A Geant4 simulation of X-ray emission for three-dimensional proton imaging of microscopic samples

被引:1
|
作者
Michelet, Claire [1 ]
Li, Zhuxin [1 ]
Jalenques, H. [1 ]
Incerti, Sebastien [1 ]
Barberet, Philippe [1 ]
Deves, Guillaume [1 ]
Delville, Marie-Helene [2 ]
Seznec, Herve [1 ]
机构
[1] Univ Bordeaux, UMR5797, CENBG, CNRS, F-33170 Gradignan, France
[2] Univ Bordeaux, UMR5026, ICMCB, CNRS, 87 Ave Dr A Schweitzer, F-33608 Pessac, France
关键词
Geant4; simulation; Proton computed tomography; PIXE tomography; Caenorhabditis elegans; PIXE-TOMOGRAPHY; SOFTWARE PACKAGE; RECONSTRUCTION; CELLS;
D O I
10.1016/j.ejmp.2021.12.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Proton computed micmtomography is a technique that reveals the inner content of microscopic samples. The density distribution of the material (in g.cm(-3)) is obtained from proton transmission tomography (STIM: Scanning Transmission Ion Microscopy) and the element content from X-ray emission tomography (PIXE: Particle Induced X-ray Emission). A precise quantification of chemical elements is difficult for thick samples, because of the variations of X-ray production cross-sections and of X-ray absorption. Both phenomena are at the origin of an attenuation of the measured X-ray spectra, which leads to an underestimation of the element content. Our aim is to quantify the accuracy of a specific correction method that we designed for thick samples. Methods: In this study, we describe how the 3D variations in the mass density were taken into account in the reconstruction code, in order to quantify the correction according to the position of the proton beam and the position and aperture angle of the X-ray detector. Moreover, we assess the accuracy of the reconstructed densities using Geant4 simulations on numerical phantoms, used as references. Results: The correction process was successfully applied and led, for the largest regions of interest (little affected by partial volume effects), to an accuracy <= 4% for phosphorus (compared to about 40% discrepancy without correction). Conclusion: This study demonstrates the accuracy of the correction method implemented in the tomographic reconstruction code for thick samples. It also points out some advantages offered by Geant4 simulations: i) they produce projection data that are totally independent of the inversion method used for the image reconstruction; ii) one or more physical processes (X-ray absorption, proton energy loss) can be artificially turned off, in order to precisely quantify the effect of the different phenomena involved in the attenuation of X-ray spectra.
引用
收藏
页码:85 / 93
页数:9
相关论文
共 50 条
  • [31] GEANT4 simulation study of over-response phenomenon of fiber x-ray sensor*
    Zhang, Bin
    Xie, Tian-Ci
    Qin, Zhuang
    Li, Hao-Peng
    Li, Song
    Zhao, Wen-Hui
    Chen, Zi-Yin
    Xu, Jun
    Lewis, Elfed
    Sun, Wei-Min
    CHINESE PHYSICS B, 2021, 30 (04)
  • [32] Three-dimensional X-ray imaging of macropore flow
    Schwenk, Anna Elena
    Jarvis, Nicholas
    Larsbo, Mats
    Koestel, John
    GEODERMA, 2023, 440
  • [33] Simulation of a Typical Wolter-I X-ray Telescope Using the Geant4 Toolkit
    Mehdi, Abbasian Motlagh
    Gohar, Rastegarzadeh
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2023, 66 (03) : 451 - 455
  • [34] Scatter in an uncollimated x-ray CT machine based on a Geant4 Monte Carlo simulation
    Wadeson, Nicola
    Morton, Edward
    Lionheart, William
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [35] ZnO nanowires in polycarbonate membrane as a high resolution X-ray detector (a Geant4 simulation)
    Taheri, Ali
    Saramad, Shahyar
    Setayeshi, Saeed
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 716 : 15 - 22
  • [36] Simulation of a Typical Wolter-I X-ray Telescope Using the Geant4 Toolkit
    Abbasian Motlagh Mehdi
    Rastegarzadeh Gohar
    Instruments and Experimental Techniques, 2023, 66 : 451 - 455
  • [37] Determination of molybdenum target parameters for transmission X-ray tube: A Geant4 simulation study
    Mohammad Mehdi Nasseri
    Pramana, 2019, 92
  • [38] GEANT4 simulation study of over-response phenomenon of fiber x-ray sensor
    张彬
    谢天赐
    秦壮
    李昊鹏
    李松
    赵文辉
    陈子印
    徐军
    Elfed Lewis
    孙伟民
    Chinese Physics B, 2021, 30 (04) : 732 - 738
  • [39] A Geant4 tool for edge-illumination X-ray phase-contrast imaging
    Brombal, L.
    Rigon, L.
    Arfelli, F.
    Menk, R. H.
    Brun, F.
    JOURNAL OF INSTRUMENTATION, 2022, 17 (01):
  • [40] Parameterization models for X-ray transition radiation in the GEANT4 package
    Apostolakis, J
    Giani, S
    Maire, M
    Bagulya, AV
    Grichine, VM
    COMPUTER PHYSICS COMMUNICATIONS, 2000, 132 (03) : 241 - 250