Simultaneous point-of-care testing of blood lipid profile and glucose: Performance evaluation of the GCare Lipid Analyzer

被引:1
|
作者
Kim, Ha Nui [1 ]
Yoon, Soo-Young [1 ]
机构
[1] Korea Univ, Dept Lab Med, Guro Hosp, 148 Gurodong Ro, Seoul 08308, South Korea
关键词
diabetes mellitus; dyslipidemia; glucose; lipid; POC; point-of-care testing; EDUCATION-PROGRAM RECOMMENDATIONS; DENSITY-LIPOPROTEIN CHOLESTEROL;
D O I
10.1002/jcla.24055
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background Point-of-care (POC) testing provides quick results and includes tests for blood glucose and lipid profiles. We evaluated the newly developed POC device, the GCare Lipid Analyzer, which is used to measure glucose, total cholesterol (TC), triglyceride (TG), and high-density lipoprotein cholesterol (HDL-C) levels. Methods Venous and capillary blood samples were collected from patients who visited Korea University Guro Hospital. The results obtained using the GCare Lipid Analyzer were compared with those obtained using the TBA 2000FR chemistry analyzer and YSI 2300 STAT Plus analyzer. The glucose system evaluation process was based on the International Organization for Standardization 15197:2013 guidelines. Results The correlation coefficients (R) for TC, TG, and HDL-C were 0.965, 0.969, and 0.943 in capillary blood and 0.969, 0.990, and 0.956 in venous blood, respectively. The total errors for TC, TG, and HDL-C of the lipid profile using venous blood were all acceptable at 6.6%, 9.3%, and 11.6%, respectively. For glucose concentrations <100 mg/dl, 96.1% of the measured glucose levels were within +/- 15 mg/dl in venous samples and 100% were within +/- 15 mg/dl in capillary samples. For glucose concentrations >= 100 mg/dl, 100% and 99.5% of the measured glucose levels were within 15% for venous and capillary blood, respectively. Conclusion The performance of the GCare Lipid Analyzer is acceptable for both blood glucose and lipid profile testing, indicating that it is reliable for use in patients with diabetic dyslipidemia.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Evaluation of the hemocue blood glucose data management analyzer for point-of-care testing.
    Wahl, HG
    Postma, Y
    Riedlinger, I
    Werner, R
    Schmuelling, RM
    CLINICAL CHEMISTRY, 1998, 44 : A66 - A66
  • [2] Evaluation of the SureStep®Pro blood glucose analyzer for point-of-care testing in neonates.
    Weber, S
    Queen, K
    Riggio, A
    CLINICAL CHEMISTRY, 1998, 44 : A66 - A66
  • [3] Performance evaluation of Barozen Lipid Plus for point-of-care testing of lipid profiles: a method comparison study
    Yu, Soojoung
    Sun, Hwa Yeon
    Yoo, Byungwook
    JOURNAL OF YEUNGNAM MEDICAL SCIENCE, 2023, 40 : S73 - S80
  • [4] Comparative performance of two point-of-care analysers for lipid testing
    Shephard, Mark Ds
    Mazzachi, Beryl C.
    Shephard, Anne K.
    CLINICAL LABORATORY, 2007, 53 (9-12) : 561 - 566
  • [5] Evaluation of a pharmacist-managed lipid clinic that uses point-of-care lipid testing
    Gerrald, Katherine R.
    Dixon, Dave L.
    Barnette, Debra J.
    Williams, Virginia G.
    JOURNAL OF CLINICAL LIPIDOLOGY, 2010, 4 (02) : 120 - 125
  • [6] A Photochemical Dongle for Point-of-Care Evaluation of Blood Lipid Level
    Huang, Xiwei
    Guo, Jinhong
    2017 IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE (BIOCAS), 2017,
  • [7] Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
    Zhu, Huiqing
    Huang, Zhifeng
    Huang, Huimin
    Wang, Caiyou
    Wu, Liting
    Lin, Runpei
    Sun, Baoqing
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2022, (181):
  • [8] PORTABLE SIMULTANEOUS MULTIPLE ANALYTE WHOLE-BLOOD ANALYZER FOR POINT-OF-CARE TESTING
    SCHEMBRI, CT
    OSTOICH, V
    LINGANE, PJ
    BURD, TL
    BUHL, SN
    CLINICAL CHEMISTRY, 1992, 38 (09) : 1665 - 1670
  • [9] Performance evaluation of the Precision™ PCx™ point-of-care blood glucose analyzer using discriminant ratio methodology
    Deyi, VYM
    Philippe, M
    Alexandre, KC
    Nayer, PD
    Hermans, MP
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2002, 40 (10) : 1052 - 1055
  • [10] ECONOMIC EVALUATION OF BLOOD GLUCOSE POINT-OF-CARE TESTING IN THE INTENSIVE CARE UNIT
    Steuten, L. M. G.
    Kip, M.
    Hooijdonk, R.
    Monteban, H.
    Spronk, P.
    Schultz, M.
    VALUE IN HEALTH, 2014, 17 (03) : A250 - A250