Safety and efficacy of flow re-direction endoluminal device (FRED) in the treatment of cerebral aneurysms: a single center experience

被引:34
|
作者
Briganti, Francesco [1 ]
Leone, Giuseppe [2 ]
Ugga, Lorenzo [2 ]
Marseglia, Mariano [1 ]
Solari, Domenico [3 ]
Caranci, Ferdinando [2 ]
Mariniello, Giuseppe [3 ]
Maiuri, Francesco [3 ]
Cappabianca, Paolo [3 ]
机构
[1] Univ Naples Federico II, Dept Adv Biomed Sci, Unit Intervent Neuroradiol, Via S Pansini 5, I-80131 Naples, Italy
[2] Federico II, Dept Adv Biomed Sci, Via S Pansini 5, I-80131 Naples, Italy
[3] Univ Naples Federico II, Div Neurosurg, Dept Neurosci Reprod & Odontostomatol Sci, Via S Pansini 5, I-80131 Naples, Italy
关键词
Intracranial aneurysm; Flow diverter devices; FRED; Endovascular treatment; PIPELINE EMBOLIZATION DEVICE; INTRACRANIAL ANEURYSMS; DIVERTOR DEVICES; FOLLOW-UP; INITIAL-EXPERIENCE; ENDOVASCULAR TREATMENT; SILK STENT; METAANALYSIS; MULTICENTER; DIVERSION;
D O I
10.1007/s00701-016-2875-4
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Experience with the endovascular treatment of cerebral aneurysms by the Flow Re-Direction Endoluminal Device (FRED) is still limited. The aim of this study is to discuss the results and complications of this new flow diverter device (FDD). Between November 2013 and April 2015, 20 patients (15 female and five male) harboring 24 cerebral aneurysms were treated with FRED FDD in a single center. Complete occlusion was obtained in 20/24 aneurysms (83 %) and partial occlusion in four (17 %). Intraprocedural technical complication occurred in one case (4 %) and post-procedural complications in three (12 %). None reported neurological deficits (mRS = 0). All FRED were patent at follow-up. No early or delayed aneurysm rupture, no subarachnoid (SAH) or intraparenchymal hemorrhage (IPH) no ischemic complications and no deaths occurred. Endovascular treatment with FRED FDD is a safe treatment for unruptured cerebral aneurysms, resulting in a high rate of occlusion. The FRED is substantially equivalent to the other known FDDs, which show similar functions and technical profiles.
引用
收藏
页码:1745 / 1755
页数:11
相关论文
共 50 条
  • [21] Safety and efficacy results of the Flow Redirection Endoluminal Device (FRED) stent system in the treatment of intracranial aneurysms: US pivotal trial
    McDougall, Cameron G.
    Diaz, Orlando
    Boulos, Alan
    Siddiqui, Adnan H.
    Caplan, Justin
    Fifi, Johanna T.
    Turk, Aquilla S.
    Kayan, Yasha
    Jabbour, Pascal
    Kim, Louis J.
    Hetts, Steven W.
    Cooke, Daniel L.
    Dowd, Christopher F.
    JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2022, 14 (06) : 577 - 584
  • [22] North American Multicenter Experience of the Flow-Redirection Endoluminal Device (FRED) in the Treatment of Intracranial Aneurysms
    Khorasanizadeh, MirHojjat
    Shutran, Max
    Schirmer, Clemens M.
    Salem, Mohamed M.
    Ringer, Andrew J.
    Grandhi, Ramesh
    Mitha, Alim P.
    Levitt, Michael Robert
    Jankowitz, Brian T.
    Taussky, Phil
    Thomas, Ajith J.
    Moore, Justin M.
    Ogilvy, Christopher S.
    NEUROSURGERY, 2023, 69 : 96 - 96
  • [23] Flow redirection endoluminal device (FRED) for treatment of intracranial aneurysms: A systematic review
    Waqas, Muhammad
    Dossani, Rimal H.
    Alkhaldi, Modhi
    Neveu, Jocelyn
    Cappuzzo, Justin M.
