Effect of central corneal thickness on Goldmann applanation tonometry measures - a different result with different pachymeters

被引:16
|
作者
Doughty, Michael J. [1 ]
Jonuscheit, Sven [1 ]
机构
[1] Glasgow Caledonian Univ, Dept Vis Sci, Glasgow G4 0BA, Lanark, Scotland
关键词
central corneal thickness; ultrasound pachymetry; orbscan; Goldmann tonometry; normal corneas;
D O I
10.1007/s00417-007-0601-x
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
To assess whether the use of Orbscan pachymetry, as an alternative to ultrasound pachymetry, had an effect on the IOP-CCT relationship for Goldmann tonometry in normal human eyes. The right eye of 50 healthy subjects, aged 19 to 82 years, were assessed by Orbscan (three scans), and then central corneal thickness (CCT) measured by ultrasound pachymetry (three readings) and Goldmann tonometry performed (three readings). Goldmann tonometry indicated average IOP values from 9.7 to 20 mm Hg (group mean 14.3 +/- 2.5 mm Hg) that were highly highly correlated with ultrasound pachymetry (r=0.608, p < 0.001). Orbscan central zone pachymetry (2-mm sample zone) yielded higher values than ultrasound, averaging 0.591 +/- 0.044 mm compared to 0.523 +/- 0.037 mm (p < 0.001), with the differences between the instruments being proportional to the average thickness (p < 0.001; r=0.461). Notwithstanding, the Goldmann tonometry values were still highly correlated with the Orbscan central zone thickness data (r=0.595, p < 0.001); but, due to the difference in the thickness data generated by the two pachymeters, the absolute slope of the IOP-CCT relationship was slightly less if Orbscan measures were used, with or without use of the acoustic factor. Orbscan measures of the thickness of a central corneal zone can be used to assess the impact of central corneal thickness on Goldmann tonometry data.
引用
收藏
页码:1603 / 1610
页数:8
相关论文
共 50 条
  • [31] Corneal thickness and Intraocular pressure After DSAEK: comparison of Dynamic Contour tonometry with Goldmann Applanation tonometry
    Borges, Barbara
    Vicente, Andre
    Anjos, Rita
    Vieira, Luisa
    Santos, Arnaldo Dias
    Ferreira, Joana
    Maduro, Vitor
    Alves, Nuno
    Feijao, Joao
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [32] The Influence of Central Corneal Thickness on Intraocular Pressure Measured by Goldmann Applanation Tonometry Among Selected Ethiopian Communities
    Gelaw, Yeshigeta
    Kollmann, Martin
    Irungu, Njuguna M.
    Ilako, Dunera R.
    JOURNAL OF GLAUCOMA, 2010, 19 (08) : 514 - 518
  • [33] GOLDMANN APPLANATION TONOMETRY IN PATIENTS WITH REGULAR CORNEAL ASTIGMATISM
    HOLLADAY, JT
    ALLISON, ME
    PRAGER, TC
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 1983, 96 (01) : 90 - 93
  • [34] Effects of corneal thickness, curvature, astigmatism and direction of gaze on Goldmann applanation tonometry readings
    Rask, G
    Behndig, A
    OPHTHALMIC RESEARCH, 2006, 38 (01) : 49 - 55
  • [35] The Influence of Central Corneal Thickness and Corneal Curvature Radius on The Intraocular Pressure as Measured By Different Tonometers: Noncontact and Goldmann Applanation Tonometers
    Harada, Yukinori
    Hirose, Naofumi
    Kubota, Toshiaki
    Tawara, Akihiko
    JOURNAL OF GLAUCOMA, 2008, 17 (08) : 619 - 625
  • [36] Corneal thickness and its effect on applanation tonometry in Japanese
    Takamura, Y
    Nakamura, S
    Morikubo, S
    Tanaka, M
    Kawai, R
    Takahashi, Y
    Akagi, Y
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1999, 40 (04) : S281 - S281
  • [37] Corneal thickness and applanation tonometry readings
    Anders Heijl
    Boel Bengtsson
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2001, 239 : 462 - 462
  • [38] IMPORTANCE OF CORNEAL THICKNESS IN APPLANATION TONOMETRY
    EHLERS, N
    BRAMSEN, T
    ACTA OPHTHALMOLOGICA, 1975, : 32 - 32
  • [39] The influence of corneal density and thickness on tonometry measurement with goldmann applanation, non-contact and iCare tonometry methods
    Ahmed Lubbad
    Irene Oluwatoba-Popoola
    Melanie Haar
    Carsten Framme
    Anna Bajor
    International Ophthalmology, 2022, 42 : 2167 - 2174
  • [40] Corneal thickness and applanation tonometry readings
    Heijl, A
    Bengtsson, B
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2001, 239 (06) : 462 - 462