Reliability of measured tidal volume in mechanically ventilated young pigs with normal lungs

被引:14
|
作者
Heulitt, MJ
Holt, SJ
Thurman, TL
Hall, RA
Jo, CH
Simpson, P
机构
[1] Arkansas Childrens Hosp, Little Rock, AR 72202 USA
[2] Univ Arkansas Med Sci, Coll Med, Dept Pediat, Sect Crit Care Med, Little Rock, AR 72205 USA
[3] Univ Arkansas Med Sci, Coll Med, Dept Pediat, Biostat Sect, Little Rock, AR 72205 USA
关键词
respiratory physiology; mechanical ventilation; lung volume measurements; reproducibility of results;
D O I
10.1007/s00134-005-2717-y
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: This study examined whether volumes can be accurately measured at the expiratory valve of a conventional ventilator using pressure support ventilation and positive end expiratory pressure with software compensation for circuit compliance available in the Servo i ventilator. Design and setting: Comparison of two methods for measuring tidal volume in an animal laboratory. Subjects: Twenty healthy, intubated, sedated, spontaneously breathing pigs. Interventions: Volume was measured in ten neonatal-sized and ten pediatric-sized pigs ventilated with the Servo i' ventilator using pressure support ventilation and positive end expiratory pressure with and without circuit compliance compensation. We compared volume measured at the airway opening by pneumotachography to volume measured at the expiratory valve of a conventional ventilator. Measurements and results: The use of circuit compliance compensation significantly improved the agreement between the two volume methods in neonatal-sized piglets (concordance correlation coefficient: with circuit compliance compensation, 0.97; without, 0.87, p=0.002). In pediatric-sized pigs there was improvement in agreement between the two measurement methods due to circuit compliance compensation (concordance correlation coefficient with circuit compliance compensation, 0.97; without, 0.88, p=0.027). With circuit compliance compensation off there was positive bias: mean difference (bias) 2.97 +/- 0.12 in neonatal-sized and 3.75 +/- 0.38 in pediatric-sized pigs. Conclusions: Our results show that volume can be accurately measured at the expiratory valve of a conventional ventilator in neonatal- and pediatric-sized animals.
引用
收藏
页码:1255 / 1261
页数:7
相关论文
共 50 条
  • [1] Reliability of measured tidal volume in mechanically ventilated young pigs with normal lungs
    Mark J. Heulitt
    Shirley J. Holt
    Tracy L. Thurman
    Renée A. Hall
    Chan-Hee Jo
    Pippa Simpson
    Intensive Care Medicine, 2005, 31 : 1255 - 1261
  • [2] Regional tidal lung strain in mechanically ventilated normal lungs
    Paula, Luis Felipe
    Wellman, Tyler J.
    Winkler, Tilo
    Spieth, Peter M.
    Gueldner, Andreas
    Venegas, Jose G.
    de Abreu, Marcelo Gama
    Carvalho, Alysson R.
    Melo, Marcos F. Vidal
    JOURNAL OF APPLIED PHYSIOLOGY, 2016, 121 (06) : 1335 - 1347
  • [3] Influence of tidal volume on left ventricular stroke volume variation measured by pulse contour analysis in mechanically ventilated patients
    Daniel A. Reuter
    Julian Bayerlein
    Matthias S. G. Goepfert
    Florian C. Weis
    Erich Kilger
    Peter Lamm
    Alwin E. Goetz
    Intensive Care Medicine, 2003, 29 : 476 - 480
  • [4] Influence of tidal volume on left ventricular stroke volume variation measured by pulse contour analysis in mechanically ventilated patients
    Reuter, DA
    Bayerlein, J
    Goepfert, MSG
    Weis, FC
    Kilger, E
    Lamm, P
    Goetz, AE
    INTENSIVE CARE MEDICINE, 2003, 29 (03) : 476 - 480
  • [5] Effect of tidal volume and frequency on airway responsiveness in mechanically ventilated rabbits
    Shen, X
    Gunst, SJ
    Tepper, RS
    JOURNAL OF APPLIED PHYSIOLOGY, 1997, 83 (04) : 1202 - 1208
  • [6] TIDAL VOLUME IS NOT ASSOCIATED WITH ACUTE KIDNEY INJURY IN MECHANICALLY VENTILATED CHILDREN
    Greidanus, Y.
    Albers, M.
    Kneyber, M.
    INTENSIVE CARE MEDICINE, 2011, 37 : S325 - S325
  • [7] Tidal Volume Requirement in Mechanically Ventilated Infants with Meconium Aspiration Syndrome
    Sharma, Saumya
    Clark, Shane
    Abubakar, Kabir
    Keszler, Martin
    AMERICAN JOURNAL OF PERINATOLOGY, 2015, 32 (10) : 916 - 919
  • [8] INFLUENCE OF TIDAL VOLUME ON SEDATIVE DOSE REQUIREMENTS IN MECHANICALLY VENTILATED ADULTS
    Jagow, Ben
    Short, Elizabeth
    Kruser, Jacky
    Lizza, Bryan
    CRITICAL CARE MEDICINE, 2018, 46 (01) : 504 - 504
  • [9] End-tidal CO2 as a function of tidal volume in mechanically ventilated infants
    Greer, KJ
    Bowen, WA
    Krauss, AN
    AMERICAN JOURNAL OF PERINATOLOGY, 2003, 20 (08) : 447 - 451
  • [10] Global and Regional Tidal Volume Distribution in Spontaneously Breathing Mechanically Ventilated Children
    van Dijk, Jefta
    Koopman, Alette A.
    Blokpoel, Robert G. T.
    Dijkstra, Sandra
    Markhorst, Dick G.
    Burgerhof, Johannes G. M.
    Kneyber, Martin C. J.
    RESPIRATORY CARE, 2022, 67 (04) : 383 - 393