL
    ANNALS OF BIOMEDICAL ENGINEERING, 1999, 27 (02) : 194 - 199
  • [10] Quantitative Assessment of Steady and Pulsatile Flow Fields in a Parallel Plate Flow Chamber
    Nauman, Eric A.
    Risic, Kurtis J.
    Keaveny, Tony M.
    Satcher, Robert L.
    Annals of Biomedical Engineering, 27 (02): : 194 - 199