Commissioning and quality assurance of Dynamic WaveArc irradiation

被引:13
|
作者
Sato, Sayaka [1 ]
Miyabe, Yuki [1 ]
Takahashi, Kunio [2 ]
Yamada, Masahiro [1 ]
Nakamura, Mitsuhiro [1 ]
Ishihara, Yoshitomo [1 ]
Yokota, Kenji [1 ]
Kaneko, Shuji [1 ]
Mizowaki, Takashi [1 ]
Monzen, Hajime [1 ,3 ]
Hiraoka, Masahiro [1 ]
机构
[1] Kyoto Univ, Grad Sch Med, Dept Radiat Oncol & Image Appl Therapy, Kyoto, Japan
[2] Mitsubishi Heavy Ind Co Ltd, Med Syst Engn Sect, Business Dev Dept, Hiroshima, Japan
[3] Kinki Univ, Grad Sch Med Sci, Dept Radiat Oncol, Osaka, Japan
来源
基金
日本科学技术振兴机构;
关键词
Vero4DRT; Dynamic WaveArc; commissioning; GUIDED RADIOTHERAPY SYSTEM; PARTIAL BREAST IRRADIATION; ARC ROTATION; DELIVERY; ACCURACY; COUCH; FEASIBILITY; THERAPY;
D O I
10.1120/jacmp.v16i2.5080
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A novel three-dimensional unicursal irradiation technique "Dynamic WaveArc" (DWA), which employs simultaneous and continuous gantry and O-ring rotation during dose delivery, has been implemented in Vero4DRT. The purposes of this study were to develop a commissioning and quality assurance procedure for DWA irradiation, and to assess the accuracy of the mechanical motion and dosimetric control of Vero4DRT. To determine the mechanical accuracy and the dose accuracy with DWA irradiation, 21 verification test patterns with various gantry and ring rotational directions and speeds were generated. These patterns were irradiated while recording the irradiation log data. The differences in gantry position, ring position, and accumulated MU (E-G, E-R, and E-MU, respectively) between the planned and actual values in the log at each time point were evaluated. Furthermore, the doses delivered were measured using an ionization chamber and spherical phantom. The constancy of radiation output during DWA irradiation was examined by comparison with static beam irradiation. The mean absolute error (MAE) of E-G and E-R were within 0.1 degrees and the maximum error was within 0.2 degrees. The MAE of E-MU was within 0.7 MU, and maximum error was 2.7 MU. Errors of accumulated MU were observed only around control points, changing gantry, and ring velocity. The gantry rotational range, in which E-MU was greater than or equal to 2.0 MU, was not greater than 3.2%. It was confirmed that the extent of the large differences in accumulated MU was negligibly small during the entire irradiation range. The variation of relative output value for DWA irradiation was within 0.2%, and this was equivalent to conventional arc irradiation with a rotating gantry. In conclusion, a verification procedure for DWA irradiation was designed and implemented. The results demonstrated that Vero4DRT has adequate mechanical accuracy and beam output constancy during gantry and ring rotation.
引用
收藏
页码:73 / 86
页数:14
相关论文
共 50 条
  • [41] BLOOD IRRADIATION - THE NEED FOR TRUE QUALITY ASSURANCE
    BLAKE, M
    POTHIAWALA, M
    TRANSFUSION, 1992, 32 (08) : S91 - S91
  • [42] A set of devices for quality assurance of remote irradiation
    Vajnberg, M.Sh.
    Meditsinskaya Radiologiya (Medical Radiology), 1992, 37 (3-4): : 42 - 43
  • [43] THE QUALITY ASSURANCE IN CRANIOSPINAL IRRADIATION IN THE SUPINE POSITION
    Ficek, K.
    Bacia, S.
    Blamek, S.
    Leszek, M.
    Slosarek, K.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S428 - S428
  • [44] Quality assurance for dynamic tumor tracking
    Witulla, Barbara
    Ziegler, Marc
    Brandt, Tobias
    Prasetio, Heru
    Fietkau, Rainer
    Bert, Christoph
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2021, 31 (04): : 388 - 393
  • [45] Helical tomotherapy dynamic quality assurance
    Balog, John
    Holmes, Tim
    Vaden, Richard
    MEDICAL PHYSICS, 2006, 33 (10) : 3939 - 3950
  • [46] Quality Assurance of Dynamic Renal Scintigraphy
    Samal, M.
    Valousek, J.
    Piepsz, A.
    Brolin, G.
    Heikkinen, J. O.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 41 : S377 - S378
  • [47] A method for quality assurance of dynamic wedge
    Schmidt, EL
    Rittler, J
    Steurer, R
    STRAHLENTHERAPIE UND ONKOLOGIE, 1999, 175 (01) : 39 - 41
  • [48] Film-based delivery quality assurance for robotic radiosurgery: Commissioning and validation
    Blanck, Oliver
    Masi, Laura
    Damme, Marie-Christin
    Hildebrandt, Guido
    Dunst, Juergen
    Siebert, Frank-Andre
    Poppinga, Daniela
    Poppe, Bjoern
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (05): : 476 - 483
  • [49] Commissioning of an A:Si active matrix flat panel dosimeter for IMRT quality assurance
    Moran, J
    Nurushev, T
    Litzenberg, D
    Antonuk, L
    El-Mohri, Y
    Fraass, B
    MEDICAL PHYSICS, 2002, 29 (06) : 1367 - 1367
  • [50] Commissioning an Anthropomorphic Spine and Lung Phantom for Remote Quality Assurance of Spinal Radiosurgery
    Caruthers, D.
    Ibbott, G.
    Shiu, A.
    Chang, E.
    White, R.
    Followill, D.
    MEDICAL PHYSICS, 2010, 37 (06)