Prediction of Left Main or 3-Vessel Disease Using Myocardial Perfusion Reserve on Dynamic Thallium-201 Single-Photon Emission Computed Tomography With a Semiconductor Gamma Camera

被引:60
|
作者
Shiraishi, Shinya [1 ]
Sakamoto, Fumi [1 ]
Tsuda, Noriko [1 ]
Yoshida, Morikatsu [1 ]
Tomiguchi, Seiji [2 ]
Utsunomiya, Daisuke [1 ]
Ogawa, Hisao [3 ]
Yamashita, Yasuyuki [1 ]
机构
[1] Kumamoto Univ, Grad Sch Life Sci, Dept Diagnost Radiol, Kumamoto 8608556, Japan
[2] Kumamoto Univ, Dept Diagnost Med Imaging, Sch Hlth Fac Life Sci, Kumamoto, Japan
[3] Kumamoto Univ, Dept Cardiovasc Med, Grad Sch Life Sci, Kumamoto, Japan
关键词
Balanced ischemia; Myocardial perfusion reserve; Semiconductor gamma camera; Thallium-201; Three-vessel disease; TRANSIENT ISCHEMIC DILATION; CORONARY-ARTERY-DISEASE; FLOW RESERVE; WALL-MOTION; BLOOD-FLOW; SPECT; QUANTIFICATION; PET; TRACER; PLAQUE;
D O I
10.1253/circj.CJ-14-0932
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Myocardial perfusion imaging (MPI) may fail to detect balanced ischemia. We evaluated myocardial perfusion reserve (MPR) using Tl dynamic single-photon emission computed tomography (SPECT) and a novel cadmium zinc telluride (CZT) camera for predicting 3-vessel or left main coronary artery disease (CAD). Methods and Results: A total of 55 consecutive patients with suspected CAD underwent SPECT-MPI and coronary angiography. The MPR index was calculated using the standard 2-compartment kinetic model. We analyzed the utility of MPR index, other SPECT findings, and various clinical variables. On multivariate analysis, MPR index and history of previous myocardial infarction (MI) predicted left main and 3-vessel disease. The area under the receiver operating characteristic curve was 0.81 for MPR index, 0.699 for history of previous MI, and 0.86 for MPR index plus history of previous MI. MPR index <= 1.5 yielded the highest diagnostic accuracy. Sensitivity, specificity, and accuracy were 86%, 78%, and 80%, respectively, for MPR index, 64%, 76%, 73% for previous MI, and 57%, 93%, and 84% for MPR index plus history of previous MI. Conclusions: Quantification of MPR using dynamic SPECT and a novel CZT camera may identify balanced ischemia in patients with left main or 3-vessel disease.
引用
收藏
页码:623 / 631
页数:9
相关论文
共 50 条
  • [21] Left bundle branch block and coronary artery disease: Accuracy of dipyridamole thallium-201 single-photon emission computed tomography in patients with exercise anteroseptal perfusion defects
    N. E. Lebtahi
    J. C. Stauffer
    A. Bischof Delaloye
    Journal of Nuclear Cardiology, 1997, 4 : 266 - 273
  • [22] Limited performance of quantitative assessment of myocardial function by thallium-201 gated myocardial single-photon emission tomography
    Lee, DS
    Ahn, JY
    Kim, SK
    Oh, BH
    Seo, JD
    Chung, JK
    Lee, MC
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 2000, 27 (02) : 185 - 191
  • [23] Left bundle branch block and coronary artery disease: Accuracy of dipyridamole thallium-201 single-photon emission computed tomography in patients with exercise anteroseptal perfusion defects
    Lebtahi, NE
    Stauffer, JC
    Delaloye, AB
    JOURNAL OF NUCLEAR CARDIOLOGY, 1997, 4 (04) : 266 - 273
  • [24] Limited performance of quantitative assessment of myocardial function by thallium-201 gated myocardial single-photon emission tomography
    Dong Soo Lee
    Ji Young Ahn
    Seok Ki Kim
    Byung Hee Oh
    Jung Don Seo
    June-Key Chung
    Myung Chul Lee
    European Journal of Nuclear Medicine, 2000, 27 : 185 - 191
  • [25] Comparison of real-time coherent contrast imaging to dipyridamole thallium-201 single-photon emission computed tomography for assessment of myocardial perfusion and left ventricular wall motion
    Oraby, MA
    Hays, J
    Maklady, FA
    El-Hawary, AA
    Yaneza, LO
    Zabalgoitia, M
    AMERICAN JOURNAL OF CARDIOLOGY, 2002, 90 (05): : 449 - 454
  • [26] Accuracy of washout rate analysis for thallium-201 single-photon emission computed tomography myocardial perfusion imaging using cadmium zinc telluride detectors: A phantom study
    Ishihara, Masaru
    Onoguchi, Masahisa
    Shibutani, Takayuki
    JRSM CARDIOVASCULAR DISEASE, 2020, 9
  • [27] Study of myocardial perfusion reserve with single-photon-emission computed tomography with thallium-201 and pharmacological stress with adenosine triphosphate in patients with cardiovascular risk factors
    Rodrigo, AH
    Velloso, MJG
    Ezquerra, EA
    Echevarria, JAR
    Caro, DM
    REVISTA ESPANOLA DE CARDIOLOGIA, 1997, 50 (10): : 696 - 708
  • [28] Comparison of contrast-enhanced magnetic resonance imaging and Thallium-201 single-photon emission computed tomography for prediction of reversible left ventricular dysfunction
    Regenfus, M
    Schlundt, C
    von Erffa, J
    Krähner, R
    Schmidt, M
    Oesingmann, N
    Platsch, G
    Kuwert, T
    Daniel, WG
    CIRCULATION, 2003, 108 (17) : 697 - 697
  • [29] Exercise Thallium-201 single-photon emission computed tomography for evaluation of coronary artery bypass graft patency
    Kureshi, SA
    AMERICAN JOURNAL OF CARDIOLOGY, 1996, 78 (05): : 608 - 608
  • [30] Systematic review of the diagnostic accuracy of thallium-201 single-photon emission computed tomography in the detection of recurrent glioma
    Vos, M
    Tony, B
    Hoekstra, O
    Postma, T
    Heimans, J
    Hooft, L
    NEURO-ONCOLOGY, 2005, 7 (03) : 396 - 396