On the vectorial calibration of a vibrating sample magnetometer for thin film measurements

被引:8
|
作者
Bolhuis, T
Abelmann, L
Lodder, JC
Samwel, EO
机构
[1] Univ Twente, MESA Res Inst, NL-7500 AE Enschede, Netherlands
[2] ADE Technol Inc, Digital Measurement Syst Div, Newton, MA 02166 USA
关键词
calibration; vector VSM; cross talk; thin film; demagnetisation; ME-tape;
D O I
10.1016/S0304-8853(98)00451-X
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A calibration method for the biaxial vector Vibrating Sample Magnetometer (VSM) is proposed to reduce the so-called cross talk' error, often observed during angular measurements on magnetic thin films. The coil system of the biaxial vector VSM consists of two sets of coils, one for each of the two co-ordinate axes. The method takes into account that each coil set is sensitive to all components of the magnetisation vector and moreover that the sensitivity to the components parallel and perpendicular to the film plane can be different. A correction for demagnetisation should be used to compensate the fact that the sample used for the calibration cannot be fully saturated with an applied field at angles close to the normal of the film plane. Curve-fitting on the sensitivity curves is used to reduce noise. These procedures result in a calibration which reduces the cross talk by a factor 4-8 depending on the coil configuration and the size and shape of the samples used. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:332 / 336
页数:5
相关论文
共 50 条
  • [41] AN OPTICAL HEATING STAGE FOR A VIBRATING SAMPLE MAGNETOMETER
    CAREY, R
    NEWMAN, DM
    SNELLING, J
    THOMAS, BWJ
    MEASUREMENT SCIENCE AND TECHNOLOGY, 1992, 3 (04) : 424 - 425
  • [42] NONUNIFORM MAGNETIZATION AND VIBRATING-SAMPLE MAGNETOMETER
    GUY, CN
    HOWARTH, W
    JOURNAL OF PHYSICS C-SOLID STATE PHYSICS, 1978, 11 (08): : 1635 - 1650
  • [43] IMPROVED HELIUM CRYOSTAT FOR A VIBRATING SAMPLE MAGNETOMETER
    LANDEE, CP
    GREENEY, RE
    LAMAS, AC
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1987, 58 (10): : 1957 - 1958
  • [44] THE VIBRATING-SAMPLE MAGNETOMETER AND PREISACH DIAGRAM
    ZELINKA, T
    HEJDA, P
    KROPACEK, V
    PHYSICS OF THE EARTH AND PLANETARY INTERIORS, 1987, 46 (1-3) : 241 - 246
  • [45] LOW-FREQUENCY VIBRATING SAMPLE MAGNETOMETER
    JOHANSSON, T
    NIELSEN, KG
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1976, 9 (10): : 852 - 854
  • [46] VERSATILE AND SENSITIVE VIBRATING-SAMPLE MAGNETOMETER
    FONER, S
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1959, 30 (07): : 548 - 557
  • [47] ELECTROSTATIC DRIVEN VIBRATING SAMPLE MAGNETOMETER ESPECIALLY SUITABLE FOR MAGNETIC-ANISOTROPY MEASUREMENTS
    NIELSEN, OV
    ZAITZEV, VI
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1973, 6 (10): : 1022 - 1024
  • [48] Computational modelling of a triaxial vibrating sample magnetometer
    Rodriguez, Leo
    Sapkota, Arjun
    Alvarado, Jonathan
    Tate, Jitendra S.
    Geerts, Wilhelmus J.
    AIP ADVANCES, 2024, 14 (02)
  • [49] IMAGE EFFECTS IN VIBRATING SAMPLE MAGNETOMETER SYSTEMS
    STONER, RE
    HERBERT, RH
    SILL, LR
    JOURNAL OF APPLIED PHYSICS, 1970, 41 (09) : 3706 - &
  • [50] A simple vibrating sample magnetometer for macroscopic samples
    Lopez-Dominguez, V.
    Quesada, A.
    Guzman-Minguez, J. C.
    Moreno, L.
    Lere, M.
    Spottorno, J.
    Giacomone, F.
    Fernandez, J. F.
    Hernando, A.
    Garcia, M. A.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2018, 89 (03):