Evaluation of two-dimensional dose distributions for pre-treatment patient-specific IMRT dosimetry

被引:5
|
作者
Radojcic, Deni Smilovic [1 ]
Rajlic, David [1 ]
Casar, Bozidar [2 ]
Kolacio, Manda Svabic [1 ]
Obajdin, Nevena [1 ]
Faj, Dario [3 ,4 ]
Jurkovic, Slaven [1 ,5 ]
机构
[1] Univ Hosp Rijeka, Med Phys Dept, Kresimirova 42, Rijeka, Croatia
[2] Inst Oncol LJubljana, Dept Radiat Phys, Ljubljana, Slovenia
[3] Univ Osijek, Fac Med, Osijek, Croatia
[4] Univ Osijek, Fac Dent Med & Hlth, Osijek, Croatia
[5] Univ Rijeka, Dept Med Phys & Biophys, Fac Med, Rijeka, Croatia
关键词
IMRT; 2D dose verification; gamma method; anthropomorphic phantom; WATER; VERIFICATION; PHOTON; TISSUE; BEAMS; FILM;
D O I
10.2478/raon-2018-0019
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background. The accuracy of dose calculation is crucial for success of the radiotherapy treatment. One of the methods that represent the current standard for patient-specific dosimetry is the evaluation of dose distributions measured with an ionization chamber array inside a homogeneous phantom using gamma method. Nevertheless, this method does not replicate the realistic conditions present when a patient is undergoing therapy. Therefore, to more accurately evaluate the treatment planning system (TPS) capabilities, gamma passing rates were examined for beams of different complexity passing through inhomogeneous phantoms. Materials and methods. The research was performed using Siemens Oncor Expression linear accelerator, Siemens Somatom Open CT simulator and Elekta Monaco TPS. A 2D detector array was used to evaluate dose distribution accuracy in homogeneous, semi-anthropomorphic and anthropomorphic phantoms. Validation was based on gamma analysis with 3%/3mm and 2%/2mm criteria, respectively. Results. Passing rates of the complex dose distributions degrade depending on the thickness of non-water equivalent material. They also depend on dose reporting mode used. It is observed that the passing rate decreases with plan complexity. Comparison of the data for all set-ups of semi-anthropomorphic and anthropomorphic phantoms shows that passing rates are higher in the anthropomorphic phantom. Conclusions. Presented results raise a question of possible limits of dose distribution verification in assessment of plan delivery quality. Consequently, good results obtained using standard patient specific dosimetry methodology do not guarantee the accuracy of delivered dose distribution in real clinical cases.
引用
收藏
页码:346 / 352
页数:7
相关论文
共 50 条
  • [21] Dosimetric evaluation of a 3D printed phantom for patient-specific pre-treatment plan verification
    Makris, D. N.
    Zoros, E.
    Boursianis, T.
    Pappas, E.
    Maris, T. G.
    Efstathopoulos, E. P.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S39 - S40
  • [22] On the use of biomathematical models in patient-specific IMRT dose QA
    Zhen, Heming
    Nelms, Benjamin E.
    Tome, Wolfgang A.
    MEDICAL PHYSICS, 2013, 40 (07)
  • [23] Evaluation of Delivered Dose Distributions and Metrics: A New Patient-specific QA Protocol
    Klaers, J.
    Saito, A.
    Goddu, S. M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S810 - S811
  • [24] APPLICATION OF TWO-DIMENSIONAL MatriXX OF DETECTORS TO VERIFICATION OF TOTAL DOSE DISTRIBUTIONS IN IMRT TECHNIQUES
    Chelminski, K.
    Bulski, W.
    Sobotka, P.
    Rostkowska, J.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : XX - XXI
  • [25] Validation of a quick three-dimensional dose verification system for pre-treatment IMRT QA
    Nakaguchi Y.
    Araki F.
    Ono T.
    Tomiyama Y.
    Maruyama M.
    Nagasue N.
    Shimohigashi Y.
    Kai Y.
    Radiological Physics and Technology, 2015, 8 (1) : 73 - 80
  • [26] Patient-Specific Pre-Treatment VMAT Plan Verification Using Gamma Passing Rates
    Arsalan, M. Z.
    Kakakhel, M. B.
    Shamshad, M.
    Afridi, T. A.
    ATOM INDONESIA, 2023, 49 (01) : 53 - 59
  • [27] Evaluation and Implementation of An IMRT Quality Assurance Procedure to Include Patient-Specific Volumetric Dose Analysis
    Mathews, B.
    Grant, E.
    Maricle, S.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [28] Patient-specific point dose measurement for IMRT monitor unit verification
    Dong, L
    Antolak, J
    Salehpour, M
    Forster, K
    O'Neill, L
    Kendall, R
    Rosen, I
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (03): : 867 - 877
  • [29] A Practical Approach of Small Field Dosimetry Measurement for Patient-Specific IMRT/SRS/SBRT
    Wu, X.
    Popple, R.
    Yuan, Y.
    FIveash, J.
    Shen, S.
    Duan, J.
    Cardan, R.
    Brezovich, I.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [30] A systematic approach to statistical analysis in dosimetry and patient-specific IMRT plan verification measurements
    Qin, Songbing
    Zhang, Miao
    Kim, Sung
    Chen, Ting
    Kim, Leonard H.
    Haffty, Bruce G.
    Yue, Ning J.
    RADIATION ONCOLOGY, 2013, 8