Evaluation of Pressure Bed Sensor for Automatic SAHS Screening

被引:32
|
作者
Guerrero Mora, Guillermina [1 ]
Kortelainen, Juha M. [2 ]
Palacios Hernandez, Elvia Ruth [3 ]
Tenhunen, Mirja [4 ]
Bianchi, Anna Maria [5 ]
Mendez, Martin O. [3 ]
机构
[1] Univ Autonoma San Luis Potosi, San Luis Potosi 78000, Mexico
[2] Valmistustekniikka Turvallisuustekniikka Tech Res, Tampere 33101, Finland
[3] Univ Autonoma San Luis Potosi, Fac Ciencias, San Luis Potosi 07360, Mexico
[4] Tampere Univ Hosp, Dept Clin Neurophysiol, Tampere 33521, Finland
[5] Politecn Milan, Dept Biomed Engn, I-20133 Milan, Italy
关键词
Polysomnography (PSG); pressure bed sensor (PBS); respiration monitoring; sleep apnea-hypopnea syndrome (SAHS); sleep monitoring; unattended portable monitor; unobtrusive measure; OBSTRUCTIVE SLEEP-APNEA; CHARGE SENSITIVE BED; AIR-FLOW; ELECTROCARDIOGRAM; PLETHYSMOGRAPHY; RESPIRATION; EVENTS; SCSB;
D O I
10.1109/TIM.2014.2366976
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We evaluate the performance of an unobtrusive sleep monitoring system in the detection of the sleep apnea-hypopnea syndrome (SAHS). The proposed system is a pressure bed sensor (PBS) that incorporates multiple pressure sensors into a bed mattress to measure several physiological signals of the sleeping subject: respiration; heart rate; and body movements. An automatic algorithm is developed to calculate a respiratory event index (REI). The recordings of 24 patients with suspected sleep problems are analyzed, and the results are compared with the gold standard methods; first with manual scoring of polysomnography to calculate the apnea-hypopnea index (AHI), and second with automatic detection of REI from the respiratory inductive plethysmography belts. The correlation coefficient between AHI and REI from PBS is up to 0.93. Evaluating the ability of PBS in the diagnosis of pathologic (AHI = 5) and nonpathologic (AHI < 5) subjects, we obtained a sensitivity, specificity, and accuracy of 100%, 92%, and 96%, respectively. To diagnose three levels of SAHS, mild, moderate, and severe, the Cohen's kappa value is 0.76. These findings support that PBS recording could provide a simple and unobtrusive method for detection of SAHS in home monitoring.
引用
收藏
页码:1935 / 1943
页数:9
相关论文
共 50 条
  • [21] EVALUATION OF AN AUTOMATIC BLOOD-PRESSURE INSTRUMENT
    HABERCAM, JW
    KEIL, JE
    SANDIFER, SH
    LOADHOLT, CB
    PREVENTIVE MEDICINE, 1979, 8 (02) : 168 - 168
  • [22] Automatic measurement of jaw movements in ruminants by means of a pressure sensor
    Nydegger, Franz
    Gygax, Lorenz
    Egli, Wendelin
    AGRARFORSCHUNG SCHWEIZ, 2011, 2 (02): : 60 - 65
  • [23] Evaluation of an automatic screening system for detecting retinal pathologies
    Lamard, Mathieu
    Rottier, Jean-Bernard
    Cochener, Beatrice
    Massin, Pascale
    Quellec, Gwenole
    Matta, Sarah Christina Zahida
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2020, 61 (07)
  • [24] Design and Evaluation of Automatic Health Screening System for University
    Imai, Yoshiro
    Miyazaki, Eiichi
    Kamano, Hiroshi
    2013 INTERNATIONAL CONFERENCE ON TECHNOLOGICAL ADVANCES IN ELECTRICAL, ELECTRONICS AND COMPUTER ENGINEERING (TAEECE), 2013, : 411 - 418
  • [25] Automatic system for integrated reliability test of silicon pressure sensor
    Wang, YH
    Yi, DX
    Xiao, QZ
    Wang, SC
    INTERNATIONAL CONFERENCE ON SENSORS AND CONTROL TECHNIQUES (ICSC 2000), 2000, 4077 : 364 - 368
  • [26] Multi-channel Automatic Calibration System of Pressure Sensor
    Jin Wanyu
    Zuo Siran
    Sun Dehui
    Wang Zhongyu
    PROCEEDINGS OF 2016 IEEE ADVANCED INFORMATION MANAGEMENT, COMMUNICATES, ELECTRONIC AND AUTOMATION CONTROL CONFERENCE (IMCEC 2016), 2016, : 506 - 510
  • [27] Screening for Atrial Fibrillation During Automatic Blood Pressure Measurements
    Lowe, Andrew
    Oh, Timothy H.
    Stewart, Ralph
    IEEE JOURNAL OF TRANSLATIONAL ENGINEERING IN HEALTH AND MEDICINE, 2018, 6
  • [28] Monitoring the Relative Blood Pressure Using a Hydraulic Bed Sensor System
    Su, Bo Yu
    Enayati, Moein
    Ho, K. C.
    Skubic, Marjorie
    Despins, Laurel
    Keller, James
    Popescu, Mihail
    Guidoboni, Giovanna
    Rantz, Marilyn
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2019, 66 (03) : 740 - 748
  • [29] Design of a test bed for performance evaluation in a Wireless Sensor Network
    Gao, Robert X.
    Boscardin, Tristan
    2006 IEEE INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE PROCEEDINGS, VOLS 1-5, 2006, : 341 - +
  • [30] Automatic Calibration of Bed-Leaving Sensor Signals Based on Genetic Evolutionary Learning
    Hiramatsu, Daiju
    Madokoro, Hirokazu
    Sato, Kazuhito
    Nakasho, Kazuhisa
    Shimoi, Nobuhiro