A comparative study of solid and liquid inner contact paroxetine hydrochloride ion-selective electrode membranes

被引:0
|
作者
Khalil, Mohamed Magdy [1 ]
Issa, Yousry Mostfa [2 ]
Mohamed, Ali Gaber [1 ]
机构
[1] Beni Suef Univ, Fac Sci, Dept Chem, Bani Suwayf, Egypt
[2] Cairo Univ, Fac Sci, Dept Chem, Giza, Egypt
关键词
Ion-selective electrodes; Solid-contact ionselective electrode; PVC membrane electrode; Paroxetine hydrochloride; SEROTONIN REUPTAKE INHIBITORS; MASS-SPECTROMETRY; HUMAN PLASMA; POTENTIOMETRIC DETERMINATION; GAS-CHROMATOGRAPHY; ACTIVE METABOLITES; PHASE EXTRACTION; NEUTRAL CARRIER; UV DETECTION; ANTIDEPRESSANTS;
D O I
10.1007/s13738-015-0638-4
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A comparative study was made between two designs of paroxetine-selective electrodes: a polyvinyl chloride membrane (liquid inner contact) called electrodes PME1 and PME2 and a solid contact called electrodes CWE1 and CWE2 based on paroxetine-phosphotungstate and paroxetine-phosphomolybdate as ion exchangers. The four electrodes, PME1, PME2, CWE1 and CWE2, show linearity over the concentration range from 1 x 10(-5) to 1 x 10(-2) M, with slopes of 56.7, 54.4, 59.8 and 55.3 mV/decade, meanwhile the limits of detection were 2.5 x 10(-6), 4 x 10(-6), 5.6 x 10(-6) and 6.2 x 10(-6) M, respectively. PME1 and PME2 showed better limit of detection than electrodes CWE1 and CWE2. The present electrodes show clear discrimination of Prx.HCl from several inorganic, and organic species. The sensors were applied efficiently for determination of Prx.HCl in its pharmaceutical preparations using standard addition and the calibration curve methods.
引用
收藏
页码:1637 / 1648
页数:12
相关论文
共 50 条
  • [1] A comparative study of solid and liquid inner contact paroxetine hydrochloride ion-selective electrode membranes
    Mohamed Magdy Khalil
    Yousry Mostfa Issa
    Ali Gaber Mohamed
    Journal of the Iranian Chemical Society, 2015, 12 : 1637 - 1648
  • [2] A comparative study of liquid and solid inner contact roxatidine acetate ion-selective electrode membranes
    Abd El-Rahman, Mohamed K.
    Zaazaa, Hala E.
    Abbas, Samah S.
    El-Zeany, Badr
    EL-Sherif, Zeinab A.
    EL-Haddad, Dalia A.
    CHINESE CHEMICAL LETTERS, 2015, 26 (06) : 714 - 720
  • [3] A comparative study of solid and liquid inner contact benzalkonium chloride ion-selective electrode membranes
    Abu Shawish, Hazem M.
    Khedr, Abdalla M.
    Abed-Almonem, Khalid I.
    Gaber, M.
    TALANTA, 2012, 101 : 211 - 219
  • [4] A comparative study of liquid and solid inner contact roxatidine acetate ion-selective electrode membranes
    Mohamed K.Abd El-Rahman
    Hala E.Zaazaa
    Samah S.Abbas
    Badr El-Zeany
    Zeinab A.EL-Sherif
    Dalia A.EL-Haddad
    Chinese Chemical Letters, 2015, 26 (06) : 714 - 720
  • [5] Green Ion Selective Electrode Potentiometric Application for the Determination of Cinchocaine Hydrochloride in Presence of Its Degradation Products and Betamethasone Valerate: A Comparative Study of Liquid and Solid Inner Contact Ion-Selective Electrode Membranes
    Rezk, Mamdouh R.
    Fayed, Ahmed S.
    Marzouk, Hoda M.
    Abbas, Samah S.
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2017, 164 (09) : H628 - H634
  • [6] A Comparative Study of Cadmium Ion-Selective Electrodes with Solid and Liquid Inner Contact
    Wardak, Cecylia
    ELECTROANALYSIS, 2012, 24 (01) : 85 - 90
  • [7] Uranyl ion-selective electrode with solid contact
    Pietrzak, Karolina
    Wardak, Cecylia
    PROCEEDINGS OF THE 16TH INTERNATIONAL STUDENTS CONFERENCE MODERN ANALYTICAL CHEMISTRY, 2020, : 45 - 50
  • [8] Solid-contact bromide ion-selective electrode
    Stefanova, OK
    Alagova, ZS
    Khripun, GA
    RUSSIAN JOURNAL OF APPLIED CHEMISTRY, 1995, 68 (08) : 1212 - 1213
  • [9] Cobalt(II) ion-selective electrode with solid contact
    Wardak, Cecylia
    CENTRAL EUROPEAN JOURNAL OF CHEMISTRY, 2008, 6 (04): : 607 - 612
  • [10] Solid-contact ion-selective electrode with stable internal electrode
    Vamvakaki, M