Simultaneous point-of-care testing of blood lipid profile and glucose: Performance evaluation of the GCare Lipid Analyzer

被引:1
|
作者
Kim, Ha Nui [1 ]
Yoon, Soo-Young [1 ]
机构
[1] Korea Univ, Dept Lab Med, Guro Hosp, 148 Gurodong Ro, Seoul 08308, South Korea
关键词
diabetes mellitus; dyslipidemia; glucose; lipid; POC; point-of-care testing; EDUCATION-PROGRAM RECOMMENDATIONS; DENSITY-LIPOPROTEIN CHOLESTEROL;
D O I
10.1002/jcla.24055
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background Point-of-care (POC) testing provides quick results and includes tests for blood glucose and lipid profiles. We evaluated the newly developed POC device, the GCare Lipid Analyzer, which is used to measure glucose, total cholesterol (TC), triglyceride (TG), and high-density lipoprotein cholesterol (HDL-C) levels. Methods Venous and capillary blood samples were collected from patients who visited Korea University Guro Hospital. The results obtained using the GCare Lipid Analyzer were compared with those obtained using the TBA 2000FR chemistry analyzer and YSI 2300 STAT Plus analyzer. The glucose system evaluation process was based on the International Organization for Standardization 15197:2013 guidelines. Results The correlation coefficients (R) for TC, TG, and HDL-C were 0.965, 0.969, and 0.943 in capillary blood and 0.969, 0.990, and 0.956 in venous blood, respectively. The total errors for TC, TG, and HDL-C of the lipid profile using venous blood were all acceptable at 6.6%, 9.3%, and 11.6%, respectively. For glucose concentrations <100 mg/dl, 96.1% of the measured glucose levels were within +/- 15 mg/dl in venous samples and 100% were within +/- 15 mg/dl in capillary samples. For glucose concentrations >= 100 mg/dl, 100% and 99.5% of the measured glucose levels were within 15% for venous and capillary blood, respectively. Conclusion The performance of the GCare Lipid Analyzer is acceptable for both blood glucose and lipid profile testing, indicating that it is reliable for use in patients with diabetic dyslipidemia.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] Analytical evaluation of SENDX 100TM blood gas/electrolyte analyzer for point-of-care testing.
    Liu, X
    Mohammad, AA
    Petersen, JR
    Okorodudu, AO
    CLINICAL CHEMISTRY, 1997, 43 : 277 - 277
  • [42] Clinical evaluation of a new point-of-care blood gas analyzer, the ''IRMA system''
    Ozaki, M
    Morioka, N
    Matsukawa, T
    Atarashi, K
    Sessler, DI
    Suzuki, H
    ANESTHESIOLOGY, 1997, 87 (03) : A428 - A428
  • [43] Multicenter study on the analytical performance of a new point-of-care blood gas analyzer
    Goerlach-Graw, A
    Engel, B
    Halwachs-Baumann, G
    Kost, G
    Kraft, M
    Küster, H
    Miedema, K
    Nagel, D
    Normand, H
    Okorodudu, A
    Olander, B
    Rolinski, B
    Wiley, B
    Wu, J
    CLINICAL CHEMISTRY, 2004, 50 (06) : A171 - A171
  • [44] Capillary blood for point-of-care testing
    Tang, Ruihua
    Yang, Hui
    Choi, Jane Ru
    Gong, Yan
    You, MinLi
    Wen, Ting
    Li, Ang
    Li, XiuJun
    Xu, Bo
    Zhang, Sufeng
    Mei, Qibing
    Xu, Feng
    CRITICAL REVIEWS IN CLINICAL LABORATORY SCIENCES, 2017, 54 (05) : 294 - 308
  • [45] Point-of-care testing .3. Determining acceptability of blood glucose meters - Evaluating a blood glucose testing system
    Voss, EM
    Bina, DM
    McNeil, LD
    Johnson, ML
    Cembrowkski, GS
    LABORATORY MEDICINE, 1996, 27 (10) : 679 - 682
  • [46] Clinical investigation of a point-of-care blood ammonia analyzer
    Goggs, Robert
    Serrano, Sergio
    Szladovits, Balazs
    Keir, Iain
    Ong, Ryan
    Hughes, Dez
    VETERINARY CLINICAL PATHOLOGY, 2008, 37 (02) : 198 - 206
  • [47] Point-of-care, portable microfluidic blood analyzer system
    Maleki, Teimour
    Fricke, Todd
    Quesenberry, J. T.
    Todd, Paul W.
    Leary, James F.
    MICROFLUIDICS, BIOMEMS, AND MEDICAL MICROSYSTEMS X, 2012, 8251
  • [48] Performance evaluation of SD A1cCare as a HbA1c analyzer for point-of-care testing
    Lee, Kyunghoon
    Jun, Sun-Hee
    Han, Minje
    Song, Sang Hoon
    Park, Kyoung Un
    Song, Woon Heung
    Song, Junghan
    CLINICAL BIOCHEMISTRY, 2015, 48 (09) : 625 - 627
  • [49] Drug interference with glucose measurements on six handheld glucose meters and a portable glucose analyzer for point-of-care testing in critical care.
    Tang, ZP
    Du, XG
    Louie, RF
    Kost, GJ
    CLINICAL CHEMISTRY, 1999, 45 (06) : A31 - A32
  • [50] Evaluation of the ability of non-laboratorians to perform lipid testing with the ABAXIS Piccolo point-of-care instrument.
    Kazmierczak, SC
    Kazmierczak, DE
    Ammirati, EB
    Hoag, DS
    Netto, MM
    Bleile, DM
    CLINICAL CHEMISTRY, 2004, 50 (06) : A165 - A165