Commissioning and beam characterization of the first gantry-mounted accelerator pencil beam scanning proton system

被引:41
|
作者
Kang, M. [1 ]
Cessac, Rob [2 ]
Pang, D. [1 ]
机构
[1] Georgetown Univ Hosp, Dept Radiat Med, Washington, DC 20007 USA
[2] Mevion Med Syst Inc, Littleton, MA USA
关键词
adaptive aperture; beam commissioning; compart PBS system; gantry-mounted accelerator; HYPERSCAN; pencil beam scanning; MONTE-CARLO; THERAPY; INTERPLAY; LUNG; DELIVERY; CANCER; MOTION; PATIENT; IMPACT; TUMOR;
D O I
10.1002/mp.13972
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To present and discuss beam characteristics and commissioning process of the first gantry-mounted accelerator single-room pencil beam scanning (PBS) proton system. Methods: The Mevion HYPERSCAN employs a design configuration with a synchrocyclotron mounted on the gantry to eliminate the traditional beamline and a nozzle that contains the dosimetry monitoring chambers, the energy modulator [energy selector (ES)], and an adaptive aperture (AA). To characterize the beam, we measured the integrated depth dose (IDD) for 12 energies. from the highest energy of 227 MeV down to 28 MeV with a range difference similar to 2 cm between the adjacent energies, using a large radius Bragg peak chamber; single spot profiles in air at five locations along the beam central axis using radiochromic EBT3 film and cross compared with a scintillation detector; and determined the output using a densely packed spot map. To access the performance of AA, we measured interleaf leakage and the penumbra reduction effect. Monte Carlo simulation using TOPAS was performed to study spot size variation along the beam path, beam divergence, and energy spectrum. Results: This proton system is calibrated to deliver 1 Gy dose at 5 cm depth in water using the highest beam energy by delivering 1 MU/spot to a 10 x 10 cm(2) map with a 2.5 mm spot spacing. The spot size in air at isocenter for a maximum range beam of 227 MeV is 4.1 mm. This system is able to reduce the beam range all the way to the patient surface; the lowest energy beam measured was 28 MeV which has a spot size of 15.7 mm. The beam divergence is 2.4 mrad at 227 MeV and 52.7 mrad for the superficial 28 MeV beam. The binary design of the ES has resulted in shifts of the effective SSD toward the isocenter as the energy is modulated lower. The pristine Bragg peaks have a constant 80%-80% width of 8.4 mm at all energies. The interleaf leakage of the AA is less than 1.5% at the highest energy. The AA reduces field penumbras. For a 10x10 optimized field, the 227 MeV beam penumbra measured at isocenter with a 5 cm air gap went from 6.8 to 4.7 mm and the 28 MeV beam penumbra went from 21.4 to 7.5 mm. Conclusions: The HYPERSCAN proton system has a unique design, which is reflected in the Bragg peak shapes, the variation of spot sizes with energy and the penumbra sharpening effect of the AA. The combination of the ES and AA makes PBS implementation possible without using a beam transport line and supplemental range shifter devices. In commissioning the TPS and designing plans these differences need to be considered. (C) 2020 American Association of Physicists in Medicine
引用
收藏
页码:3496 / 3510
页数:15
相关论文
共 50 条
  • [1] An Overview of the Commissioning of a Two-Gantry Proton Pencil Beam Scanning System
    Zhang, Y.
    Lee, E.
    Xiao, Z.
    Lamba, M.
    Mascia, A.
    MEDICAL PHYSICS, 2017, 44 (06) : 2870 - 2870
  • [2] Commissioning and initial experience with the first clinical gantry-mounted proton therapy system
    Zhao, Tianyu
    Sun, Baozhou
    Grantham, Kevin
    Rankine, Leith
    Cai, Bin
    Goddu, Sreekrishna M.
    Santanam, Lakshmi
    Knutson, Nels
    Zhang, Tiezhi
    Reilly, Michael
    Bottani, Beth
    Bradley, Jeffrey
    Mutic, Sasa
    Klein, Eric E.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02): : 24 - 40
  • [3] High spatiotemporal resolution scintillation imaging of pulsed pencil beam scanning proton beams produced by a gantry-mounted synchrocyclotron
    Goddu, S. Murty
    Hao, Yao
    Ji, Zhen
    Setianegara, Jufri
    Liu, Fengwei
    Green, Winter
    Sobotka, Lee G.
    Zhao, Tianyu
    Perkins, Stephanie
    Darafsheh, Arash
    MEDICAL PHYSICS, 2024, 51 (07) : 4996 - 5006
  • [4] Image quality characterisation of a proton gantry-mounted cone-beam CT system
    Andersson, K.
    Dahlgren, C. Vallhagen
    Dasu, A.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S556 - S557
  • [5] Commissioning the beam model of apertures in a compact proton system using pencil beam scanning
    Perez Moreno, J. M.
    Vera Sanchez, J. A.
    Cerron Campoo, F.
    Castro Novais, J.
    Canals de las Casas, E.
    Lorenzo Villanueva, I.
    Sallabanda Hajro, M.
    Montero Feijoo, M.
    de Pablo Gonzalez, A.
    Matute Martin, R.
    Miralbell, R.
    Mazal, A.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S1750 - S1751
  • [6] Commissioning of a Two-Gantry Proton Pencil Beam Scanning System, the First Multi-Room Varian ProBeam360°
    Lee, E.
    Faught, A. M.
    Lee-Poprocki, H.
    Batin, E.
    Ayan, A. S.
    MEDICAL PHYSICS, 2024, 51 (10) : 7722 - 7723
  • [7] Commissioning of the world's first compact pencil-beam scanning proton therapy system
    Pidikiti, Rajesh
    Patel, Bijal C.
    Maynard, Matthew R.
    Dugas, Joseph P.
    Syh, Joseph
    Sahoo, Narayan
    Wu, Hsinshun Terry
    Rosen, Lane R.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (01): : 94 - 105
  • [8] Siddarth-II Superia: The First Indian Made Ring Gantry-Mounted IGRT Capable VMAT Delivery System Commissioning and Beam Characterization.
    Muralidhar, K. Raja
    Ramachandran, V.
    Subrahmanyam, G.
    MEDICAL PHYSICS, 2024, 51 (10) : 7968 - 7969
  • [9] Monte Carlo model of a gantry for proton pencil beam scanning
    Winterhalter, Carla
    Safai, Sairos
    Grassberger, Clemens
    Oxley, David
    Weber, Damien C.
    Lomax, Tony
    STRAHLENTHERAPIE UND ONKOLOGIE, 2016, 192 (11) : 837 - 837
  • [10] Feasibility of SOBP Proton FLASH Beam Delivery Using a Gantry-Mounted Clinical Synchrocyclotron
    Darafsheh, A.
    Hao, Y.
    Zhao, X.
    Zwart, T.
    Wagner, M.
    Evans, T.
    Reynoso, F.
    Zhao, T.
    MEDICAL PHYSICS, 2021, 48 (06)