Evaluation of the use of automatic exposure control and automatic tube potential selection in low-dose cerebrospinal fluid shunt head CT

被引:7
|
作者
Wallace, Adam N. [1 ,2 ]
Vyhmeister, Ross [2 ]
Bagade, Swapnil [1 ,2 ]
Chatterjee, Arindam [1 ,2 ]
Hicks, Brandon [1 ,2 ]
Ramirez-Giraldo, Juan Carlos [3 ]
McKinstry, Robert C. [1 ,2 ]
机构
[1] Barnes Jewish Hosp, Mallinckrodt Inst Radiol, St Louis, MO 63110 USA
[2] Washington Univ, Sch Med, St Louis, MO 63110 USA
[3] Siemens Healthcare, Malvern, PA USA
关键词
Head CT; Cerebrospinal fluid shunt; Automatic tube current modulation; Automatic tube potential selection; CareDose4D; IMAGE QUALITY; ITERATIVE RECONSTRUCTION; TREATED HYDROCEPHALUS; COMPUTED-TOMOGRAPHY; RADIATION; REDUCTION; CHILDREN; SCANNER; CANCER; RISKS;
D O I
10.1007/s00234-015-1508-6
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Cerebrospinal fluid shunts are primarily used for the treatment of hydrocephalus. Shunt complications may necessitate multiple non-contrast head CT scans resulting in potentially high levels of radiation dose starting at an early age. A new head CT protocol using automatic exposure control and automated tube potential selection has been implemented at our institution to reduce radiation exposure. The purpose of this study was to evaluate the reduction in radiation dose achieved by this protocol compared with a protocol with fixed parameters. A retrospective sample of 60 non-contrast head CT scans assessing for cerebrospinal fluid shunt malfunction was identified, 30 of which were performed with each protocol. The radiation doses of the two protocols were compared using the volume CT dose index and dose length product. The diagnostic acceptability and quality of each scan were evaluated by three independent readers. The new protocol lowered the average volume CT dose index from 15.2 to 9.2 mGy representing a 39 % reduction (P < 0.01; 95 % CI 35-44 %) and lowered the dose length product from 259.5 to 151.2 mGy/cm representing a 42 % reduction (P < 0.01; 95 % CI 34-50 %). The new protocol produced diagnostically acceptable scans with comparable image quality to the fixed parameter protocol. A pediatric shunt non-contrast head CT protocol using automatic exposure control and automated tube potential selection reduced patient radiation dose compared with a fixed parameter protocol while producing diagnostic images of comparable quality.
引用
收藏
页码:639 / 644
页数:6
相关论文
共 50 条
  • [31] A Study of Fixed Current and Automatic Tube Current Modulation at a Fixed Low Tube Voltage in Low-dose Computed Tomography (CT) Imaging of Infant Lungs
    Chen, Wushuang
    Tang, Shilong
    Xu, Zhen
    Chen, Xin
    Zheng, Yinglan
    He, Ling
    IRANIAN JOURNAL OF RADIOLOGY, 2023, 20 (02)
  • [32] Estimating patient dose from CT exams that use automatic exposure control: Development and validation of methods to accurately estimate tube current values
    McMillan, Kyle
    Bostani, Maryam
    Cagnon, Christopher H.
    Yu, Lifeng
    Leng, Shuai
    McCollough, Cynthia H.
    McNitt-Gray, Michael F.
    MEDICAL PHYSICS, 2017, 44 (08) : 4262 - 4275
  • [33] DOSE AND IMAGE QUALITY IN LOW-DOSE CT FOR URINARY STONE DISEASE: ADDED VALUE OF AUTOMATIC TUBE CURRENT MODULATION AND ITERATIVE RECONSTRUCTION TECHNIQUES
    Soenen, Olivier
    Balliauw, Christophe
    Oyen, Raymond
    Zanca, Federica
    RADIATION PROTECTION DOSIMETRY, 2017, 174 (02) : 242 - 249
  • [34] Automatic Lung Segmentation of Volumetric Low-Dose CT Scans Using Graph Cuts
    Ali, Asem M.
    Farag, Aly A.
    ADVANCES IN VISUAL COMPUTING, PT I, PROCEEDINGS, 2008, 5358 : 258 - 267
  • [35] AUTOMATIC DETECTION OF BONE MARROW INFILTRATION BY MULTIPLE MYELOMA DETECTION IN LOW-DOSE CT
    Martinez-Martinez, F.
    Kybic, J.
    Lambert, L.
    2015 IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING (ICIP), 2015, : 4813 - 4817
  • [36] Active Shape Dictionary for Automatic Segmentation of Pathological Lung in Low-dose CT Image
    Chen, Geng
    Xiang, Dehui
    Tian, Haihong
    Zhang, Bin
    Zhu, Weifang
    Shi, Fei
    Chen, Xinjian
    MEDICAL IMAGING 2019: IMAGE PROCESSING, 2019, 10949
  • [37] Effects of tube potential selection together with computed tomography automatic tube current modulation on CT imaging performance
    Sookpeng, S.
    Martin, C. J.
    Krisanachinda, A.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2021, 41 (04) : 809 - 831
  • [38] Low-Dose CT With Automatic Tube Current Modulation, Adaptive Statistical Iterative Reconstruction, and Low Tube Voltage for the Diagnosis of Renal Colic: Impact of Body Mass Index
    Gervaise, Alban
    Naulet, Pierre
    Beuret, Florence
    Henry, Christelle
    Pernin, Matthieu
    Portron, Yann
    Lapierre-Combes, Marie
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 202 (03) : 553 - 560
  • [39] Reduction of breast dose in abdominal CT examinations: Effectiveness of automatic exposure control system
    Matsubara, Kosuke
    Koshida, Kichiro
    Noto, Kimiya
    Takata, Tadanori
    RADIATION MEASUREMENTS, 2011, 46 (12) : 2056 - 2059
  • [40] OVERVIEW, PRACTICALTIPS AND POTENTIAL PITFALLS OF USING AUTOMATIC EXPOSURE CONTROL IN CT: SIEMENS CARE DOSE 4D
    Soderberg, Marcus
    RADIATION PROTECTION DOSIMETRY, 2016, 169 (1-4) : 84 - 91