The impact of dual energy CT imaging on dose calculations for pre-clinical studies

被引:18
|
作者
Vaniqui, Ana [1 ]
Schyns, Lotte E. J. R. [1 ]
Almeida, Isabel P. [1 ]
van der Heyden, Brent [1 ]
van Hoof, Stefan J. [1 ]
Verhaegen, Frank [1 ]
机构
[1] Maastricht Univ, Med Ctr, GROW Sch Oncol & Dev Biol, Dept Radiat Oncol MAASTRO, Maastricht, Netherlands
关键词
Dual energy CT; Small animal radiotherapy; Dose calculations; Tissue segmentation; Monte Carlo; Pre-clinical; TISSUE SEGMENTATION; RADIOTHERAPY; BRACHYTHERAPY; IRRADIATION; ACCURACY; NUMBERS;
D O I
10.1186/s13014-017-0922-9
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: To investigate the feasibility of using dual-energy CT (DECT) for tissue segmentation and kilovolt (kV) dose calculations in pre-clinical studies and assess potential dose calculation accuracy gain. Methods: Two phantoms and an ex-vivo mouse were scanned in a small animal irradiator with two distinct energies. Tissue segmentation was performed with the single-energy CT (SECT) and DECT methods. A number of different material maps was used. Dose calculations were performed to verify the impact of segmentations on the dose accuracy. Results: DECT showed better overall results in comparison to SECT. Higher number of DECT segmentation media resulted in smaller dose differences in comparison to the reference. Increasing the number of materials in the SECT method yielded more instability. Both modalities showed a limit to which adding more materials with similar characteristics ceased providing better segmentation results, and resulted in more noise in the material maps and the dose distributions. The effect was aggravated with a decrease in beam energy. For the ex-vivo specimen, the choice of only one high dense bone for the SECT method resulted in large volumes of tissue receiving high doses. For the DECT method, the choice of more than one kind of bone resulted in lower dose values for the different tissues occupying the same volume. For the organs at risk surrounded by bone, the doses were lower when using the SECT method in comparison to DECT, due to the high absorption of the bone. SECT material segmentation may lead to an underestimation of the dose to OAR in the proximity of bone. Conclusions: The DECT method enabled the selection of a higher number of materials thereby increasing the accuracy in dose calculations. In phantom studies, SECT performed best with three materials and DECT with seven for the phantom case. For irradiations in preclinical studies with kV photon energies, the use of DECT segmentation combined with the choice of a low-density bone is recommended.
引用
收藏
页数:15
相关论文
共 50 条
  • [31] Immunotoxins and other conjugates: Pre-clinical studies
    Bolognesi, A
    Polito, L
    MINI-REVIEWS IN MEDICINAL CHEMISTRY, 2004, 4 (05) : 563 - 583
  • [32] RH1, pre-clinical studies
    Hargreaves, RHJ
    Winski, SL
    Ross, D
    Alley, M
    Sausville, EA
    McGown, AT
    Butler, J
    BRITISH JOURNAL OF CANCER, 1999, 80 : 95 - 95
  • [33] Cardiac Dual-Energy CT Applications and Clinical Impact
    Albrecht M.H.
    De Cecco C.N.
    Nance J.W., Jr.
    Varga-Szemes A.
    De Santis D.
    Eid M.
    Tesche C.
    Apfaltrer G.
    von Knebel Doeberitz P.L.
    Jacobs B.
    Vogl T.J.
    Joseph Schoepf U.
    Current Radiology Reports, 5 (9)
  • [34] A Combined Dose Delivery and Transmission Dose Verification Model for a Small Animal Precision Irradiator for Pre-Clinical Studies
    Granton, P. V.
    van Hoof, S.
    Podesta, M.
    Nijsten, S.
    van Elmpt, W.
    Verhaegen, F.
    MEDICAL PHYSICS, 2011, 38 (06) : 3734 - +
  • [35] Evaluation in animal imaging of the new generation of pre-clinical all-digital PET/CT
    Xiao, W.
    Yan, H.
    Wan, L.
    Xie, Q.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (SUPPL 1) : S864 - S865
  • [36] A simulation framework for pre-clinical studies on dose and image quality:TB concept and first validation
    Smans, Kristien
    Pauwels, Herman
    Rogge, Frank
    Struelens, Lara
    Dragusin, Octavian
    Vanhavere, Filip
    Bosmans, Hilde
    MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913
  • [37] Tissue Doppler imaging for the pre-clinical diagnosis of cardiomyopathy
    Nagueh, SF
    EUROPEAN HEART JOURNAL, 2004, 25 (21) : 1865 - 1866
  • [38] An Online Repository for Pre-Clinical Imaging Protocols (PIPs)
    Gammon, Seth T.
    Cohen, Allison S.
    Lehnert, Adrienne L.
    Sullivan, Daniel C.
    Malyarenko, Dariya
    Manning, Henry Charles
    Hormuth, David A.
    Daldrup-Link, Heike E.
    An, Hongyu
    Quirk, James D.
    Shoghi, Kooresh
    Pagel, Mark David
    Kinahan, Paul E.
    Miyaoka, Robert S.
    Houghton, A. McGarry
    Lewis, Michael T.
    Larson, Peder
    Sriram, Renuka
    Blocker, Stephanie J.
    Pickup, Stephen
    Badea, Alexandra
    Badea, Cristian T.
    Yankeelov, Thomas E.
    Chenevert, Thomas L.
    TOMOGRAPHY, 2023, 9 (02) : 750 - 758
  • [39] Pre-clinical imaging with artificial intelligence and digital twins
    Currie, Geoffrey
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65
  • [40] Multimodal pre-clinical imaging of liver inflammation and fibrosis
    Schmidt-Christensen, Anja
    Nilsson, Julia
    Mayans, Sofia
    Holmberg, Dan
    JOURNAL OF HEPATOLOGY, 2019, 70 (01) : E316 - E316