A comparison of the different features of quadripolar left ventricular pacing leads to deliver cardiac resynchronization therapy

被引:5
|
作者
Antoniadis, Antonios P. [1 ]
Behar, Jonathan M.
Sieniewicz, Ben
Gould, Justin
Niederer, Steven
Rinaldi, Christopher A.
机构
[1] St Thomas Hosp, Guys & St Thomas NHS Fdn Trust, Cardiovasc Dept, 6th Floor East Wing,Westminster Bridge Rd, London SE1 7EH, England
基金
英国工程与自然科学研究理事会;
关键词
Quadripolar left ventricular leads; cardiac resynchronization therapy; heart failure; biventricular pacing; multisite pacing; PHRENIC-NERVE STIMULATION; CHRONIC HEART-FAILURE; ACUTE HEMODYNAMIC-RESPONSE; FOLLOW-UP; IMPLANTABLE DEFIBRILLATOR; MECHANICAL DYSSYNCHRONY; MULTISITE; IMPROVES; MULTICENTER; EXPERIENCE;
D O I
10.1080/17434440.2017.1369404
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Introduction: Cardiac Resynchronization therapy (CRT) improves the quality of life and reduces morbidity and mortality of certain patients with heart failure. However, not all patients respond positively after CRT and about one third of cases do not experience benefit. Suboptimal biventricular pacing may account for this and quadripolar left ventricular (LV) leads have emerged in the last years to address issues relating to inadequate delivery of CRT. Areas covered: This review article concisely summarizes the main technical characteristics of the quadripolar LV leads either currently available in the market today or under final stages of development. Focus is given in recent advancements in the area and challenging aspects and controversies, future implications as well as opportunities for further development. Expert commentary: Quadripolar LV pacing leads have now become the standard of care in CRT. Currently a multitude of lead options is available to the clinician. The selection process of the most appropriate lead is far from the 'one size fits all' concept. Further development of quadripolar LV leads is currently ongoing and it is anticipated to contribute towards the release of more technologically advantageous leads which will enable the delivery of optimal CRT therapy with the lowest rate of complications.
引用
收藏
页码:697 / 706
页数:10
相关论文
共 50 条
  • [21] Practical approach to implanting left ventricular pacing leads for cardiac resynchronization
    León, AR
    Delurgio, DB
    Mera, F
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2005, 16 (01) : 100 - 105
  • [22] Can We Use the Intrinsic Left Ventricular Delay (QLV) to Optimize the Pacing Configuration for Cardiac Resynchronization Therapy With a Quadripolar Left Ventricular Lead?
    van Everdingen, Wouter M.
    Zweerink, Alwin
    Cramer, Maarten J.
    Doevendans, Pieter A.
    Uyen Chau Nguyen
    van Rossum, Albert C.
    Prinzen, Frits W.
    Vernooy, Kevin
    Allaart, Cornelis P.
    Meine, Mathias
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2018, 11 (03):
  • [23] CAN WE USE THE INTRINSIC LEFT VENTRICULAR DELAY (QLV) TO OPTIMIZE THE PACING CONFIGURATION FOR CARDIAC RESYNCHRONIZATION THERAPY WITH A QUADRIPOLAR LEFT VENTRICULAR LEAD?
    van Everdingen, Wouter
    Zweerink, Alwin
    Cramer, Maarten Jan
    Doevendans, Pieter
    Uyen Chau Nguyen
    van Rossum, Albert
    Prinzen, Frits
    Vernooy, Kevin
    Allaart, Cornelis
    Meine, Mathias
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2018, 71 (11) : 945 - 945
  • [24] Repetitive optimizing left ventricular pacing configurations with quadripolar leads improves response to cardiac resynchronization therapy A single-center randomized clinical trial
    Gu, Min
    Jin, Han
    Hua, Wei
    Fan, Xiao-Han
    Ding, Li-Gang
    Wang, Jing
    Niu, Hong-Xia
    Cai, Chi
    Zhang, Shu
    MEDICINE, 2017, 96 (37)
  • [25] Stent-stabilization of left ventricular pacing leads for cardiac resynchronization therapy: A promising concept?
    Ermis, Cengiz
    Benditt, David G.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2007, 18 (03) : 308 - 309
  • [26] Results of patients with cardiac resynchronization therapy using a quadripolar left ventricular lead
    D'Ortencio, Alfredo
    Bevacqua, Raul J.
    INSUFICIENCIA CARDIACA, 2013, 8 (03) : 142 - 148
  • [27] Multisite pacing via a quadripolar lead for cardiac resynchronization therapy
    Alexandre Bodin
    Arnaud Bisson
    Clémentine Andre
    Bertrand Pierre
    Laurent Fauchier
    Dominique Babuty
    Nicolas Clementy
    Journal of Interventional Cardiac Electrophysiology, 2019, 56 : 117 - 125
  • [28] Optimizing left ventricular pacing configurations for cardiac resynchronization therapy
    Lee, N
    Disch, D
    Tocco, P
    Botteron, GW
    Craddock, H
    Huelsing, DJ
    JOURNAL OF CARDIAC FAILURE, 2005, 11 (06) : S197 - S197
  • [29] Evaluation of effective left ventricular pacing in cardiac resynchronization therapy
    Mendez Zurita, F. Francisco
    Santos, J. Quispe
    Font, E. Rodriguez
    Martin, C. Alonso
    Ramos, J. Guerra
    Garcia, B. Campos
    Weidmann, Z. Moreno
    Bravo, E. Gilces
    De Diego, I. Ramirez
    Martin, L. Sanchez
    Osorio, M. Grande
    Ramirez, D. Majo
    Jaen, E. Paz
    Nieves, A. Velasco
    Prat, X. Vinolas
    EUROPEAN JOURNAL OF HEART FAILURE, 2024, 26 : 31 - 31
  • [30] Occurrence of simultaneous cathodal-anodal capture with left ventricular quadripolar leads for cardiac resynchronization therapy: an electrocardiogram evaluation
    Occhetta, Eraldo
    Dell'Era, Gabriele
    Giubertoni, Ailia
    Magnani, Andrea
    Rametta, Francesco
    Blandino, Alessandro
    Magnano, Vincenzo
    Malacrida, Maurizio
    Marino, Paolo
    EUROPACE, 2017, 19 (04): : 596 - 601