High spatiotemporal resolution scintillation imaging of pulsed pencil beam scanning proton beams produced by a gantry-mounted synchrocyclotron

被引:1
|
作者
Goddu, S. Murty [1 ]
Hao, Yao [1 ]
Ji, Zhen [1 ]
Setianegara, Jufri [1 ]
Liu, Fengwei [1 ]
Green, Winter [1 ]
Sobotka, Lee G. [2 ]
Zhao, Tianyu [1 ]
Perkins, Stephanie [1 ]
Darafsheh, Arash [1 ]
机构
[1] Washington Univ, Sch Med St Louis, Dept Radiat Oncol, St Louis, MO 63110 USA
[2] Washington Univ St Louis, Dept Chem, St Louis, MO USA
关键词
imaging; proton therapy; quality assurance; scintillation; synchrocyclotron; PLASTIC SCINTILLATOR; QUENCHING CORRECTION; QUALITY-ASSURANCE; MAGIC CUBE; DOSIMETRY; DETECTORS; THERAPY; IRRADIATION; WATER; TOOL;
D O I
10.1002/mp.17116
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundA dosimeter with high spatial and temporal resolution would be of significant interest for pencil beam scanning (PBS) proton beams' characterization, especially when facing small fields and beams with high temporal dynamics. Optical imaging of scintillators has potential in providing sub-millimeter spatial resolution with pulse-by-pulse basis temporal resolution when the imaging system is capable of operating in synchrony with the beam-producing accelerator.PurposeWe demonstrate the feasibility of imaging PBS proton beams as they pass through a plastic scintillator detector to simultaneously obtain multiple beam parameters, including proton range, pencil beam's widths at different depths, spot's size, and spot's position on a pulse-by-pulse basis with sub-millimeter resolution.Materials and methodsA PBS synchrocyclotron was used for proton irradiation. A BC-408 plastic scintillator block with 30 x 30 x 5 cm3 size, and another block with 30 x 30 x 0.5 cm3 size, positioned in an optically sealed housing, were used sequentially to measure the proton range, and spot size/location, respectively. A high-speed complementary metal-oxide-semiconductor (CMOS) camera system synchronized with the accelerator's pulses through a gating module was used for imaging. Scintillation images, captured with the camera directly facing the 5-cm-thick scintillator, were corrected for background (BG), and ionization quenching of the scintillator to obtain the proton range. Spots' position and size were obtained from scintillation images of the 0.5-cm-thick scintillator when a 45 degrees mirror was used to reflect the scintillation light toward the camera.ResultsScintillation images with 0.16 mm/pixel resolution corresponding to all proton pulses were captured. Pulse-by-pulse analysis showed that variations of the range, spots' position, and size were within +/- 0.2% standard deviation of their average values. The absolute ranges were within +/- 1 mm of their expected values. The average spot-positions were mostly within +/- 0.8 mm and spots' sigma agreed within 0.2 mm of the expected values.ConclusionScintillation-imaging PBS beams with high-spatiotemporal resolution is feasible and may help in efficient and cost-effective acceptance testing and commissioning of existing and even emerging technologies such as FLASH, grid, mini-beams, and so forth.
引用
收藏
页码:4996 / 5006
页数:11
相关论文
共 17 条
  • [1] Synchronized high-speed scintillation imaging of proton beams, generated by a gantry-mounted synchrocyclotron, on a pulse-by-pulse basis
    Goddu, Sreekrishna Murty
    Westphal, Gregory T.
    Sun, Baozhou
    Wu, Yu
    Bloch, Charles D.
    Bradley, Jeffrey D.
    Darafsheh, Arash
    MEDICAL PHYSICS, 2022, 49 (09) : 6209 - 6220
  • [2] Pulse-By-Pulse Ultra-High Resolution Scintillation Imaging Dosimetry for Validation of Ultra-High Dose Rate (FLASH) Proton Beams Produced By a Gantry-Mounted Synchrocyclotron
    Goddu, S. M.
    Hao, Y.
    Hollingsworth, S.
