Implantable Cardioverter-Defibrillator Shock after Stenting Across the Device Leads

被引:3
|
作者
Mehra, Sanjay [1 ]
Chelu, Mihail Gabriel [1 ,2 ]
机构
[1] Brody Sch Med, East Carolina Heart Inst, Dept Cardiovasc Sci, Greenville, NC 27834 USA
[2] Univ Utah, Sch Med, Div Electrophysiol, Salt Lake City, UT USA
来源
TEXAS HEART INSTITUTE JOURNAL | 2016年 / 43卷 / 01期
关键词
Cardiac pacing; artificial; death; sudden; cardiac/prevention; control; defibrillators; implantable/adverse effects; device removal/education; equipment failure; kidney failure; chronic/therapy; prosthesis implantation/adverse effects; stents/adverse effects; stress; psychological/etiology; treatment refusal;
D O I
10.14503/THIJ-14-4759
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A 45-year-old man with nonischemic cardiomyopathy and end-stage renal disease had lived uneventfully with a cardiac resynchronization therapy defibrillator (CRT-D) for 5 years. Less than a month before presenting at our institution, he had undergone stenting of his partially occluded subclavian vein, to relieve stenosis of the ipsilateral arteriovenous fistula that was used for his hemodialysis. The CRT-D subsequently discharged. Device interrogation revealed that electrical noise originating from leads damaged by the stent had caused the inappropriate shock and intermittent electrical discharges thereafter. The patient was highly traumatized by these events and insisted upon device removal, which deprived him of a potentially life-saving intervention. He later had a cardiac arrest that resulted in sustained profound hypoxic ischemic encephalopathy with minimal neurologic recovery: his family placed him in a long-term care facility on ventilator support, with a tracheostomy and feeding tube. This situation might have been avoided through collaboration between the interventional radiologist and the electrophysiologist. To our knowledge, this is the first report of a patient with nonischemic cardiomyopathy and end-stage renal disease who presented with inappropriate defibrillator discharge caused by lead damage secondary to stenting across the leads.
引用
收藏
页码:88 / 90
页数:3
相关论文
共 50 条
  • [21] Inappropriate implantable cardioverter-defibrillator shock in unusual circumstances
    Suran, Maria-Claudia-Berenice
    Margulescu, Andrei-Dumitru
    Siliste, Calin
    Vinereanu, Dragos
    ACTA CARDIOLOGICA, 2016, 71 (03) : 369 - 370
  • [22] Implantable cardioverter-defibrillator failure unmasked by a "lucky" shock
    Eng, Lim K.
    Walters, Tomos E.
    Gould, Paul A.
    JOURNAL OF ARRHYTHMIA, 2012, 28 (04) : 254 - 257
  • [23] Survival After Implantable Cardioverter-Defibrillator Shocks
    Aktas, Mehmet K.
    Younis, Arwa
    Zareba, Wojciech
    Kutyifa, Valentina
    Klein, Helmut
    Daubert, James P.
    Estes, Mark
    McNitt, Scott
    Polonsky, Bronislava
    Goldenberg, Ilan
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2021, 77 (20) : 2453 - 2462
  • [24] IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - LIFESAVER OR A DEVICE LOOKING FOR A DISEASE
    ZIPES, DP
    CIRCULATION, 1994, 89 (06) : 2934 - 2936
  • [25] The effect of device advisories on implantable cardioverter-defibrillator therapy
    Amin M.S.
    Ellenbogen K.A.
    Current Cardiology Reports, 2010, 12 (5) : 361 - 366
  • [26] SILICONE POUCH FOR PROTECTION OF AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR LEADS
    CILLEY, JH
    CERNAIANU, AC
    LIBBY, JA
    BALDINO, WA
    DELROSSI, AJ
    ANNALS OF THORACIC SURGERY, 1991, 51 (03): : 504 - 505
  • [27] Mechanical and electrical dysfunction of Riata implantable cardioverter-defibrillator leads
    Demirel, Fatma
    Adiyaman, Ahmet
    Delnoy, Peter Paul H. M.
    Smit, Jaap Jan J.
    Misier, Anand R. Ramdat
    Elvan, Arif
    EUROPACE, 2014, 16 (12): : 1787 - 1794
  • [28] Occurrence and extraction of implantable cardioverter-defibrillator leads with conductor externalization
    Maciag, Aleksander
    Syska, Pawel
    Sterlinski, Maciej
    Kolodzinska, Agnieszka
    Oreziak, Artur
    Kusmierski, Krzysztof
    Marciniak-Emmons, Marta
    Przybylski, Andrzej
    Szumowski, Lukasz
    Opolski, Grzegorz
    Szwed, Hanna
    CARDIOLOGY JOURNAL, 2018, 25 (02) : 221 - 228
  • [29] Failure of implantable cardioverter-defibrillator leads: A matter of lead size?
    Rordorf, Roberto
    Poggio, Luca
    Savastano, Simone
    Vicentini, Alessandro
    Petracci, Barbara
    Chieffo, Enrico
    Klersy, Catherine
    Landolina, Maurizio
    HEART RHYTHM, 2013, 10 (02) : 184 - 190
  • [30] Psychological Effects of Implantable Cardioverter-Defibrillator Leads under Advisory
    Heatherly, Steven J.
    Simmons, Tony
    Fitzgerald, David M.
    Mitchell, Mark
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2011, 34 (06): : 694 - 699