Implantable Cardioverter-Defibrillator Shock after Stenting Across the Device Leads

被引:3
|
作者
Mehra, Sanjay [1 ]
Chelu, Mihail Gabriel [1 ,2 ]
机构
[1] Brody Sch Med, East Carolina Heart Inst, Dept Cardiovasc Sci, Greenville, NC 27834 USA
[2] Univ Utah, Sch Med, Div Electrophysiol, Salt Lake City, UT USA
来源
TEXAS HEART INSTITUTE JOURNAL | 2016年 / 43卷 / 01期
关键词
Cardiac pacing; artificial; death; sudden; cardiac/prevention; control; defibrillators; implantable/adverse effects; device removal/education; equipment failure; kidney failure; chronic/therapy; prosthesis implantation/adverse effects; stents/adverse effects; stress; psychological/etiology; treatment refusal;
D O I
10.14503/THIJ-14-4759
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A 45-year-old man with nonischemic cardiomyopathy and end-stage renal disease had lived uneventfully with a cardiac resynchronization therapy defibrillator (CRT-D) for 5 years. Less than a month before presenting at our institution, he had undergone stenting of his partially occluded subclavian vein, to relieve stenosis of the ipsilateral arteriovenous fistula that was used for his hemodialysis. The CRT-D subsequently discharged. Device interrogation revealed that electrical noise originating from leads damaged by the stent had caused the inappropriate shock and intermittent electrical discharges thereafter. The patient was highly traumatized by these events and insisted upon device removal, which deprived him of a potentially life-saving intervention. He later had a cardiac arrest that resulted in sustained profound hypoxic ischemic encephalopathy with minimal neurologic recovery: his family placed him in a long-term care facility on ventilator support, with a tracheostomy and feeding tube. This situation might have been avoided through collaboration between the interventional radiologist and the electrophysiologist. To our knowledge, this is the first report of a patient with nonischemic cardiomyopathy and end-stage renal disease who presented with inappropriate defibrillator discharge caused by lead damage secondary to stenting across the leads.
引用
收藏
页码:88 / 90
页数:3
相关论文
共 50 条
  • [41] Development of the Implantable Cardioverter-Defibrillator
    Maron, Barry J.
    Estes, N. A. Mark
    Rowin, Ethan J.
    Maron, Martin S.
    Reynolds, Matthew R.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2023, 82 (04) : 353 - 373
  • [42] THE IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - AN UPDATE
    MIROWSKI, M
    JOURNAL OF CARDIOVASCULAR MEDICINE, 1984, 9 (03): : 191 - &
  • [43] The subcutaneous implantable cardioverter-defibrillator
    Grace, Andrew
    CURRENT OPINION IN CARDIOLOGY, 2014, 29 (01) : 10 - 19
  • [44] A prophylactic implantable cardioverter-defibrillator?
    Brugada, P
    Wellens, F
    Andries, E
    AMERICAN JOURNAL OF CARDIOLOGY, 1996, 78 : 128 - 133
  • [45] Implantable cardioverter-defibrillator advances
    Lerman, RD
    Cannom, DS
    CURRENT OPINION IN CARDIOLOGY, 1996, 11 (01) : 16 - 22
  • [46] AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    LATTUCA, J
    VELTRI, EP
    AMERICAN HEART JOURNAL, 1989, 118 (06) : 1359 - 1359
  • [47] Indications for implantable Cardioverter-Defibrillator
    Tebbenjohanns, J
    Niehaus, M
    Korte, T
    DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 2002, 127 (41) : 2120 - 2124
  • [48] AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    EPSTEIN, AE
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1991, 102 (01): : 160 - 161
  • [49] The implantable cardioverter-defibrillator: An update
    Miller, Jared D.
    Yousuf, Omair
    Berger, Ronald D.
    TRENDS IN CARDIOVASCULAR MEDICINE, 2015, 25 (07) : 606 - 611
  • [50] Implantable cardioverter-defibrillator utilization
    Camm, John
    EUROPACE, 2011, 13 (03): : 448 - 448