Dose rate and energy distributions of X rays from a linear accelerator at the maze entrance of a radiotherapy room by measurement and Monte Carlo simulation

被引:7
|
作者
Al-Affan, IAM [1 ]
Smith, CW
Morgan, HM
Lillicrap, SC
机构
[1] Velindre NHS Trust, Cardiff CF4 7XL, S Glam, Wales
[2] Royal United Hosp, Wessex Reg Med Phys Serv, Bath BA1 3NG, Avon, England
关键词
D O I
10.1093/oxfordjournals.rpd.a032360
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The dose rates in a radiotherapy room and maze were measured and computed, using the Monte Carlo code for neutron and photon transport, MCNP, for 6 MV X rays at different field sizes and gantry angles. The contribution from leakage was also measured and computed. When a phantom (patient) was positioned at the isocentre the measured and computed doses at the maze entrance were increased by a factor of 2-3 times over that without a phantom present. The size of the phantom on which the primary beam was incident was not a sensitive factor in increasing the dose rate at the maze entrance. There was good agreement between the measurements and MCNP computations. The energy spectra of the photons (both the scattered and leakage components) near the maze entrance were measured and computed for a primary beam with a field size 20 mm x 20 mm for scatter and leakage.
引用
收藏
页码:273 / 277
页数:5
相关论文
共 50 条
  • [41] Monte Carlo simulation on low-energy electrons from gold nanoparticle in radiotherapy
    Chow, James C. L.
    Leung, Michael K. K.
    Fahey, Sean
    Chithrani, Devika B.
    Jaffray, David A.
    HIGH PERFORMANCE COMPUTING SYMPOSIUM 2011, 2012, 341
  • [42] CALIBRATION OF ENERGY AND ABSORBED DOSE, AND MEASUREMENT OF DOSE DISTRIBUTIONS AND TISSUE-AIR RATIOS FOR AN 8 MV LINEAR ACCELERATOR
    WRIGHT, AE
    BUSHONG, SC
    HUDGINS, PT
    BELLINGE.DL
    PHYSICS IN MEDICINE AND BIOLOGY, 1970, 15 (02): : 366 - &
  • [43] Estimation of absorbed gamma dose rate from granite by Monte Carlo simulation approach
    Knezevic, J.
    Kuzmanovic, P.
    Mrdja, D.
    Todorovic, N.
    Bikit, I
    Hansman, J.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2020, 40 (02) : 596 - 611
  • [44] Modeling 6 MeV Electron Beam From Medical Linear Accelerator Using Monte Carlo Simulation
    Dechsupa, P.
    Tannanonta, C.
    Phaisangittisakul, N.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [45] Axial absorbed dose distributions during abdominal computed tomography acquisitions: Measurement and the Monte Carlo simulation study
    Matsubara, K.
    Nagata, H.
    Okubo, R.
    Ogawa, Y.
    Chusin, T.
    Hirosawa, A.
    18TH ASIA-OCEANIA CONGRESS OF MEDICAL PHYSICS (AOCMP) & 16TH SOUTH-EAST ASIA CONGRESS OF MEDICAL PHYSICS (SEACOMP), 2019, 1248
  • [46] Monte Carlo simulation and measurement of radiation leakage from applicators used in external electron radiotherapy
    Shimozato, Tomohiro
    Okudaira, Kuniyasu
    Fuse, Hiraku
    Tabushi, Katsuyoshi
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (04): : 388 - 396
  • [47] DEPENDENCE BETWEEN THE SIZE OF THE TREATMENT ROOM AND THE FLUENCE OF NEUTRONS UNDESIRABLE IN RADIOTHERAPY FOR THE HIGH-ENERGY THERAPEUTIC X-RAYS GENERATED BY THE LINEAR MEDICAL ACCELERATOR
    Laciak, Marcin
    Konefal, Adam
    ACTA PHYSICA POLONICA B, 2014, 45 (02): : 559 - 564
  • [48] CONVERGING STEREOTACTIC RADIOTHERAPY USING KILOVOLTAGE X-RAYS: EXPERIMENTAL IRRADIATION OF NORMAL RABBIT LUNG AND DOSE-VOLUME ANALYSIS WITH MONTE CARLO SIMULATION
    Kawase, Takatsugu
    Kunieda, Etsuo
    Deloar, Hossain M.
    Tsunoo, Takanori
    Seki, Satoshi
    Oku, Yohei
    Saitoh, Hidetoshi
    Saito, Kimiaki
    Ogawa, Eileen N.
    Ishizaka, Akitoshi
    Kameyama, Kaori
    Kubo, Atsushi
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (02): : 468 - 474
  • [49] Evaluation of the electron energy fluence and angular distributions from a clinical accelerator. A BEAMnrc Monte Carlo study
    Jabbari, N.
    Nedaie, H.
    Zeinali, A.
    IRANIAN JOURNAL OF RADIATION RESEARCH, 2011, 9 (01): : 29 - 36
  • [50] Monte Carlo simulation of stereotactic microbeam radiation therapy: evaluation of the usage of a linear accelerator as the x-ray source
    Gokeri, Gurdal
    Kocar, Cemil
    Tombakoglu, Mehmet
    Cecen, Yigit
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (13): : 4621 - 4642