Dose rate and energy distributions of X rays from a linear accelerator at the maze entrance of a radiotherapy room by measurement and Monte Carlo simulation

被引:7
|
作者
Al-Affan, IAM [1 ]
Smith, CW
Morgan, HM
Lillicrap, SC
机构
[1] Velindre NHS Trust, Cardiff CF4 7XL, S Glam, Wales
[2] Royal United Hosp, Wessex Reg Med Phys Serv, Bath BA1 3NG, Avon, England
关键词
D O I
10.1093/oxfordjournals.rpd.a032360
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The dose rates in a radiotherapy room and maze were measured and computed, using the Monte Carlo code for neutron and photon transport, MCNP, for 6 MV X rays at different field sizes and gantry angles. The contribution from leakage was also measured and computed. When a phantom (patient) was positioned at the isocentre the measured and computed doses at the maze entrance were increased by a factor of 2-3 times over that without a phantom present. The size of the phantom on which the primary beam was incident was not a sensitive factor in increasing the dose rate at the maze entrance. There was good agreement between the measurements and MCNP computations. The energy spectra of the photons (both the scattered and leakage components) near the maze entrance were measured and computed for a primary beam with a field size 20 mm x 20 mm for scatter and leakage.
引用
收藏
页码:273 / 277
页数:5
相关论文
共 50 条
  • [1] Estimation of the dose at the maze entrance for x-rays from radiotherapy linear accelerators
    Al-Affan, IAM
    MEDICAL PHYSICS, 2000, 27 (01) : 231 - 238
  • [2] On a method of dose rate calculation at the maze entrance of medical linear accelerator rooms
    Falcao, RC
    Okuno, E
    RADIATION PROTECTION DOSIMETRY, 1999, 86 (03) : 225 - 227
  • [3] Monte carlo simulation study on the dose and dose-averaged linear energy transfer distributions in carbon ion radiotherapy
    Ishikawa, Akihisa
    Koba, Yusuke
    Furuta, Takuya
    Chang, Weishan
    Yonai, Shunsuke
    Matsumoto, Shinnosuke
    Hashimoto, Shintaro
    Hirai, Yuta
    Sato, Tatsuhiko
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2024, 17 (02) : 553 - 560
  • [4] Monte Carlo simulation of a medical linear accelerator for radiotherapy use
    Serrano, B.
    Hachem, A.
    Franchisseur, E.
    Herault, J.
    Marcie, S.
    Costa, A.
    Bensadoun, R. J.
    Barthe, J.
    Gerard, J. P.
    RADIATION PROTECTION DOSIMETRY, 2006, 119 (1-4) : 506 - 509
  • [5] MONTE CARLO SIMULATION OF ELECTRON DOSE DISTRIBUTIONS PRODUCED BY THE ELEKTA SYNERGY TM LINEAR ACCELERATOR
    Nedaie, H. A.
    Mosleh Shirazi, M. A.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S475 - S475
  • [6] Dose rates and energy spectra in the maze of a linear accelerator treatment room
    Morgan, S
    Morgan, HM
    Lillicrap, SC
    BRITISH JOURNAL OF RADIOLOGY, 1996, 69 (826): : 977 - 977
  • [7] Monte Carlo simulation approaches to dose distributions for 6 MV photon beams in clinical linear accelerator
    Tartar, Ahmet
    BIOCYBERNETICS AND BIOMEDICAL ENGINEERING, 2014, 34 (02) : 90 - 100
  • [8] Determining optimization of the initial parameters in Monte Carlo simulation for linear accelerator radiotherapy
    Chang, Kwo-Ping
    Wang, Zhi-Wei
    Shiau, An-Cheng
    RADIATION PHYSICS AND CHEMISTRY, 2014, 95 : 161 - 165
  • [9] Monte Carlo study of electron dose distributions produced by the elekta precise linear accelerator
    Nedaie, Hassan Ali
    Mosleh-Shirazi, M. A.
    Shariary, M.
    Gharaati, H.
    Allahverdi, M.
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2006, 11 (06) : 287 - 292
  • [10] Monte Carlo Simulation of Photoneutron Dose in Radiotherapy Room as A Function of Gantry Angles
    Kara, U.
    Mesbahi, A.
    Akkurt, I.
    ACTA PHYSICA POLONICA A, 2015, 128 (2B) : B378 - B380