Comparison of Real-Time Two-Dimensional and Three-Dimensional Contrast-Enhanced Ultrasound to Quantify Flow in an In Vitro Model: A Feasibility Study

被引:4
|
作者
Ruan, Si-Min [1 ]
Zheng, Qiao [2 ]
Wang, Zhu [1 ]
Hu, Hang-Tong [1 ]
Chen, Li-Da [1 ]
Guo, Huan-Ling [1 ]
Xie, Xiao-Yan [1 ]
Lu, Ming-De [1 ,3 ]
Li, Wei [1 ]
Wang, Wei [1 ]
机构
[1] Sun Yat Sen Univ, Inst Diagnost & Intervent Ultrasound, Dept Med Ultrason, Ultrason Artificial Intelligence X Lab,Affiliated, Guangzhou, Guangdong, Peoples R China
[2] Sun Yat Sen Univ, Fetal Med Ctr, Dept Med Ultrason, Affiliated Hosp 1, Guangzhou, Guangdong, Peoples R China
[3] Sun Yat Sen Univ, Dept Hepatobiliary Surg, Affiliated Hosp 1, Guangzhou, Guangdong, Peoples R China
来源
MEDICAL SCIENCE MONITOR | 2019年 / 25卷
关键词
Contrast Media; Imaging; Three-Dimensional; In Vitro; Ultrasonography; FOCAL LIVER-LESIONS; QUANTIFICATION; SUPERIOR; US;
D O I
10.12659/MSM.919160
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background: This feasibility study aimed to compare real-time two-dimensional contrast-enhanced ultrasound (2D-CEUS) and three-dimensional contrast-enhanced ultrasound (3D-CEUS) to quantify flow in an in vitro model. Material/Methods: Five polyvinyl chloride (PVC) tubes were used for the perfusion models and used SonoVue ultrasound contrast agent with a perfusion volume ratio of 1: 2: 4: 8: 16. The contrast was injected at a constant speed to compare the raw quantitative data of 2D-CEUS and 3D-CEUS at angles of 0 degrees, 45 degrees, and 90 degrees. The coefficient of variation (CV) of the peak intensity (PI) in the model were compared and the correlations between weighted PI and perfusion volume were analyzed. Results: In the three angles used, real-time 3D-CEUS resulted in a more comprehensive view of the spatial relationships in the perfusion model. Using real-time 2D-CEUS, the mean CV was 0.92 +/- 0.36, and the mean CV in the real-time 3D-CEUS model was significantly less at 0.48 +/- 0.32 (p<0.001). Quantitative 3D-CEUS parameters showed a good correlation with those of 2D-CEUS with an r-value of 0.93 (p=0.02). The r-value of weighted PI and the perfusion ratio using 2D-CEUS was 0.66 (p=0.23) compared with values in 3D-CEUS of 0.84 (p=0.08). Conclusions: The combination of real-time 3D-CEUS and quantitative analysis identified the spatial distribution of the changes in angle in the model, which was less influenced by sectional planes, and was more representative of the perfusion volume when compared with 2D-CEUS. Quantitative real-time 3D-CEUS requires in vivo studies to evaluate the potential role in the clinical evaluation of vascular perfusion of malignant tumors.
引用
收藏
页码:10029 / 10035
页数:7
相关论文
共 50 条
  • [1] Feasibility of dynamic three-dimensional contrast-enhanced ultrasound in focal liver lesions: Image quality evaluation and correlation of quantification with two-dimensional contrast-enhanced ultrasound
    Cao, Jiaying
    Dong, Yi
    Fan, Peili
    Mao, Feng
    Wang, Wenping
    CLINICAL HEMORHEOLOGY AND MICROCIRCULATION, 2019, 72 (03) : 305 - 316
  • [2] Real-time three-dimensional color Doppler flow: Feasibility and initial quantitative comparison with two-dimensional methods
    Sugeng, L
    Weinert, L
    Spencer, KT
    Furlong, KT
    Lang, RM
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (05) : 356A - 356A
  • [3] Preliminary study of real-time three-dimensional contrast-enhanced ultrasound of sentinel lymph nodes in breast cancer
    Ziyue Hu
    Xueqing Cheng
    Juan Li
    Jingzhen Jiang
    Zirui Jiang
    Hui Li
    Tingting Li
    Zhenqi Zhang
    Bo Tan
    Man Lu
    European Radiology, 2020, 30 : 1426 - 1435
  • [4] Preliminary study of real-time three-dimensional contrast-enhanced ultrasound of sentinel lymph nodes in breast cancer
    Hu, Ziyue
    Cheng, Xueqing
    Li, Juan
    Jiang, Jingzhen
    Jiang, Zirui
    Li, Hui
    Li, Tingting
    Zhang, Zhenqi
    Tan, Bo
    Lu, Man
    EUROPEAN RADIOLOGY, 2020, 30 (03) : 1426 - 1435
  • [5] Real-Time Three-Dimensional Flexible Needle Tracking using Two-Dimensional Ultrasound
    Vrooijink, Gustaaf J.
    Abayazid, Momen
    Misra, Sarthak
    2013 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION (ICRA), 2013, : 1688 - 1693
  • [6] Quantitative assessment of left ventricular size and function in cardiac transplant recipients: Side-by-side comparison of real time two-dimensional echocardiography, contrast-enhanced two-dimensional echocardiography, three-dimensional echocardiography, and contrast-enhanced three-dimensional echocardiography as compared to magnetic resonance imaging
    Rodriguez-Manero, Moises
    Azcarate-Aguero, Pedro
    Kreidieh, Bahij
    Teresa Alvez, Maria
    Martinez-Monzonis, Amparo
    Diaz-Dorronsoro, Agnes
    Cid-Menendez, Adrian
    Ramon Gonzalez-Juanatey, Jose
    Barba-Cosials, Joaquin
    Rabago, Gregorio
    Bastarrika, Gorka
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2019, 36 (02): : 306 - 311
  • [7] Contrast enhanced real-time three-dimensional echocardiography, is it feasible?
    Kapetenakis, S
    Murray, O
    Rance, K
    Platts, D
    Monaghan, M
    EUROPEAN HEART JOURNAL, 2003, 24 : 700 - 700
  • [9] Double contrast-enhanced two-dimensional and three-dimensional ultrasonography for evaluation of gastric lesions
    Shi, Hong
    Yu, Xiu-Hua
    Guo, Xin-Zhang
    Guo, Yuan
    Zhang, Hong
    Qian, Bin
    Wei, Zhang-Rui
    Li, Li
    Wang, Xian-Chen
    Kong, Zi-Xiang
    WORLD JOURNAL OF GASTROENTEROLOGY, 2012, 18 (31) : 4136 - 4144
  • [10] Magnetic resonance colonography: Comparison of contrast-enhanced three-dimensional vibe with two-dimensional FISP sequences - Preliminary experience
    Lauenstein, TC
    Ajaj, W
    Kuehle, CA
    Goehde, SC
    Schlosser, TW
    Ruehm, SG
    INVESTIGATIVE RADIOLOGY, 2005, 40 (02) : 89 - 96