Neutron therapy using medical linacs

被引:0
|
作者
Adnani, N [1 ]
Fallone, BG [1 ]
机构
[1] Cross Canc Inst, Edmonton, AB T6G 1Z2, Canada
关键词
neutrons; linacs; BNCT; BNCEPT;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Measurements of the neutron flux generated at depth in water by an 18 MV photon beam of a Varian 2100C linac are presented, The thermal neutron flux at depth exhibits a maximum at around 5 cm of 2.37 x 10(6) n/cm(2)/s fur a 40 x 40 cm(2) field size, 100 cm SSD and a photon output of 400 MU/min. Because of the low thermal neutron flux, a large concentration of boron-10 at the tumor is required to achieve significant dose deposition by boron neutron capture reaction, To reduce the concentration required at the tumor it is necessary to increase the irradiation time without a significant increase in the photon dose. This is found to be possible by completely closing the collimator jaws which results in a 1000 times reduction in the photon dose. Also the thermal neutron flux can be further increased by reducing the SSD to the minimum possible (50 cm in our case). In such conditions, the linac can be considered to be operating in a neutron beam mode. la addition, a 25 MV photon beam will generated 10 times more neutrons than an 18 MV beam This report demonstrates that boron neutron capture therapy and boron neutron capture enhanced photon therapy can be successfully performed using the neutron beam generated by high energy medical linacs.
引用
收藏
页码:133 / 136
页数:4
相关论文
共 50 条
  • [1] Neutron imaging using medical linacs
    Adnani, N
    Fallone, BG
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 2457 - 2460
  • [2] USE OF A NEUTRON REMMETER TO MEASURE LEAKAGE NEUTRONS FROM MEDICAL LINACS
    ROGERS, DWO
    VANDYK, G
    MEDICAL PHYSICS, 1980, 7 (04) : 426 - 426
  • [3] MEASUREMENTS OF THE PARASITIC NEUTRON DOSE AT ORGANS FROM MEDICAL LINACS AT DIFFERENT ENERGIES BY USING BUBBLE DETECTORS
    Alikaniotis, K.
    Severgnini, M.
    Giannini, G.
    Milan, V.
    RADIATION PROTECTION DOSIMETRY, 2018, 180 (1-4) : 267 - 272
  • [4] COMPARISON OF NEUTRON MEASUREMENTS AT LINACS USING BUBBLE DOSIMETERS TO OTHER NEUTRON DETECTORS
    NELSON, ME
    GORDON, R
    RADIATION PROTECTION DOSIMETRY, 1993, 47 (1-4) : 547 - 550
  • [5] Neutron and photon spectra in LINACs
    Vega-Carrillo, H. R.
    Martinez-Ovalle, S. A.
    Lallena, A. M.
    Mercado, G. A.
    Benites-Rengifo, J. L.
    APPLIED RADIATION AND ISOTOPES, 2012, 71 : 75 - 80
  • [6] New medical linacs
    Schonberg, RG
    Mishin, AV
    APPLICATION OF ACCELERATORS IN RESEARCH AND INDUSTRY - PROCEEDINGS OF THE FOURTEENTH INTERNATIONAL CONFERENCE, PTS 1 AND 2, 1997, (392): : 1289 - 1292
  • [7] Active neutron monitor for radiotherapy linacs
    Gomez, F.
    Brualla, L.
    Gonzalez-Castano, D. M.
    Guerre, F. X.
    Gutierrez, F.
    Iglesias, A.
    Pazos, A.
    Pena, J.
    Pombar, M.
    Rosello, J. V.
    Zapata, M.
    Sanchez-Doblado, F.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S255 - S255
  • [8] Linacs for Medical Isotope Production
    Pramudita, A.
    ATOM INDONESIA, 2011, 37 (01) : 1 - 4
  • [9] Medical applications of electron linacs
    Wambersie, A
    Gahbauer, RA
    PROCEEEDINGS OF THE XVIII INTERNATIONAL LINEAR ACCELERATOR CONFERENCE, VOLS 1 AND 2, 1996, 96 (07): : 610 - 610
  • [10] H*(10) and neutron spectra around linacs
    Hector Rene Vega-Carrillo
    Arturo Ortíz-Hernandez
    Victor Martin Hernandez-Davila
    Berenice Hernández-Almaraz
    Teodoro Rivera Montalvo
    Journal of Radioanalytical and Nuclear Chemistry, 2010, 283 : 537 - 540