Detection efficiency evaluation for low energy of a NaI(Tl) scintillation detector

被引:2
|
作者
Maeng, Seongjin [1 ]
Lee, Sang Hoon [1 ,2 ]
Park, Seong Jin [1 ,3 ]
Choi, Woo Cheol [4 ,5 ]
机构
[1] Kyungpook Natl Univ, Sch Energy Engn, Daegu 41566, South Korea
[2] Kyungpook Natl Univ, Radiat Sci Res Inst, Daegu 41566, South Korea
[3] Korea Fdn Nucl Safety, Seongnam 13493, Gyeonggi Do, South Korea
[4] Daegu Catholic Univ, Dept Radiol, Gyongsan 38430, Gyeongbuk, South Korea
[5] BRNC, Seoul 08507, South Korea
关键词
NaI(Tl) scintillation detector; X-ray escape; Low energy ragne; Detection efficiency; COINCIDENCE-SUMMING CORRECTION; RAY;
D O I
10.1016/j.radphyschem.2022.110325
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
To improve the characterization of the detection efficiency of a NaI(Tl) scintillation detector in the low energy range, measurements and Monte Carlo simulations were performed with a certified reference material (CRM) and disk sources. The detection efficiency as modeled. In the MCNP simulation, the detection efficiency dropped as expected at 33.17 keV. The X-ray escape correction factors of a Tl-204 disk source and MCNP simulations were compared, a factor of the Tl-204 was similar. Moreover, Correcting the X-ray escape and using fewer sources than the CRM resulted in better detection efficiency for the NaI(Tl) detector in the low energy range.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] More on the scintillation response of NaI(Tl)
    Valentine, JD
    Rooney, BD
    Dorenbos, P
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) : 1750 - 1756
  • [32] Scintillation process in NaI(Tl): Comparison with scintillation models
    Kubota, S
    Shiraishi, F
    Takami, Y
    JOURNAL OF THE PHYSICAL SOCIETY OF JAPAN, 2000, 69 (10) : 3435 - 3440
  • [33] CHARACTERISTICS OF A NAI(TL) SCINTILLATION CHAMBER
    BRUNINX, E
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1975, 26 (6-7): : 429 - 430
  • [34] CONTRIBUTION OF NAI(TL) CRYSTAL TO TOTAL LINEWIDTH OF NAI(TL) SCINTILLATION COUNTERS
    NARAYAN, GH
    PRESCOTT, JR
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1968, NS15 (03) : 162 - +
  • [35] Improved NaI(Tl) scintillation detectors
    Andryushchenko, LA
    Grinev, BV
    Litichevskii, AM
    Udovichenko, LV
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1997, 40 (01) : 59 - 63
  • [36] Improved NaI(Tl) Scintillation Detectors
    Instrum Exp Tech, (59):
  • [37] More on the scintillation response of NaI(Tl)
    Valentine, J.D.
    Rooney, B.D.
    Dorenbos, P.
    IEEE Transactions on Nuclear Science, 1998, 45 (3 pt 4): : 1750 - 1756
  • [38] Comparison of 3/8" and 5/8" NaI(Tl)-scintillation crystal (NaI) efficiency for different radioisotopes
    Happel, C.
    Groothuis, M.
    Mader, A.
    Bockisch, B.
    Gruenwald, F.
    Kranert, W. T.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 : S514 - S515
  • [39] DETECTION OF X-RAYS BY MEANS OF NAI(TL) SCINTILLATION COUNTERS
    WEST, HI
    MEYERHOF, WE
    HOFSTADTER, R
    PHYSICAL REVIEW, 1951, 81 (01): : 141 - 142
  • [40] Radioactive source strength effect on gamma radiation monitoring with a NaI (Tl) scintillation detector
    Yeboah, E.
    Yakovlev, G. A.
    Smirnov, S. V.
    Yakovleva, V. S.
    JOURNAL OF INSTRUMENTATION, 2022, 17 (05)