Scope of detection and determination of gallium(III) in industrial ground water by square wave anodic stripping voltammetry on bismuth film electrode

被引:21
|
作者
Kamat, J. V. [1 ]
Guin, Saurav K. [1 ]
Pillai, Jisha S. [1 ]
Aggarwal, Suresh K. [1 ]
机构
[1] Bhabha Atom Res Ctr, Div Fuel Chem, Bombay 400085, Maharashtra, India
关键词
Bismuth-film electrode; Gallium; Square wave anodic stripping voltammetry; Gaussian peak fit; Atomic force microscope; SILVER-BASED ELECTRODE; NEUTRON-ACTIVATION; PRECONCENTRATION; SEPARATION; ELECTROANALYSIS; SPECTROMETRY; EXTRACTION; ALUMINUM; SAMPLES; TRACES;
D O I
10.1016/j.talanta.2011.09.010
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Gallium(III) in ground water may cause human health hazards due to the antineoplastic and antimicrobial activities of gallium. However, the exposure limit of Ga(III) has not been set. This paper demonstrates the scope of employing the square wave anodic stripping voltammetry (SWASV) on bismuth film electrode (BiFE) for selective and sensitive detection of Ga(III) as well as its accurate and precise determination. The key parameters were optimized and the bismuth film morphology was characterized. The performance of BiFE was also compared with that of the mercury film electrode (MFE). The performance of BiFE was also studied for interferences of Zn(II), Cd(II), TI(I) and Cu(II) ions. Gaussian peak fitting was performed to improve the calibration curve and the fitting process revealed almost similar stripping peak heights as obtained from the experimentally observed data though slight improvement in calibration was obtained from the peak area analysis. A good linear dynamic range (R(2) = 0.996) was obtained in the concentration range of 20-100 mu g/L with a limit of detection (LOD) of 6.6 mu g/L (S/N = 3) of Ga(III). A relative standard deviation of 2.9% (n = 10) was obtained for 20 mu g/L of Ga(III) solution. The practical analytical utility of the method was verified by the determination of Ga(III) in spiked water samples, where 100-105% recovery of Ga(III) was achieved. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:256 / 265
页数:10
相关论文
共 50 条
  • [1] Determination of cadmium(II) by square wave anodic stripping voltammetry using bismuth-antimony film electrode
    Yi, Wen Jiao
    Li, Yang
    Ran, Gu
    Luo, Hong Qun
    Li, Nian Bing
    SENSORS AND ACTUATORS B-CHEMICAL, 2012, 166 : 544 - 548
  • [2] Determination of total antimony(III,V) by square-wave anodic stripping voltammetry with in situ plated bismuth-film electrode
    Zong, Ping
    Long, Jiajie
    Nagaosa, Yukio
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 2011, 91 (05) : 421 - 430
  • [3] SQUARE-WAVE ANODIC STRIPPING VOLTAMMETRY (SWASV) FOR THE DETERMINATION OF ECOTOXIC METALS, USING A BISMUTH-FILM ELECTRODE
    Anastasiadou, Z. D.
    Sipaki, I.
    Jannakoudakis, P. D.
    Girousi, S. T.
    ANALYTICAL LETTERS, 2011, 44 (05) : 761 - 777
  • [4] Enhanced resolution of copper and bismuth by addition of gallium in anodic stripping voltammetry with the bismuth film electrode
    Prior, Chad
    Lenehan, Claire E.
    Walker, G. Stewart
    ELECTROANALYSIS, 2006, 18 (24) : 2486 - 2489
  • [5] Determination of thallium in a flow system by anodic stripping voltammetry at a bismuth film electrode
    Korolczuk, Mieczyslaw
    Surmacz, Weronika
    Tyszczuk, Katarzyna
    ELECTROANALYSIS, 2007, 19 (21) : 2217 - 2221
  • [6] Simultaneous determination of lead and tin at the bismuth film electrode by square wave stripping voltammetry and chemometric methods
    Tang, Wenxian
    Bin, Jun
    Fan, Wei
    Zhang, Zhiming
    Yun, Yonghuan
    Liang, Yizeng
    ANALYTICAL METHODS, 2016, 8 (27) : 5475 - 5486
  • [7] Determination of Lead in Ambient Aerosol Samples by Anodic Stripping Voltammetry on a Bismuth Film Electrode
    Palomo Marin, Ma Rosario
    Pinilla Gil, Eduardo
    Calvo Blazquez, Lorenzo
    ELECTROANALYSIS, 2011, 23 (01) : 215 - 221
  • [8] A study on the utility of bismuth-film electrodes for the determination of In(III) in the presence of Pb(II) and Cd(II) by square wave anodic stripping voltammetry
    Charalambous, A
    Economou, A
    ANALYTICA CHIMICA ACTA, 2005, 547 (01) : 53 - 58
  • [9] Determination of Trace Chromium by Square-Wave Adsorptive Cathodic Stripping Voltammetry at an Improved Bismuth Film Electrode
    Zhang, Qing
    Zhong, Shang-wei
    Su, Jia-li
    Li, Xiang-jun
    Zou, Hong
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2013, 160 (04) : H237 - H242
  • [10] DIRECT DETERMINATION OF BISMUTH (III) IN SEA-WATER BY SQUARE-WAVE ANODIC-STRIPPING VOLTAMMETRY AT A GLASSY-CARBON ROTATING-DISK ELECTRODE
    KOMORSKYLOVRIC, S
    ANALYTICA CHIMICA ACTA, 1988, 204 (1-2) : 161 - 167