Low-Dose 3D Rotational Angiography in Measuring the Size of Intracranial Aneurysm: In Vitro Feasibility Study Using Aneurysm Phantom

被引:2
|
作者
Ki, Hee Jong [1 ]
Kim, Bum-Soo [2 ]
Kim, Jun-Ki
Choi, Jai Ho [1 ]
Shin, Yong Sam [1 ]
Choi, Yangsean [2 ]
Shin, Na-Young [2 ]
Jang, Jinhee [2 ]
Ahn, Kook-jin [2 ]
机构
[1] Catholic Univ Korea, Seoul St Marys Hosp, Dept Neurosurg, Seoul, South Korea
[2] Catholic Univ Korea, Seoul St Marys Hosp, Dept Radiol, Seoul, South Korea
关键词
Cerebral angiography; Radiation dosage; Intracranial aneurysm;
D O I
10.5469/neuroint.2020.00437
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Purpose: Three-dimensional (3D) measurement of intracranial aneurysms is important in planning endovascular treatment, and 3D rotational angiography (RA) is effective in accurate measurement. The purpose of this study was to evaluate the feasibility of low dose 3D RA (5 seconds 0.10 mu Gy/frame) in measuring an intracranial aneurysm using an in vitro phantom. Materials and Methods: We investigated an in vitro 3D phantom of an intracranial aneurysm with 10 acquisitions of 3D RA with a conventional dose (5 seconds 0.36 mu Gy/frame) and 10 acquisitions with a low-dose (5 seconds 0.10 mu Gy/frame). 3D size and neck diameters of the aneurysm were measured and compared between the 2 groups (conventional and low-dose) using noninferiority statistics. Results: The aneurysm measurements were well-correlated between the 2 readers, and noninferiority in the measurement of aneurysmal size of low-dose 3D RA was demonstrated, as the upper margin of the 1-sided 97.5% confidence interval did not cross the pre-defined noninferiority margin of 0.2 mm by the 2 readers. Conclusion: Low-dose (5 seconds 0.10 mu Gy/frame) cerebral 3D RA is technically feasible and not inferior in in vitro 3D measurement of an intracranial aneurysm. Thus, low-dose 3D RA is promising and needs further evaluation for its clinical utility in the planning of endovascular treatment of an intracranial aneurysm.
引用
收藏
页码:59 / 63
页数:5
相关论文
共 50 条
  • [21] Utility of 3D rotational angiography road map in flow diverter deployment in a distal dissecting MCA aneurysm
    Rustemi, Oriela
    Di Clemente, Loris
    Raneri, Fabio
    Volpin, Lorenzo
    Iannucci, Giuseppe
    NEUROSURGICAL FOCUS, 2019, 46
  • [22] Comparison of Image Enlargement according to 3D Reconstruction in a CT Scan: Using an Aneurysm Phantom
    Lee, Seong-yong
    Kim, Ham-Gyum
    Kim, Hwa-Sun
    Choi, Jae-Ho
    Cho, Jae-Hwan
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2018, 72 (07) : 805 - 810
  • [23] 3D manufacturing of intracranial aneurysm biomodels for flow visualizations: Low cost fabrication processes
    Souza, A.
    Souza, M. S.
    Pinho, D.
    Agujetas, R.
    Ferrera, C.
    Lima, R.
    Puga, H.
    Ribeiro, J.
    MECHANICS RESEARCH COMMUNICATIONS, 2020, 107
  • [24] Comparison of Image Enlargement according to 3D Reconstruction in a CT Scan: Using an Aneurysm Phantom
    Seong-yong Lee
    Ham-Gyum Kim
    Hwa-Sun Kim
    Jae-Ho Choi
    Jae-Hwan Cho
    Journal of the Korean Physical Society, 2018, 72 : 805 - 810
  • [25] Intracranial aneurysm segmentation in 3D CT angiography: Method and quantitative validation with and without prior noise filtering
    Firouzian, Azadeh
    Manniesing, Rashindra
    Flach, Zwenneke H.
    Risselada, Roelof
    van Kooten, Fop
    Sturkenboom, Miriam C. J. M.
    van der Lugt, Aad
    Niessen, Wiro J.
    EUROPEAN JOURNAL OF RADIOLOGY, 2011, 79 (02) : 299 - 304
  • [26] Shape Trumps Size: Image-Based Morphological Analysis Reveals That the 3D Shape Discriminates Intracranial Aneurysm Disease Status Better Than Aneurysm Size
    Juchler, Norman
    Schilling, Sabine
    Bijlenga, Philippe
    Kurtcuoglu, Vartan
    Hirsch, Sven
    FRONTIERS IN NEUROLOGY, 2022, 13
  • [27] 3D hindfoot alignment measurements based on low-dose biplanar radiographs: a clinical feasibility study
    Andrea B. Rosskopf
    Reto Sutter
    Christian W. A. Pfirrmann
    Florian M. Buck
    Skeletal Radiology, 2019, 48 : 707 - 712
  • [28] 3D hindfoot alignment measurements based on low-dose biplanar radiographs: a clinical feasibility study
    Rosskopf, Andrea B.
    Sutter, Reto
    Pfirrmann, Christian W. A.
    Buck, Florian M.
    SKELETAL RADIOLOGY, 2019, 48 (05) : 707 - 712
  • [29] Comparison of Cerebral Aneurysm Volumes as Determined by Digitally Measured 3D Rotational Angiography and Approximation from Three Diameters
    Sadato, A.
    Hayakawa, M.
    Tanaka, T.
    Hirose, Y.
    INTERVENTIONAL NEURORADIOLOGY, 2011, 17 (02): : 154 - 158
  • [30] Dose comparison of classical 2-plane DSA and 3D rotational angiography for the assessment of intracranial aneurysms
    N. Guberina
    U. Lechel
    M. Forsting
    C. Mönninghoff
    U. Dietrich
    A. Ringelstein
    Neuroradiology, 2016, 58 : 673 - 678