AGREEMENT AND REPRODUCIBILITY OF RETINAL PIGMENT EPITHELIAL DETACHMENT VOLUMETRIC MEASUREMENTS THROUGH OPTICAL COHERENCE TOMOGRAPHY

被引:11
|
作者
Ho, Joseph [1 ]
Adhi, Mehreen [1 ]
Baumal, Caroline [1 ]
Liu, Jonathan [2 ]
Fujimoto, James G. [2 ]
Duker, Jay S. [1 ]
Waheed, Nadia K. [1 ]
机构
[1] Tufts Med Ctr, Dept Ophthalmol, New England Eye Ctr, Boston, MA USA
[2] MIT, Dept Elect Engn & Comp Sci, Elect Res Lab, Cambridge, MA 02139 USA
关键词
automated RPED volume; intraclass coefficient; MACULAR DEGENERATION; AUTOMATIC SEGMENTATION; DRUSEN;
D O I
10.1097/IAE.0000000000000355
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the agreement and reproducibility of retinal pigment epithelial detachment (RPED) volumetric measurements using a commercially available optical coherence tomography software available for the Zeiss Cirrus HD-OCT. Methods: Twelve eyes of 10 patients with a diagnosis of neovascular age-related macular degeneration with RPED, seen at the New England Eye Center between October 2012 and December 2012, were enrolled in the study. Three separate scans per affected eye were obtained using the "Macular Cube 512 x 128" protocol. "Retinal pigment epithelial (RPE) elevation analysis" software was used to measure RPED volumes in the central 3-mm and 5-mm circles by calculating the volume between the "RPE fit" and "true RPE" lines. All 128 raster scans for each eye were exported into the AMIRA software for manual segmentation of RPED volumes in the central 3-mm and 5-mm circles. Interscan reproducibility and manual-to-automated agreement were assessed by intraclass correlation coefficient. Incidence of automated segmentation line error for both RPE fit and true RPE lines in the central 1 mm region was calculated. Results: Average RPED volumes through automated segmentation software were 0.14 mm(3) and 0.21 mm(3) in the central 3-mm and 5-mm circles, respectively. Manual segmentation yielded average RPED volumes of 0.50 mm(3) in the 3-mm circles and 0.92 mm(3) in the 5-mm circles. Manual segmentation yielded significantly greater RPED volumes compared with automated measurements (P < 0.05). Intraclass correlation coefficients across the 3 automated measurements were 0.954 and 0.983 for volume in the 3-mm and 5-mm circles, respectively. Intraclass correlation coefficients between the manual and automatic volumes were 0.296 and 0.337 for the 3-mm and 5-mm circles, respectively. In the central 1 mm region, 11 of the 12 scans had breakdown in RPE fit line, whereas 8 of the 12 scans showed true RPE line breakdown. Conclusion: Automated "RPED elevation" software demonstrated high interscan reproducibility. However, it showed low agreement with manual measurements from high rates of segmentation line breakdown, especially at the level of the RPE fit line (91.7%). Manual measurements resulted in greater volumes compared with automated measurements.
引用
收藏
页码:467 / 472
页数:6
相关论文
共 50 条
  • [11] Optical coherence tomography in the diagnosis of retinal pigment epithelial tear
    Giovannini, A
    Amato, G
    Scassellati-Sforzolini, B
    D'Altobrando, E
    Mariotti, C
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1999, 40 (04) : S315 - S315
  • [12] Reproducibility of retinal thickness measurements using Optical Coherence Tomography
    Wang, RC
    Koozekanani, D
    Herderick, EE
    Roberts, C
    Katz, SE
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1999, 40 (04) : S125 - S125
  • [13] OPTICAL COHERENCE TOMOGRAPHY-MEASURED PIGMENT EPITHELIAL DETACHMENT HEIGHT AS A PREDICTOR FOR RETINAL PIGMENT EPITHELIAL TEARS ASSOCIATED WITH INTRAVITREAL BEVACIZUMAB INJECTIONS
    Chan, Clement K.
    Abraham, Prema
    Meyer, Carsten H.
    Kokame, Gregg T.
    Kaiser, Peter K.
    Rauser, Michael E.
    Gross, Jeffrey G.
    Nuthi, Asha S. D.
    Lin, Steven G.
    Daher, Noha S.
    RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2010, 30 (02): : 203 - 211
  • [14] Dilemmas in treating drusenoid pigment epithelial detachment inactive on optical coherence tomography
    Tripathy, Koushik
    ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2018, 81 (01) : 81 - 82
  • [15] Reproducibility of retinal thickness measurements using optical coherence tomography (OCT)
    Liebmann, JM
    Baumann, M
    Gentile, RC
    DeLuca, R
    Ozols, A
    Chew, SJ
    Ritch, R
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1996, 37 (03) : 4464 - 4464
  • [16] Reproducibility of RTVue Retinal Nerve Fiber Layer Thickness and Optic Disc Measurements and Agreement with Stratus Optical Coherence Tomography Measurements
    Gonzalez-Garcia, Alberto O.
    Vizzeri, Gianmarco
    Bowd, Christopher
    Medeiros, Felipe A.
    Zangwill, Linda M.
    Weinreb, Robert N.
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 147 (06) : 1067 - 1074
  • [17] Enhanced Depth Imaging Optical Coherence Tomography of Retinal Pigment Epithelial Detachment in Age-related Macular Degeneration
    Spaide, Richard F.
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 147 (04) : 644 - 652
  • [18] Correlation of optical coherence tomography with angiography in retinal pigment epithelial detachment associated with age-related macular degeneration
    Sato, T
    Iida, T
    Hagimura, N
    Kishi, S
    RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2004, 24 (06): : 910 - 914
  • [19] Intraoperative optical coherence tomography in retinal detachment
    Junker, B.
    Maier, M.
    Agostini, H.
    Hattenbach, L. -O.
    Pielen, A.
    Framme, C.
    OPHTHALMOLOGE, 2016, 113 (08): : 663 - 667
  • [20] Prediction of Pigment Epithelial Detachment in Optical Coherence Tomography Images using Machine Learning
    Sheeba T.M.
    Raj S.A.A.
    Intl. J. Adv. Comput. Sci. Appl., 2024, 4 (537-546): : 537 - 546