Validity and reliability of the T-Scan®III for measuring force under laboratory conditions

被引:45
|
作者
Cerna, M. [1 ]
Ferreira, R. [2 ]
Zaror, C. [2 ]
Navarro, P. [3 ]
Sandoval, P. [2 ]
机构
[1] Univ La Frontera, Fac Sci & Engn, Dept Phys Sci, Temuco, Chile
[2] Univ La Frontera, Fac Dent, Dept Pediat Dent & Orthodont, Destristry Program, Temuco, Chile
[3] Univ La Frontera, Fac Dent, Temuco, Chile
关键词
validity; reliability; T-Scan((R))III; bite force; bite registration; dental occlusion;
D O I
10.1111/joor.12284
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Although measuring bite force is an important indicator of the health of the masticatory system, few commercially available transducers have been validated for routine clinical use. T-Scan((R))III Occlusal Analysis System allows to record the bite force distribution, indicating its relative intensity and occlusal timing. Nevertheless, even fewer studies have evaluated the validity and reliability of the latest generation of the T-Scan((R)) occlusal analysis system. To determine the validity and reliability of the T-Scan((R))III system when measuring total absolute bite force under laboratory conditions. Known forces were applied to 18 T-Scan((R))III sensors, which were classified into two groups differentiated by their production series. Both Lin's concordance correlation coefficient (CCC) and the intra-class correlation coefficient (ICC) were used to assess the system's reliability and validity. Considering all the sensors studied, a substantial level (Lin's CCC 0969) and a very good level of reliability (CCI 0994) were obtained. When evaluating the validity of the system, a poor (Lin's CCC 0530) and moderate (ICC 0693) agreement were also obtained. The main factor that negatively influenced the validity of the T-Scan((R))III under these study conditions was the significant difference in the behaviour of the two sensor groups. The T-Scan((R))III showed a high degree of reliability when used to perform consecutive measurements. However, the system showed an insufficient degree of validity for measuring absolute force when estimating total occlusal force under laboratory conditions.
引用
收藏
页码:544 / 551
页数:8
相关论文
共 50 条
  • [21] Using the T-scan III System to Analyze Occlusal Function in Mandibular Reconstruction Patients: A Pilot Study
    Liu, Chao-Wei
    Chang, Yang-Ming
    Shen, Yu-Fu
    Hong, Hsiang-Hsi
    BIOMEDICAL JOURNAL, 2015, 38 (01) : 52 - 57
  • [22] Comparison of closure occlusal force parameters in post-orthodontic and non-orthodontic subjects using T-Scan® III DMD occlusal analysis
    Qadeer, Sarah
    Yang, Lili
    Sarinnaphakorn, Letrit
    Kerstein, Robert B.
    CRANIO-THE JOURNAL OF CRANIOMANDIBULAR & SLEEP PRACTICE, 2016, 34 (06): : 395 - 401
  • [23] T-Scan III® HD sensors' accuracy influenced by in vitro occlusal simulation of anterior and posterior teeth
    Martins, Mario
    Pereira, Joana
    Fonseca, Julio
    Nicolau, Pedro
    ANNALS OF MEDICINE, 2018, 50 : S85 - S85
  • [24] A study on the mandibular arc of closure reproducibility using the T-Scan III computerized occlusal analysis system
    Balos, Monica Dora
    Soaita, Claudia
    Cerghizan, Diana
    Popsor, Sorin
    8TH INTERNATIONAL CONFERENCE INTERDISCIPLINARITY IN ENGINEERING, INTER-ENG 2014, 2015, 19 : 916 - 920
  • [25] RELIABILITY AND VALIDITY EFFECTS UNDER MAIL SURVEY CONDITIONS
    JONES, WH
    LANG, JR
    JOURNAL OF BUSINESS RESEARCH, 1982, 10 (03) : 339 - 353
  • [26] Association between Occlusal Forces and Periodontal Conditions in Untreated Chronic Periodontitis Using T-Scan: A Retrospective Study
    Soundarajan, Subasree
    Gajendran, Priyalochana
    JOURNAL OF RESEARCH IN MEDICAL AND DENTAL SCIENCE, 2022, 10 (01):
  • [27] The Reliability and Validity of the Loadsol® under Various Walking and Running Conditions
    Renner, Kristen E.
    Williams, D. S. Blaise
    Queen, Robin M.
    SENSORS, 2019, 19 (02)
  • [28] Adjustment of Occlusal Splint with Synchronized T-Scan III Digital Occlusal Analysis System and Bio-EMG III in a Patient with Sleep Bruxism
    Karakis, Duygu
    Demirdag, Elif Didem
    JOURNAL OF ADVANCED ORAL RESEARCH, 2021, 12 (01) : 170 - 175
  • [29] Reliability and validity of the Actiwatch and Clouclip for measuring illumination in real-world conditions
    Howell, Colleen M.
    McCullough, Sara J.
    Doyle, Lesley
    Murphy, Marie H.
    Saunders, Kathryn J.
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2021, 41 (05) : 1048 - 1059
  • [30] Criterion validity and reliability of an instrumented mouthguard under pendulum impactor conditions
    Austin, Kieran
    Nicholas, Kieran Jai
    Jones, Christopher M.
    Loosemore, Mike
    SPORTS ENGINEERING, 2023, 26 (01)