Helical tomotherapy dynamic quality assurance

被引:30
|
作者
Balog, John [1 ]
Holmes, Tim
Vaden, Richard
机构
[1] Mohawk Valley Med Phys, Rome, NY 13440 USA
[2] St Agnes HealthCare, Baltimore, MD 21229 USA
[3] EC Rost Oncol LLC, Tallahassee, FL USA
关键词
D O I
10.1118/1.2351952
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A multifaceted tomotherapy quality assurance procedure has been developed. This procedure tests most of the features inherent in the tomotherapy Hi-Art device. This includes the megavoltage imaging quality, spatial and temporal accuracy of the dynamic delivery properties, as well as more traditional beam output characteristics. This is accomplished with a specialized multichannel electrometer that measures collected charge every 100 ms and a Virtual Water cylindrical phantom that holds many ion chambers and differing density insert plugs. Both devices are offered with the Hi-Art product. These tests are presented as well as their sensitivity to beam and delivery variations. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:3939 / 3950
页数:12
相关论文
共 50 条
  • [21] TomoMQA: Automated analysis program for MVCT quality assurance of helical tomotherapy
    Han, Min Cheol
    Hong, Chae-Seon
    Chang, Kyung Hwan
    Kim, Jihun
    Han, Su Chul
    Kim, Dong Wook
    Park, Kwangwoo
    Kim, Min-Joo
    Shin, Han Back
    Lee, Se Hyung
    Kim, Jin Sung
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (06): : 151 - 157
  • [22] Quality Assurance for Total Marrow Irradiation (TMI) Using Helical Tomotherapy
    Zhang, X.
    Penagaricano, J.
    Yan, Y.
    Hancock, S.
    Ratanatharathorn, V.
    Corry, P.
    Moros, E.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [23] Performance of ArcCHECK based quality assurance in helical tomotherapy with TomoEdge technology
    Yang, B.
    Geng, H.
    Lam, W. W.
    Cheung, K. Y.
    Yu, S. K.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S941 - S941
  • [24] Dosimetry quality assurance of helical tomotherapy treatments, effect of density inhomogeneities
    Tomsej, M.
    Tondeur, V.
    De Patoul, N.
    Vynckier, S.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S508 - S508
  • [25] Commissioning of PTW Seven29 for Helical TomoTherapy Quality Assurance
    Myers, P.
    Stathakis, S.
    Gutierrez, A.
    Mavroidis, P.
    Mihailidis, D.
    Papanikolaou, N.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [26] Independent quality assurance of a helical tomotherapy machine using the dose magnifying glass
    Wong, J. H. D.
    Hardcastle, N.
    Tome, W. A.
    Bayliss, A.
    Tolakanahalli, R.
    Lerch, M. L. F.
    Petasecca, M.
    Carolan, M.
    Metcalfe, P.
    Rosenfeld, A. B.
    MEDICAL PHYSICS, 2011, 38 (04) : 2256 - 2264
  • [27] FILMLESS QUALITY ASSURANCE FOR HELICAL TOMOTHERAPY WITH A CYLINDRICAL STAIR-LIKE ALUMINIUM PHANTOM
    Althof, V.
    Kramer, D.
    Westendorp, R.
    van Haaren, P.
    Wispelweij, M.
    Minken, A.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S92 - S93
  • [28] Dynamic Jaws and Dynamic Couch in Helical Tomotherapy
    Sterzing, F.
    Uhl, M.
    Sroka-Perez, G.
    Schubert, K.
    Chen, Y.
    Lu, W.
    Mackie, T. R.
    Debus, J.
    Oliveira, G.
    Herfarth, K.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S719 - S719
  • [29] DYNAMIC JAWS AND DYNAMIC COUCH IN HELICAL TOMOTHERAPY
    Sterzing, Florian
    Uhl, Matthias
    Hauswald, Henrik
    Schubert, Kai
    Sroka-Perez, Gabriele
    Chen, Yu
    Lu, Weiguo
    Mackie, Rock
    Debus, Juergen
    Herfarth, Klaus
    Oliveira, Gustavo
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (04): : 1266 - 1273
  • [30] Tomotherapy daily quality assurance phantom
    Jaradat, H
    Paliwal, B
    Mehta, M
    MEDICAL PHYSICS, 2005, 32 (06) : 1989 - 1989