Evaluation of a 4D Cone-Beam CT Reconstruction Approach using a Simulation Framework

被引:3
|
作者
Hartl, Alexander [1 ]
Yaniv, Ziv [1 ]
机构
[1] Georgetown Univ, Med Ctr, Imaging Sci & Informat Syst Ctr, Dept Radiol, Washington, DC 20007 USA
关键词
COMPUTED-TOMOGRAPHY; MOTION;
D O I
10.1109/IEMBS.2009.5333125
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Current image-guided navigation systems for thoracic abdominal interventions utilize three dimensional (3D) images acquired at breath-hold. As a result they can only provide guidance at a specific point in the respiratory cycle. The intervention is thus performed in a gated manner, with the physician advancing only when the patient is at the same respiratory cycle in which the 3D image was acquired. To enable a more continuous workflow we propose to use 4D image data. We describe an approach to constructing a set of 4D images from a diagnostic CT acquired at breath-hold and a set of intraoperative cone-beam CT (CBCT) projection images acquired while the patient is freely breathing. Our approach is based on an initial reconstruction of a gated 4D CBCT data set. The 3D CBCT images for each respiratory phase are then non-rigidly registered to the diagnostic CT data. Finally the diagnostic CT is deformed based on the registration results, providing a 4D data set with sufficient quality for navigation purposes. In this work we evaluate the proposed reconstruction approach using a simulation framework. A 3D CBCT dataset of an anthropomorphic phantom is deformed using internal motion data acquired from an animal model to create a ground truth 4D CBCT image. Simulated projection images are then created from the 4D image and the known CBCT scan parameters. Finally, the original 3D CBCT and the simulated X-ray images are used as input to our reconstruction method. The resulting 4D data set is then compared to the known ground truth by normalized cross correlation(NCC). We show that the deformed diagnostic CTs are of better quality than the gated reconstructions with a mean NCC value of 0.94 versus a mean 0.81 for the reconstructions.
引用
收藏
页码:5729 / 5732
页数:4
相关论文
共 50 条
  • [31] Low-dose cardiac 4D cone-beam CT image reconstruction using two-cycle data
    Wi, Sunhee
    Lee, Yunjeong
    Lee, Jiseoc
    Abbas, Sajid
    Cho, Seungryong
    Son, Kihong
    Kim, Jin Sung
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [32] Evaluation of gantry speed on image quality and imaging dose for 4D cone-beam CT acquisition
    Santoso, Andrew P.
    Song, Kwang H.
    Qin, Yujiao
    Gardner, Stephen J.
    Liu, Chang
    Chetty, Indrin J.
    Movsas, Benjamin
    Ajlouni, Munther
    Wen, Ning
    RADIATION ONCOLOGY, 2016, 11
  • [33] Framework for iterative cone-beam micro-CT reconstruction
    Benson, TM
    Gregor, J
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (05) : 1335 - 1340
  • [34] Evaluation of Clinical Acceptability of DIR Mapped Contours for Adaptive Radiotherapy with 4D Cone-Beam CT
    Balik, S.
    Weiss, E.
    Jan, N.
    Zhang, L.
    Roman, N.
    Sleeman, W.
    Christensen, G.
    Williamson, J.
    Hugo, G.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [35] Self-contained deep learning-based boosting of 4D cone-beam CT reconstruction
    Madesta, Frederic
    Sentker, Thilo
    Gauer, Tobias
    Werner, Rene
    MEDICAL PHYSICS, 2020, 47 (11) : 5619 - 5631
  • [36] Framework for iterative cone-beam micro-CT reconstruction
    Benson, T
    Gregor, J
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3253 - 3257
  • [37] Evaluation of gantry speed on image quality and imaging dose for 4D cone-beam CT acquisition
    Andrew P. Santoso
    Kwang H. Song
    Yujiao Qin
    Stephen J. Gardner
    Chang Liu
    Indrin J. Chetty
    Benjamin Movsas
    Munther Ajlouni
    Ning Wen
    Radiation Oncology, 11
  • [38] Individualizing 4D cone-beam CT (CBCT) acquisition protocol for external beam radiotherapy
    Mao, W.
    Li, T.
    Munro, P.
    Chao, M.
    Xing, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (03): : S146 - S147
  • [39] Impact of baseline shifts on 4D cone-beam CT images using a 4D phantom driven by lung tumor motions
    Moriwaki, H.
    Shiraishi, K.
    Sakumi, A.
    Ikeda, T.
    Shimizu, W.
    Yoda, K.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S981 - S981
  • [40] Assess Interfractional Tumor Motion in SBRT Lung Treatment Using 4D Cone-Beam CT
    Li, J.
    Yu, Y.
    Xiao, Y.
    MEDICAL PHYSICS, 2013, 40 (06)