Evaluation of a treatment planning system developed for clinical boron neutron capture therapy and validation against an independent Monte Carlo dose calculation system

被引:42
|
作者
Hu, Naonori [1 ,2 ]
Tanaka, Hiroki [2 ]
Kakino, Ryo [1 ]
Yoshikawa, Syuushi [1 ]
Miyao, Mamoru [3 ]
Akita, Kazuhiko [1 ]
Isohashi, Kayako [1 ]
Aihara, Teruhito [1 ]
Nihei, Keiji [1 ,4 ]
Ono, Koji [1 ]
机构
[1] Osaka Med & Pharmaceut Univ, Kansai BNCT Med Ctr, Osaka Fu Takatsuki Shi Daigakumachi 2-7, Takatsuki, Osaka, Japan
[2] Kyoto Univ, Inst Integrated Radiat & Nucl Sci, Kyoto, Japan
[3] Osaka Med & Pharmaceut Univ Hosp, Cent Dept Radiol, Takatsuki, Osaka, Japan
[4] Osaka Med & Pharmaceut Univ Hosp, Dept Radiat Oncol, Takatsuki, Osaka, Japan
关键词
Boron neutron capture therapy; Treatment planning system; Commissioning; Monte Carlo simulation; COMPUTATIONAL DOSIMETRY; VERIFICATION;
D O I
10.1186/s13014-021-01968-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Boron neutron capture therapy (BNCT) for the treatment of unresectable, locally advanced, and recurrent carcinoma of the head and neck cancer has been approved by the Japanese government for reimbursement under the national health insurance as of June 2020. A new treatment planning system for clinical BNCT has been developed by Sumitomo Heavy Industries, Ltd. (Sumitomo), NeuCure (R) Dose Engine(.) To safely implement this system for clinical use, the simulated neutron flux and gamma ray dose rate inside a water phantom was compared against experimental measurements. Furthermore, to validate and verify the new planning system, the dose distribution inside an anthropomorphic head phantom was compared against a BNCT treatment planning system SERA and an in-house developed Monte Carlo dose calculation program. The simulated results closely matched the experimental results, within 5% for the thermal neutron flux and 10% for the gamma ray dose rate. The dose distribution inside the head phantom closely matched with SERA and the in-house developed dose calculation program, within 3% for the tumour and a difference of 0.3 Gy(w) for the brain.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Evaluation of a treatment planning system developed for clinical boron neutron capture therapy and validation against an independent Monte Carlo dose calculation system
    Naonori Hu
    Hiroki Tanaka
    Ryo Kakino
    Syuushi Yoshikawa
    Mamoru Miyao
    Kazuhiko Akita
    Kayako Isohashi
    Teruhito Aihara
    Keiji Nihei
    Koji Ono
    Radiation Oncology, 16
  • [2] Development of a New Monte-Carlo Treatment Planning System for Boron Neutron Capture Therapy
    Kumada, Hiroaki
    Nakamura, Takemi
    Matsumura, Akira
    Ono, Koji
    PROCEEDINGS OF THE 17TH INTERNATIONAL CONFERENCE ON NUCLEAR ENGINEERING, VOL 2, 2009, : 823 - 829
  • [3] Development of Monte Carlo based real-time treatment planning system with fast calculation algorithm for boron neutron capture therapy
    Takada, Kenta
    Kumada, Hiroaki
    Liem, Peng Hong
    Sakurai, Hideyuki
    Sakae, Takeji
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (12): : 1846 - 1851
  • [4] Whole-body dose evaluation with an adaptive treatment planning system for boron neutron capture therapy
    Takada, Kenta
    Kumada, Hiroaki
    Isobe, Tomonori
    Terunuma, Toshiyuki
    Kamizawa, Satoshi
    Sakurai, Hideyuki
    Sakae, Takeji
    Matsumura, Akira
    RADIATION PROTECTION DOSIMETRY, 2015, 167 (04) : 584 - 590
  • [5] Clinical implementation and validation of a commercial Monte Carlo dose calculation system
    Chuang, C
    Curran, B
    Verhey, L
    MEDICAL PHYSICS, 2003, 30 (06) : 1453 - 1453
  • [6] Monte Carlo based treatment planning systems for Boron Neutron Capture Therapy in Petten, The Netherlands
    Nievaart, V. A.
    Daquino, G. G.
    Moss, R. L.
    FIRST EUROPEAN WORKSHOP ON MONTE CARLO TREATMENT PLANNING, 2007, 74
  • [7] Development and evaluation of dose calculation algorithm with a combination of Monte Carlo and point-kernel methods for boron neutron capture therapy
    Nojiri, Mai
    Takata, Takushi
    Hu, Naonori
    Sakurai, Yoshinori
    Suzuki, Minoru
    Tanaka, Hiroki
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2023, 9 (03)
  • [8] Evaluation of the Monte Carlo Dose Calculation Engine of Eclipse Treatment Planning System for Electron Beams
    Kamta, G. Lagmago
    Reniers, B.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [9] MONTE-CARLO DOSIMETRY CALCULATION FOR BORON NEUTRON-CAPTURE THERAPY IN TREATMENT OF BRAIN TUMORS
    DEUTSCH, OL
    MURRAY, BW
    NUCLEAR TECHNOLOGY, 1975, 26 (03) : 320 - 339
  • [10] Dose calculation using Monte Carlo algorithm of Monaco treatment planning system
    Li, J.
    Doemer, A.
    Harrison, A. S.
    Galvin, J. M.
    Xiao, Y.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S674 - S674