Calibration of an Indexing Table Using a High Precision Angle Comparator

被引:1
|
作者
Huang, Ming [1 ,2 ]
Liu, PinKuan [1 ]
Xia, Yangqiu [2 ,3 ]
Hu, Qiu [2 ]
Li, Mengyang [2 ]
Tang, Qiang [2 ,3 ]
机构
[1] Shanghai Jiao Tong Univ, State Key Lab Mech Syst & Vibrat, Shanghai, Peoples R China
[2] China Acad Engn Phys, Inst Machinery Mfg Technol, Mianyang, Peoples R China
[3] Natl Machine Tool Prod Qual Supervis Testing Ctr, Mianyang, Peoples R China
关键词
D O I
10.1051/matecconf/20179507022
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
The calibration of indexing table is important to ensure high accuracy and reliability in angle metrology. In this paper, we present calibrations of an indexing table and uncertainty analysis to evaluate the performance of the high precision angle comparator. The configuration of the angle comparator is explained. In the calibration, a complete closure method is applied to obtain the angle comparator's position error and the indexing table's angular deviation simultaneously based on the least square method. The position error of the angle comparator is evaluated to be about 0.08 ''. The calibration result of an indexing table is compatible with that of the manufacturer and National Institute of Metrology of China within the expanded uncertainty of calibration, 0.04 ''.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] A dynamic CMOS comparator with high precision and resolution
    Wu, R
    Wu, XB
    Yan, XL
    2004: 7TH INTERNATIONAL CONFERENCE ON SOLID-STATE AND INTEGRATED CIRCUITS TECHNOLOGY, VOLS 1- 3, PROCEEDINGS, 2004, : 1567 - 1570
  • [32] Self-calibration of precision angle sensors
    Kiyono, Satoshi
    Zhang, Shizhou
    Seimitsu Kogaku Kaishi/Journal of the Japan Society for Precision Engineering, 1994, 60 (11): : 1591 - 1595
  • [33] A HIGH PRECISION SPECTROMETER TABLE
    KEY, P
    HAYWARD, G
    HARRISON, RG
    LITTLE, VI
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1969, 2 (04): : 374 - &
  • [34] A digital backstage calibration using comparator dithering
    Xiong, Zhaoxin
    Cai, Min
    He, Xiaoyong
    Huazhong Keji Daxue Xuebao (Ziran Kexue Ban)/Journal of Huazhong University of Science and Technology (Natural Science Edition), 2013, 41 (08): : 24 - 29
  • [35] Calibration of non-contact incremental linear encoders using a macro-micro dual-drive high-precision comparator
    Yu, Haoyu
    Liu, Hongzhong
    Li, Xuan
    Ye, Guoyong
    Shi, Yongsheng
    Yin, Lei
    Jiang, Weitao
    Chen, Bangdao
    Liu, Xiaokang
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2015, 26 (09)
  • [36] Using of high precision multimeter digitizer for high transient current calibration
    Roccato, P. E.
    Capra, P. P.
    Bellavia, L.
    Cinnirella, L.
    Lanzillotti, M.
    2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [37] A functional-precision analysis of the Vertical Comparator for the Calibration of geodetic Levelling Systems
    Kuchmister, Janusz
    Goluch, Piotr
    Cmielewski, Kazimierz
    Rzepka, Janusz
    Budzyn, Grzegorz
    MEASUREMENT, 2020, 163
  • [38] HIGH-PRECISION LOAD CELL MASS COMPARATOR
    SCHOONOVER, RM
    JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1979, 84 (05): : 347 - 351
  • [39] HIGH PRECISION LOAD CELL MASS COMPARATOR.
    Schoonover, Randall M.
    Journal of Research of the National Bureau of Standards (United States), 1979, 84 (05): : 347 - 351
  • [40] High-precision diameter measurement of ring and plug gages using a new comparator
    Neugebauer, M
    Lüdicke, F