STUDY ON ON-SITE CALIBRATION METHOD OF STATIONARY NEUTRON DOSIMETER

被引:0
|
作者
Lv, Jinxu [1 ]
Lv, Ning [1 ]
Guo, Huiping [1 ]
Sun, Mingyan [1 ]
Zhao, Kuo [1 ]
Lv, Wenhui [1 ]
Wei, Zhihao [1 ]
Xiao, Qizhan [1 ]
Li, Daji [1 ]
Liu, Yang [2 ]
机构
[1] Xian Res Inst Hitec, Xian, Peoples R China
[2] Xian Guanneng Neutron Detect Technol Co Ltd, Xian, Peoples R China
关键词
Neutron ambient dose equivalent; Energy correction; Radiation measurement; on site calibration;
D O I
暂无
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The on-site calibration system ofstationary neutron ambient dose equivalent instrument is mainly comprised of a small controllable neutron source and a reference neutron ambient dose equivalent instrument. According to the principle of "relative calibration method", a small controllable neutron source continuously emits neutrons at the calibration site to construct a neutron radiationfield. The calibration factor (NB) can be obtained by comparing the response numbers from two instrument, the instrument to be calibrated and the reference instrument, which are symmetrically placed in the neutron radiation field. In order to complete the transferofthe ambient dose equivalent calibration coefficient ofthe reference instrument from National Metrology Center (252CfStandard radiation field) to nuclearfacility site,the calibration coefficient needs to be corrected, that is, multiplied by the "energy correction coefficient". Energy correction coefficient includes: (1) "Instrument Energy Response Correction Coefficient" 1c, for neutron fluence to neutron counting of the instrument, (2) "Conversion Correction Coefficient" kp for neutron fluence to neutron ambient dose equivalent. Tests have shown that the ambient dose equivalent rate measurement error ofthe instrument which have been calibrated but without energy correction was 16.5%, by contrast the measurement error was 2.9% with energy correction. It can be seen that the energy correction is necessary and effective for the measuring instrument ofneutron ambient dose equivalent during the on-site calibration process.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] On-site calibration of a current transformer using a Rogowski coil
    Suomalainen, E-P
    Hallstrom, J.
    2008 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2008, : 468 - 469
  • [32] AC Electronic Load for On-Site Calibration of Energy Meters
    Aurilio, Gianluca
    Gallo, Daniele
    Landi, Carmine
    Luiso, Mario
    2013 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC), 2013, : 768 - 773
  • [33] On-site calibration of a phase fraction meter by an inverse technique
    Rolnik, V.P.
    Seleghim Jr., P.
    Revista Brasileira de Ciencias Mecanicas/Journal of the Brazilian Society of Mechanical Sciences, 2002, 24 (04): : 266 - 270
  • [34] On-site infrasound calibration to correct wave parameter estimation
    Kristoffersen, S. K.
    Le Pichon, A.
    Schwardt, M.
    Vincent, P.
    Doury, B.
    Larsonnier, F.
    Pilger, C.
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2024, 155 (03): : 1780 - 1798
  • [35] Eurotherm gains UKAS accreditation for on-site calibration service
    不详
    HEAT TREATMENT OF METALS, 2004, 31 (03): : 76 - 76
  • [36] ON-SITE CALIBRATION OF WEIGH-IN-MOTION SYSTEMS.
    Izadmehr, Bahman
    Lee, Clyde E.
    Transportation Research Record, 1987, : 136 - 144
  • [37] A reflectance based vicarious calibration with on-site instruments monitoring
    Arai, K
    CALIBRATION AND CHARACTERIZATION OF SATELLITE SENSORS AND ACCURACY OF DERIVED PHYSICAL PARAMETERS, 2001, 28 (01): : 11 - 20
  • [38] Transportable system for on-site calibration of permafrost temperature sensors
    Merlone, Andrea
    Sanna, Francesca
    Coppa, Graziano
    Massano, Laura
    Musacchio, Chiara
    PERMAFROST AND PERIGLACIAL PROCESSES, 2020, 31 (04) : 610 - 620
  • [39] PORTABLE GENERATOR FOR ON-SITE CALIBRATION OF PEROXYACETYL NITRATE ANALYZERS
    GROSJEAN, D
    FUNG, K
    COLLINS, J
    HARRISON, J
    BREITUNG, E
    ANALYTICAL CHEMISTRY, 1984, 56 (03) : 569 - 573
  • [40] Experimental analysis of on-site calibration of electronic instrument transformer
    Tan, Hong-En
    Hu, Hao-Liang
    Lei, Min
    Zhang, Shu-Han
    Li, Qian
    Li, He
    Li, Deng-Yun
    Gaodianya Jishu/High Voltage Engineering, 2010, 36 (12): : 2990 - 2995