Evaluation of two-dimensional dose distributions for pre-treatment patient-specific IMRT dosimetry

被引:5
|
作者
Radojcic, Deni Smilovic [1 ]
Rajlic, David [1 ]
Casar, Bozidar [2 ]
Kolacio, Manda Svabic [1 ]
Obajdin, Nevena [1 ]
Faj, Dario [3 ,4 ]
Jurkovic, Slaven [1 ,5 ]
机构
[1] Univ Hosp Rijeka, Med Phys Dept, Kresimirova 42, Rijeka, Croatia
[2] Inst Oncol LJubljana, Dept Radiat Phys, Ljubljana, Slovenia
[3] Univ Osijek, Fac Med, Osijek, Croatia
[4] Univ Osijek, Fac Dent Med & Hlth, Osijek, Croatia
[5] Univ Rijeka, Dept Med Phys & Biophys, Fac Med, Rijeka, Croatia
关键词
IMRT; 2D dose verification; gamma method; anthropomorphic phantom; WATER; VERIFICATION; PHOTON; TISSUE; BEAMS; FILM;
D O I
10.2478/raon-2018-0019
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background. The accuracy of dose calculation is crucial for success of the radiotherapy treatment. One of the methods that represent the current standard for patient-specific dosimetry is the evaluation of dose distributions measured with an ionization chamber array inside a homogeneous phantom using gamma method. Nevertheless, this method does not replicate the realistic conditions present when a patient is undergoing therapy. Therefore, to more accurately evaluate the treatment planning system (TPS) capabilities, gamma passing rates were examined for beams of different complexity passing through inhomogeneous phantoms. Materials and methods. The research was performed using Siemens Oncor Expression linear accelerator, Siemens Somatom Open CT simulator and Elekta Monaco TPS. A 2D detector array was used to evaluate dose distribution accuracy in homogeneous, semi-anthropomorphic and anthropomorphic phantoms. Validation was based on gamma analysis with 3%/3mm and 2%/2mm criteria, respectively. Results. Passing rates of the complex dose distributions degrade depending on the thickness of non-water equivalent material. They also depend on dose reporting mode used. It is observed that the passing rate decreases with plan complexity. Comparison of the data for all set-ups of semi-anthropomorphic and anthropomorphic phantoms shows that passing rates are higher in the anthropomorphic phantom. Conclusions. Presented results raise a question of possible limits of dose distribution verification in assessment of plan delivery quality. Consequently, good results obtained using standard patient specific dosimetry methodology do not guarantee the accuracy of delivered dose distribution in real clinical cases.
引用
收藏
页码:346 / 352
页数:7
相关论文
共 50 条
  • [41] Linking VMAT pre-treatment QA and patient treatment QA by EPID dosimetry
    Sassowsky, M.
    Manser, P.
    Fix, M. K.
    Frauchiger, D.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2012, 188 (08) : 754 - 754
  • [42] Multi-dimensional patient-specific plan QA on IMRT and RapidArc techniques
    Chen, J.
    EUROPEAN JOURNAL OF CANCER, 2013, 49 : S231 - S231
  • [43] Is it still necessary to perform measured based pre-treatment patient-specific QA for SRS HyperArc treatments?
    Cavalli, Nina
    Bonanno, Elisa
    Borzi, Giuseppina R.
    D'Anna, Alessia
    Pace, Martina
    Stella, Giuseppe
    Zirone, Lucia
    Marino, Carmelo
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (02):
  • [44] Amorphous silicon flat panel imaging devices for pre-treatment verification and in vivo dosimetry of IMRT
    Wendling, M
    McDermott, LN
    van Asselen, B
    Sonke, JJ
    Stroom, J
    van Herk, M
    Mijnheer, BJ
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S15 - S15
  • [45] On the Evaluation of Patient-Specific IMRT QA Using EPID, Dynalog Files, and Patient Anatomy
    Defoor, D.
    Mavroidis, P.
    Gutierrez, A.
    Papanikolaou, N.
    Stathakis, S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S885 - S885
  • [46] Gantry angle dependence pre-treatment patient-specific quality assurance and gamma index variation study
    Thekkadath, S.
    Kummali, A.
    Mm, M.
    Raman, G. R.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S155 - S156
  • [47] Comparison of IMRT Verifications with a Two-Dimensional Array of Diodes and a Commercial Portal Dosimetry
    Guerrero, M.
    Patel, I.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [48] Evaluation of QA Dosimetry System for SRS and SBRT Patient-Specific QA
    Hillman, Y.
    Kim, J.
    Chetty, I.
    Wen, N.
    MEDICAL PHYSICS, 2017, 44 (06) : 3273 - 3273
  • [49] A method for predicting results of γ-evaluation of portal dose images in pre-treatment QA of dynamic IMRT-plans
    Vestergaard, A.
    Petersen, J.
    Hofmann, L.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S182 - S182
  • [50] Absolute dose pre-treatment Portal Dosimetry using the Varian MAASTRO implementation
    Taborda, A.
    Stroom, J.
    Baltes, C.
    Seabra, A.
    Dikaiou, K.
    Greco, C.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S433 - S434