Helical tomotherapy dynamic quality assurance

被引:30
|
作者
Balog, John [1 ]
Holmes, Tim
Vaden, Richard
机构
[1] Mohawk Valley Med Phys, Rome, NY 13440 USA
[2] St Agnes HealthCare, Baltimore, MD 21229 USA
[3] EC Rost Oncol LLC, Tallahassee, FL USA
关键词
D O I
10.1118/1.2351952
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A multifaceted tomotherapy quality assurance procedure has been developed. This procedure tests most of the features inherent in the tomotherapy Hi-Art device. This includes the megavoltage imaging quality, spatial and temporal accuracy of the dynamic delivery properties, as well as more traditional beam output characteristics. This is accomplished with a specialized multichannel electrometer that measures collected charge every 100 ms and a Virtual Water cylindrical phantom that holds many ion chambers and differing density insert plugs. Both devices are offered with the Hi-Art product. These tests are presented as well as their sensitivity to beam and delivery variations. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:3939 / 3950
页数:12
相关论文
共 50 条
  • [1] Helical tomotherapy quality assurance
    Balog, John
    Soisson, Emilie
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (01): : S113 - S117
  • [2] Helical tomotherapy quality assurance with ArcCHECK
    Chapman, David
    Barnett, Rob
    Yartsev, Slav
    MEDICAL DOSIMETRY, 2014, 39 (02) : 159 - 162
  • [3] Quality Assurance Procedures for Helical Tomotherapy
    Haaren, P.
    Althof, V.
    Kramer, D.
    Westendorp, H.
    Minken, A.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [4] Quality assurance of a helical tomotherapy machine
    Fenwick, JD
    Tomé, WA
    Jaradat, HA
    Hui, SK
    James, JA
    Balog, JP
    DeSouza, CN
    Lucas, DB
    Olivera, GH
    Mackie, TR
    Paliwal, BR
    PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (13): : 2933 - 2953
  • [5] Mechanical and dosimetric quality assurance for helical tomotherapy
    Ramsey, C. R.
    Gibbons, J.
    Holmes, T.
    Jeong, K.
    Langen, K.
    Low, D.
    Meeks, S.
    Papanikolaou, N.
    Schubert, K.
    Tomsej, M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (03): : S699 - S699
  • [6] Enhanced efficacy in helical tomotherapy quality assurance
    Goddu, S.
    Mutic, S.
    Chaudhari, S.
    Grigsby, J.
    Low, D.
    MEDICAL PHYSICS, 2006, 33 (06) : 2248 - 2248
  • [7] Simplified clinical quality assurance for helical tomotherapy
    Grant, W.
    McGary, J.
    Childress, N.
    CravenS, R.
    Krumbach, T.
    Olivera, G.
    MEDICAL PHYSICS, 2007, 34 (06) : 2561 - 2561
  • [8] Dose reconstruction quality assurance for helical tomotherapy
    Chase, D.
    Ramsey, C.
    Seibert, R.
    MEDICAL PHYSICS, 2006, 33 (06) : 2256 - 2256
  • [9] Computed radiography for helical tomotherapy quality assurance
    Ramsey, C.
    Harris, C.
    MEDICAL PHYSICS, 2006, 33 (06) : 2193 - 2193
  • [10] Patient specific quality assurance for helical tomotherapy
    Thomas, S
    MacKenzie, M
    Field, G
    Syme, A
    Fallone, B
    MEDICAL PHYSICS, 2005, 32 (07) : 2414 - 2414