MOSFET Assessment of Radiation Dose Delivered to Mice Using the Small Animal Radiation Research Platform (SARRP)

被引:9
|
作者
Ngwa, Wilfred
Korideck, Houari
Chin, Lee M.
Makrigiorgos, G. Mike
Berbeco, Ross I. [1 ]
机构
[1] Brigham & Womens Hosp, Dept Radiat Oncol, Div Med Phys & Biophys, Dana Farber Canc Inst, Boston, MA 02115 USA
关键词
HIGH-RESOLUTION; DOSIMETRY; RADIOTHERAPY; IRRADIATION; DETECTORS; GUIDANCE; BUILDUP; SYSTEM;
D O I
10.1667/RR2536.1
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Ngwa, W., Korideck, H., Chin, L. M., Makrigiorgos, G. M. and Berbeco, R. I. MOSFET Assessment of Radiation Dose Delivered to Mice Using the Small Animal Radiation Research Platform (SARRP). Radiat. Res. 176, 816-820 (2011). The Small Animal Radiation Research Platform (SARRP) is a novel isocentric irradiation system that enables state-of-the-art image-guided radiotherapy research to be performed with animal models. This paper reports the results obtained from investigations assessing the radiation dose delivered by the SARRP to different anatomical target volumes in mice. Surgically implanted metal oxide semiconductor field effect transistors (MOSFET) dosimeters were employed for the dose assessment. The results reveal differences between the calculated and measured dose of -3.5 to 0.5%, -5.2 to -0.7%, -3.9 to 0.5%, -5.9 to 2.5%, -5.5 to 0.5%, and -4.3 to 0% for the left kidney, liver, pancreas, prostate, left lung, and brain, respectively. Overall, the findings show less than 6% difference between the delivered and calculated dose, without tissue heterogeneity corrections. These results provide a useful assessment of the need for tissue heterogeneity corrections in SARRP dose calculations for clinically relevant tumor model sites. (C) 2011 by Radiation Research Society
引用
收藏
页码:816 / 820
页数:5
相关论文
共 50 条
  • [41] Small animal photon counting cone-beam CT on a preclinical radiation research platform to improve radiation dose calculation accuracy
    Hu, Xiaoyu
    Zhong, Yuncheng
    Lai, Youfang
    Shen, Chenyang
    Yang, Kai
    Jia, Xun
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (19):
  • [42] Establishing MicroCT Imaging Dose to Mouse Organs Using a Validated Monte Carlo Model of the Small Animal Radiation Research Platform
    Johnstone, C. D.
    Bazalova-Carter, M.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [43] Novel Quality Assurance Procedure for the Small Animal Radiation Research Platform Using Commercial OSLDs
    Keller, J.
    Lockamy, V.
    Glenn, S.
    Harrison, A.
    MEDICAL PHYSICS, 2015, 42 (06) : 3468 - 3468
  • [44] On Achievable Dose-Rates in a Small Animal Radiation Research Platform (SARRP) and Energy Dependency of EBT3, EBT-XD, and HD-V2 Radiochromic Films
    Mulrow, D.
    Hao, Y.
    Lo, W.
    Mazur, T.
    Sobotka, L.
    Darafsheh, A.
    MEDICAL PHYSICS, 2022, 49 (06) : E365 - E365
  • [45] Characterization of Radiation Response of MRI-Guided 3D Conformal Treatment of U251 and 9L Glioblastoma Rat Models Using a Small Animal Radiation Research Platform (SARRP)
    Valadie, G.
    Cabral, G.
    Shadaia, S.
    Farmer, K.
    Panda, S.
    Tavarekere, N.
    Lee, I.
    Knight, R.
    Joiner, M.
    Brown, S.
    Ewing, J.
    MEDICAL PHYSICS, 2020, 47 (06) : E715 - E716
  • [46] Measure the Actual Radiation Dose Delivered for Prostate IMRT Treatment Using An Implantable MOSFET Dosimeter
    Cao, J.
    Showalter, T.
    Buzurovic, I.
    Peng, C.
    Harrison, A.
    Xiao, Y.
    Yu, Y.
    MEDICAL PHYSICS, 2011, 38 (06) : 3517 - U72
  • [47] Validation of GPU-accelerated superposition-convolution dose computations for the Small Animal Radiation Research Platform
    Cho, Nathan
    Tsiamas, Panagiotis
    Velarde, Esteban
    Tryggestad, Erik
    Jacques, Robert
    Berbeco, Ross
    McNutt, Todd
    Kazanzides, Peter
    Wong, John
    MEDICAL PHYSICS, 2018, 45 (05) : 2252 - 2265
  • [48] Quality Assessment of Stereotactic Radiosurgery of a Melanoma Brain Metastases Model Using a Mouselike Phantom and the Small Animal Radiation Research Platform
    Wu, Cheng-Chia
    Chaudhary, Kunal R.
    Na, Yong Hum
    Welch, David
    Black, Paul J.
    Sonabend, Adam M.
    Canoll, Peter
    Saenger, Yvonne M.
    Wang, Tony J. C.
    Wuu, Cheng-Shie
    Hei, Tom K.
    Cheng, Simon K.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (01): : 191 - 201
  • [49] Quality assurance of the dose delivered by small radiation segments
    Hansen, VN
    Evans, PM
    Budgell, GJ
    Mott, JHL
    Williams, PC
    Brugmans, MJP
    Wittkamper, FW
    Mijnheer, BJ
    Brown, K
    PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (09): : 2665 - 2675
  • [50] The small animal radiation research platform: Benchtop cone-beam CT
    Ford, E.
    Kennedy, C.
    McNutt, T.
    Armour, E.
    Iordachita, I.
    Kazanzides, P.
    Wong, J.
    MEDICAL PHYSICS, 2006, 33 (06) : 2017 - 2017