Evaluation of the use of automatic exposure control and automatic tube potential selection in low-dose cerebrospinal fluid shunt head CT

被引:7
|
作者
Wallace, Adam N. [1 ,2 ]
Vyhmeister, Ross [2 ]
Bagade, Swapnil [1 ,2 ]
Chatterjee, Arindam [1 ,2 ]
Hicks, Brandon [1 ,2 ]
Ramirez-Giraldo, Juan Carlos [3 ]
McKinstry, Robert C. [1 ,2 ]
机构
[1] Barnes Jewish Hosp, Mallinckrodt Inst Radiol, St Louis, MO 63110 USA
[2] Washington Univ, Sch Med, St Louis, MO 63110 USA
[3] Siemens Healthcare, Malvern, PA USA
关键词
Head CT; Cerebrospinal fluid shunt; Automatic tube current modulation; Automatic tube potential selection; CareDose4D; IMAGE QUALITY; ITERATIVE RECONSTRUCTION; TREATED HYDROCEPHALUS; COMPUTED-TOMOGRAPHY; RADIATION; REDUCTION; CHILDREN; SCANNER; CANCER; RISKS;
D O I
10.1007/s00234-015-1508-6
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Cerebrospinal fluid shunts are primarily used for the treatment of hydrocephalus. Shunt complications may necessitate multiple non-contrast head CT scans resulting in potentially high levels of radiation dose starting at an early age. A new head CT protocol using automatic exposure control and automated tube potential selection has been implemented at our institution to reduce radiation exposure. The purpose of this study was to evaluate the reduction in radiation dose achieved by this protocol compared with a protocol with fixed parameters. A retrospective sample of 60 non-contrast head CT scans assessing for cerebrospinal fluid shunt malfunction was identified, 30 of which were performed with each protocol. The radiation doses of the two protocols were compared using the volume CT dose index and dose length product. The diagnostic acceptability and quality of each scan were evaluated by three independent readers. The new protocol lowered the average volume CT dose index from 15.2 to 9.2 mGy representing a 39 % reduction (P < 0.01; 95 % CI 35-44 %) and lowered the dose length product from 259.5 to 151.2 mGy/cm representing a 42 % reduction (P < 0.01; 95 % CI 34-50 %). The new protocol produced diagnostically acceptable scans with comparable image quality to the fixed parameter protocol. A pediatric shunt non-contrast head CT protocol using automatic exposure control and automated tube potential selection reduced patient radiation dose compared with a fixed parameter protocol while producing diagnostic images of comparable quality.
引用
收藏
页码:639 / 644
页数:6
相关论文
共 50 条
  • [21] Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: Clinical evaluation
    Mulkens, TH
    Bellinck, P
    Baeyaert, M
    Ghysen, D
    Van Dijck, X
    Mussen, E
    Venstermans, C
    Termote, JL
    RADIOLOGY, 2005, 237 (01) : 213 - 223
  • [22] Combined Use of Automatic Tube Potential Selection with Tube Current Modulation and Iterative Reconstruction Technique in Coronary CT Angiography
    Suh, Young Joo
    Kim, Young Jin
    Hong, Sae Rom
    Hong, Yoo Jin
    Lee, Hye-Jeong
    Hur, Jin
    Choi, Byoung Wook
    RADIOLOGY, 2013, 269 (03) : 722 - 729
  • [23] Combining Automatic Tube Current Modulation with Adaptive Statistical Iterative Reconstruction for Low-Dose Chest CT Screening
    Chen, Jiang-Hong
    Jin, Er-Hu
    He, Wen
    Zhao, Li-Qin
    PLOS ONE, 2014, 9 (04):
  • [24] Procedure for optimal implementation of automatic tube potential selection in pediatric CT to reduce radiation dose and improve workflow
    Browne, Jacinta E.
    Bruesewitz, Michael R.
    Thomas, Vrieze
    Thomas, Kristen B.
    Hull, Nathan C.
    McCollough, Cynthia H.
    Yu, Lifeng
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (02): : 194 - 202
  • [25] A Phantom Evaluation of the Use of Ct Automatic Tube Current Modulation with Low Tube Potentials for Iodinated Contrast Studies
    Sookpeng, Supawitoo
    Martin, Colin J.
    RADIATION PROTECTION DOSIMETRY, 2022, 198 (03) : 188 - 195
  • [26] Low-dose protocol for head CT in evaluation of hydrocephalus in children
    Jonczyk-Potoczna, Katarzyna
    Frankiewicz, Marta
    Warzywoda, Malgorzata
    Strzyzewski, Krzysztof
    Pawlak, Bogdan
    POLISH JOURNAL OF RADIOLOGY, 2012, 77 (01) : 7 - 11
  • [27] Performance evaluation of CT-automatic exposure control devices
    Gutierrez, Daniel
    Schmidt, Sabine
    Denys, Alban
    Schnyder, Pierre
    Bochud, Francois O.
    Verdun, Francis R.
    MEDICAL IMAGING 2007: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2007, 6515
  • [28] Optimization of the pediatric head computed tomography scan image quality: Reducing dose with an automatic tube potential selection in infants
    Morel, Baptiste
    Bouette, Aurelien
    Levy, Pierre
    Antoni, Guillemette
    Chalard, Francois
    Blondiaux, Eleonore
    Le Pointe, Hubert Ducou
    JOURNAL OF NEURORADIOLOGY, 2016, 43 (06) : 398 - 403
  • [29] Automatic Lung Segmentation in Low-Dose CT Image with Contrastive Attention Module
    Yang, Changxing
    Tian, Haihong
    Xiang, Dehui
    Zhu, Weifang
    Shi, Fei
    Chen, Xinjian
    MEDICAL IMAGING 2020: IMAGE PROCESSING, 2021, 11313
  • [30] Automatic detection of pulmonary nodules in low-dose screening thoracic CT examinations
    Fiebich, M
    Wietholt, C
    Renger, BC
    Armato, SG
    Hoffmann, KR
    Wormanns, D
    Diederich, S
    MEDICAL IMAGING 1999: IMAGE PROCESSING, PTS 1 AND 2, 1999, 3661 : 1434 - 1439