    Lim, Jaims
    Khan, Asham
    Lazarov, Victoria
    Monteiro, Andre
    Davies, Jason M.
    Siddiqui, Adnan H.
    Levy, Elad, I
    INTERVENTIONAL NEURORADIOLOGY, 2022, 28 (03) : 347 - 357
  • [24] Flow Redirection Endoluminal Device (FRED) with or without Adjunctive Coiling in Treatment of Very Large and Giant Cerebral Aneurysms
    Sedat Giray Kandemirli
    Feyyaz Baltacioglu
    Jessica Jesser
    Osman Kizilkilic
    Civan Islak
    Markus Möhlenbruch
    Naci Kocer
    Clinical Neuroradiology, 2022, 32 : 471 - 480
  • [25] Flow Redirection Endoluminal Device (FRED) with or without Adjunctive Coiling in Treatment of Very Large and Giant Cerebral Aneurysms
    Kandemirli, Sedat Giray
    Baltacioglu, Feyyaz
    Jesser, Jessica
    Kizilkilic, Osman
    Islak, Civan
    Mohlenbruch, Markus
    Kocer, Naci
    CLINICAL NEURORADIOLOGY, 2022, 32 (02) : 471 - 480
  • [26] Comparison of Pipeline Embolization Device and Flow Re-Direction Endoluminal Device Flow Diverters for Internal Carotid Artery Aneurysms: A Propensity Score-Matched Cohort Study
    Griessenauer, Christoph J.
    Thomas, Ajith J.
    Enriquez-Marulanda, Alejandro
    Deshmukh, Aviraj
    Jain, Abhi
    Ogilvy, Christopher S.
    Kocer, Naci
    Engelhorn, Tobias
    Moehlenbruch, Markus
    Holtmannspotter, Markus
    Janssen, Hendrik
    Finkenzeller, Thomas
    Reith, Wolfgang
    Sonnberger, Michael
    Buhk, Jan-Hendrik
    Schirmer, Clemens M.
    Killer-Oberpfalzer, Monika
    NEUROSURGERY, 2019, 85 (02) : E249 - E255
  • [27] Commentary: Comparison of Pipeline Embolization Device and Flow Re-Direction Endoluminal Device Flow Diverters for Internal Carotid Artery Aneurysms: A Propensity-Score Matched Cohort Study
    Bender, Matthew T.
    Colby, Geoffrey P.
    Coon, Alexander L.
    NEUROSURGERY, 2019, 85 (02) : E256 - E257
  • [28] Feasibility, complications, morbidity, and mortality results at 6 months for aneurysm treatment with the Flow Re-Direction Endoluminal Device: report of SAFE study
    Pierot, Laurent
    Spelle, Laurent
    Berge, Jerome
    Januel, Anne-Christine
    Herbreteau, Denis
    Aggour, Mohamed
    Piotin, Michel
    Biondi, Alessandra
    Barreau, Xavier
    Mounayer, Charbel
    Papagiannaki, Chrisanthi
    Lejeune, Jean-Paul
    Gauvrit, Jean-Yves
    Costalat, Vincent
    JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2018, 10 (08) : 765 - +
  • [29] Difference in the cumulative incidence of aneurysmal occlusion by Flow Re-direction Endoluminal Device and Pipeline Embolization Device in the treatment of unruptured internal carotid artery aneurysms: a propensity score-matched cohort study
    Hidetoshi Matsukawa
    Kazutaka Uchida
    Saujanya Rajbhandari
    Manabu Shirakawa
    Shinichi Yoshimura
    Neurosurgical Review, 46
  • [30] Difference in the cumulative incidence of aneurysmal occlusion by Flow Re-direction Endoluminal Device and Pipeline Embolization Device in the treatment of unruptured internal carotid artery aneurysms: a propensity score-matched cohort study
    Matsukawa, Hidetoshi
    Uchida, Kazutaka
    Rajbhandari, Saujanya
    Shirakawa, Manabu
    Yoshimura, Shinichi
    NEUROSURGICAL REVIEW, 2023, 46 (01)