    Ji, Z.
    Green, W.
    Owen, D. R.
    Sobotka, L. G.
    Zhang, T.
    Zhao, T.
    Perkins, S.
    Darafsheh, A.
    MEDICAL PHYSICS, 2024, 51 (10) : 7661 - 7661
  • [3] Commissioning and beam characterization of the first gantry-mounted accelerator pencil beam scanning proton system
    Kang, M.
    Cessac, Rob
    Pang, D.
    MEDICAL PHYSICS, 2020, 47 (08) : 3496 - 3510
  • [4] Feasibility of SOBP Proton FLASH Beam Delivery Using a Gantry-Mounted Clinical Synchrocyclotron
    Darafsheh, A.
    Hao, Y.
    Zhao, X.
    Zwart, T.
    Wagner, M.
    Evans, T.
    Reynoso, F.
    Zhao, T.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [5] Ultra-fast, high spatial resolution single-pulse scintillation imaging of synchrocyclotron pencil beam scanning proton delivery
    Clark, Megan
    Ding, Xuanfeng
    Zhao, Lewei
    Pogue, Brian
    Gladstone, David
    Rahman, Mahbubur
    Zhang, Rongxiao
    Bruza, Petr
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (04):
  • [6] Feasibility of Proton Beam Delivery at Ultra-High Dose Rate FLASH Using a Gantry-Mounted Clinical Synchrocyclotron
    Darafsheh, A.
    Hao, Y.
    Zwart, T.
    Wagner, M.
    Catanzano, C.
    Williamson, J.
    Knutson, N.
    Sun, B.
    Mutic, S.
    Zhao, T.
    MEDICAL PHYSICS, 2020, 47 (06) : E461 - E461
  • [7] Performance of a scintillation imaging system for relative dosimetry in pencil beam scanning proton therapy
    Liu, Qi
    Gong, Liangde
    Li, Xiufang
    Grossmann, Martin
    Wang, Jie
    Guo, Mengya
    Gu, Shuaizhe
    Lin, Ye
    Zhang, Manzhou
    Pu, Yuehu
    Chen, Zhiling
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2022, 1040
  • [8] A 2D strip ionization chamber array with high spatiotemporal resolution for proton pencil beam scanning FLASH radiotherapy
    Yang, Yunjie
    Shi, Chengyu
    Chen, Chin-Cheng
    Tsai, Pingfang
    Kang, Minglei
    Huang, Sheng
    Lin, Chih-Hsun
    Chang, Fu-Xiong
    Chhabra, Arpit M.
    Choi, J. Isabelle
    Tome, Wolfgang A.
    Simone, Charles B.
    Lin, Haibo
    MEDICAL PHYSICS, 2022, 49 (08) : 5464 - 5475
  • [9] A High Spatiotemporal Resolution 2D Strip Ionization Chamber Array for Proton Pencil Beam Scanning FLASH Radiotherapy
    Yang, Y.
    Shi, C.
    Chen, C.
    Tsai, P.
    Huang, S.
    Kang, M.
    Lin, C.
    Chang, F.
    Chhabra, A.
    Choi, I.
    Tome, W.
    Simone, C.
    Lin, H.
    MEDICAL PHYSICS, 2022, 49 (06) : E520 - E520
  • [10] Characterization of proton pencil beam scanning and passive beam using a high spatial resolution solid-state microdosimeter
    Tran, Linh T.
    Chartier, Lachlan
    Bolst, David
    Pogossov, Alex
    Guatelli, Susanna
    Petasecca, Marco
    Lerch, Michael L. F.
    Prokopovich, Dale A.
    Reinhard, Mark I.
    Clasie, Benjamin
    Depauw, Nicolas
    Kooy, Hanne
    Flanz, Jacob B.
    McNamara, Aimee
    Paganetti, Harald
    Beltran, Chris
    Furutani, Keith
    Perevertaylo, Vladimir L.
    Jackson, Michael
    Rosenfeld, Anatoly B.
    MEDICAL PHYSICS, 2017, 44 (11) : 6085 - 6095